8.2 Gas, Combustion and Moisture Analyzers

Key Takeaways

  • Stack oxygen can inform combustion trim. Verify the sample, reference and applicable limits before using the measurement.

  • At fixed cell temperature, ideal zirconia voltage is logarithmic in the ratio of oxygen partial pressures. The named Yokogawa cell uses 50.74 mV per base-10 decade at 750 °C.

  • Verify oxygen and catalyst exposure limits for catalytic sensors. Select infrared detectors by the actual target gas and approved application.

  • Use the appropriate gas for functional checks and calibration. Verify toxic-gas cross-sensitivities and optical detector condition.

  • Report dew point at a specified pressure. For particle counts use the designated size channels and sample flow.

Last updated: November 2026

Gas, Combustion and Moisture Analyzers

This is supplemental instrumentation study from NCCER’s Analyzers and Monitors curriculum. Decide what information the measurement must provide. Stack oxygen can inform combustion trim. A combustible gas detector measures a specified fuel or calibration gas basis. A toxic gas sensor has its own target gas and range. A dew point measurement helps assess moisture at a stated pressure.

Oxygen in combustion service

With too little excess air, poor combustion can leave carbon monoxide and other unburned material. Too much excess air increases heat carried out with flue gas. Follow the combustion operating curves and limits for the equipment. Use the oxygen signal within the approved trim and burner protection architecture.

A zirconia cell has electrodes on a heated solid electrolyte that conducts oxygen ions. The electrodes see the sample and a reference with known oxygen. With the polarity used here, the ideal voltage for oxygen partial pressures is

E=RT4Fln⁡(prefpsample).E=\frac{RT}{4F}\ln\left(\frac{p_{\mathrm{ref}}}{p_{\mathrm{sample}}}\right).

Here RR is the gas constant, TT is absolute cell temperature and FF is the Faraday constant. Four electrons are exchanged per oxygen molecule in the electrode reactions. Equal reference and sample partial pressures give zero ideal cell voltage. The real instrument can exhibit zero and slope deviations. Verify it with the specified reference and calibration gases.

The Yokogawa ZR22 example operates the cell at 750 °C and gives a base-10 coefficient of 50.74 mV. At equal total pressures, the partial-pressure ratio equals the ratio of oxygen fractions. Use a consistent actual wet or dry basis and reference concentration. With a given 20.9% reference, 4.0% sample oxygen gives 50.74log⁡10(20.9/4.0)≈36.450.74\log_{10}(20.9/4.0)\approx36.4 mV. At 1.0%, it rises to about 67.0 mV. These are ideal voltages for the stated orientation and conditions. Reference gas humidity and sample pressure can affect the result. Combustibles such as carbon monoxide or hydrogen can react at the hot detector, consuming oxygen and biasing the observed oxygen low. Verify application suitability and check whether changes in reference gas, calibration or process conditions explain an unexpected reading.

A paramagnetic analyzer uses oxygen's attraction to a magnetic field. In the specified Rosemount X-STREAM design a suspended dumbbell experiences torque. A feedback current returns it toward its reference position, providing the measurement signal. Use the analyzer's stated pressure, background gas, flow and conditioning limits. Compare instruments by the actual sample, accuracy needs and installation.

Select a gas detector for the atmosphere

SensorMeasurement principleWhat to verify in service
Catalytic beadOxidation of combustible gas warms the active bead, changing its resistance relative to a reference bead.Oxygen needed for the reaction, actual gas and calibration response, catalyst contaminants, and overrange recovery.
Nondispersive infrared (NDIR)Measures attenuation at selected infrared wavelengths. A reference band can help track common optical changes.The target absorbs in the selected band. Verify gas-specific calibration and check fouling, liquid and optical faults.
Electrochemical toxic gas sensorGas reaching a working electrode reacts and generates a current that depends on exposure within the specified range. Some designs also have a separate reference electrode.Target, range, cross-sensitivities, ambient conditions, functional response and sensor condition.

For the MSA ALTAIR 5X/5X IR catalytic channel, the maker specifies oxygen above 10% for accurate combustible readings. This is a device performance limit. The work area has separate oxygen acceptance criteria. Inadequate oxygen can suppress the combustible reading. Silicone and other contaminants can reduce sensitivity, so check the instrument with the specified challenge gas and follow the manufacturer's test guidance. For an approved oxygen-deficient application, a hydrocarbon infrared detector can be suitable if its approval and gas response match the target. Infrared absorption needs no combustion reaction. A conventional hydrocarbon NDIR channel can detect its supported hydrocarbons but misses a hydrogen leak. Check optics and device faults even where the detector avoids catalytic poisoning.

A bump test exposes the sensor to a known gas to check indicated response and applicable alarms. Calibration sets the response against the specified gas values. Check the applicable instructions and device status after a failed test before relying on a reading. For example, general-industry permit-space atmospheric testing under 29 CFR 1910.146 uses the specified order of oxygen, combustible gases and vapors, then toxic contaminants. Site procedures specify acceptable conditions and the instruments for the hazards.

Measure moisture or particles as separate quantities

A chilled-mirror hygrometer observes the formation of condensate and measures the mirror temperature at equilibrium. A metal-oxide sensor measures capacitance changes as its sensing layer takes up water vapor. State the pressure at which the dew point is measured or to which it was converted. An optical particle counter uses a light beam and measures scattering from particles in its sample. Compare count and size channels at the specified sample flow. Gas concentration, particle count and dew point require different sensing conditions.

Test Your Knowledge

A Yokogawa zirconia cell is held at 750 °C. The reference is 20.9% oxygen and the reference and sample are at the same total pressure. With the reference-over-sample voltage convention, what happens as sample oxygen drops from 6.0% to 2.0%?

A

The signal drops from 50.7 mV to about 16.9 mV.

B

The signal reaches 0 mV because the sample is below 6.0% oxygen.

C

The signal is fixed because the total pressures are equal.

D

The ideal signal rises from about 27.5 mV to 51.7 mV.

Test Your Knowledge

During a permitted remote sampling check, an ALTAIR 5X oxygen channel indicates 8% oxygen and its catalytic combustible channel reads zero. What does the manual’s stated oxygen requirement imply about the combustible reading?

A

It proves there is no gas because the instrument also measures oxygen.

B

The sample is outside the device’s oxygen condition for an accurate catalytic reading, so a zero reading cannot establish that combustible gas is absent.

C

It is a valid reading whenever the oxygen channel is above zero.

D

It is valid after converting the display from volume percent to ppm.

Test Your Knowledge

Several installed methane catalytic detectors lose sensitivity during an approved challenge after silicone exposure. What is the supported next step?

A

Change the display scaling until the instruments agree with each other.

B

Accept their readings if they still report zero in clean air.

C

Treat catalyst contamination as plausible, check the affected detectors under the approved procedure, and evaluate a methane-rated infrared detector if the application permits it.

D

Replace them with infrared detectors based on an expectation that every combustible gas has the same optical response.