3.5 Occupational and Public Radiation Protection

Key Takeaways

  • Under the EU Directive, 20 mSv is the default annual occupational effective-dose limit. A higher authorised year is an exception with a five-year averaging condition.

  • The eye-lens equivalent-dose limit is separate from whole-body effective dose and from the absorbed dose at which a lens tissue reaction may occur.

  • Staff protection combines shorter exposure, distance from scatter, suitable shielding and dosimeters worn in the position specified by the local programme.

  • After a worker declares pregnancy, review conditions so exposure to the unborn child is as low as reasonably achievable and unlikely to exceed 1 mSv during at least the remainder of pregnancy.

Last updated: November 2026

Identify the person and the dose quantity

Council Directive 2013/59/Euratom distinguishes effective dose for overall radiological protection from equivalent dose to a specified tissue. A lens dose is not the whole-body effective dose. First establish who is exposed, which quantity the question asks for and which jurisdiction applies. The EU Directive provides the baseline for national rules, and workers should follow the requirements implemented in their country. Its dose-limit provisions do not set limits for the justified medical exposure of a patient.

Group or tissueLimit or protection provision under the Directive
Occupational effective dose, Article 920 mSv in any single year by default. In circumstances specified in national legislation, the competent authority may authorise up to 50 mSv in one year if the average in any five consecutive years still does not exceed 20 mSv per year.
Occupational eye lens equivalent dose, Article 920 mSv in a single year or 100 mSv in any five consecutive years with no single year above 50 mSv, as specified in national legislation.
Occupational skin or extremities, Article 9500 mSv in a year. The skin value is averaged over any 1 cm² of skin.
Public effective dose, lens and skin, Article 121 mSv effective dose, 15 mSv to the eye lens and 50 mSv to skin per year. Article 12 does not list a separate extremity limit in this table.
Students and apprentices aged 16 to under 18 who must work with sources during training, Article 116 mSv effective dose per year, 15 mSv to the lens, and 150 mSv to skin and extremities. At 18 or over, occupational limits apply in the specified work-with-sources setting.
Unborn child of a worker after pregnancy is declared, Article 10Organise the working conditions so the equivalent dose to the unborn child is as low as reasonably achievable and unlikely to exceed 1 mSv during at least the remainder of pregnancy.

The ICRP recommendation for the lens and the legally implemented limit address protection. The approximate absorbed dose threshold for a tissue reaction, expressed in Gy, is a different concept. Likewise, occupational categorisation thresholds determine monitoring and surveillance. They do not grant permission to receive a given annual dose. Under the Directive, a worker liable to exceed 6 mSv effective dose per year, 15 mSv to the lens, or 150 mSv to skin or extremities is in category A. Appropriate individual monitoring is required.

Worked example

Suppose four consecutive years of occupational effective dose total 59 mSv. An additional 42 mSv would yield 101 mSv over five years, averaging 20.2 mSv per year. That would fail the five-year condition for an Article 9 exception even though 42 mSv is below the exceptional single-year ceiling of 50. A smaller total that fits the numerical ceilings does not on its own authorise higher exposure. The exception has to apply under national law and be authorised. Good planning aims below the limits.

Control exposure during fluoroscopy

The patient is the main source of scattered radiation to staff. Reduce exposure by combining better technique with positioning, protective equipment and equipment-mounted shields. Keep hands out of the primary beam. A lead glove does not make direct exposure acceptable and may influence automatic exposure if it enters the imaging field.

  • Time. Use the lowest pulse rate adequate for the task and review stored images or last-image hold when possible. Minimise acquisition runs while retaining diagnostic content. Reduce exposure to the patient and scatter to the team by keeping the field limited to the task.
  • Distance. When a procedure permits, step back during imaging. An ideal point-source model predicts one quarter of the intensity at twice the distance. A patient is an extended scatter source, so this is a teaching approximation. During a lateral projection, work on the detector side where feasible because scatter is generally higher on the X-ray tube side.
  • Shielding. Position a ceiling-suspended shield between the patient and the operator's head and use table skirts where available. Wear appropriate protective garments selected and tested under departmental policy. Lead spectacles and thyroid collars help protect specific tissues when used correctly. Garment performance depends on material, fit and beam quality.

Follow the local dosimetry programme. For interventional work, one dosimeter may be worn under the apron to assess protected exposure and another outside near the collar. Lens or extremity monitoring may require a separate dosimeter and position. The collar reading alone does not give the protected whole-body dose. Declared pregnancy calls for a workplace review and may require an abdominal dosimeter under the apron in line with local policy.

Use quality assurance to find recurring problems

A dose review needs a representative sample and the corresponding clinical indication. A cluster of high fluoroscopy dose values can reflect more complex cases, a poorly positioned detector or insufficiently optimised acquisition settings. Check the case mix and record the changes. An unexpected pattern of low staff badge values can reveal missed badge use as well as better protection. Escalate anomalous readings, inspect technique and equipment, and use follow-up measurements to see whether the correction worked.

See Directive 2013/59/Euratom, Articles 8–12 and 35–41 and ICRP Publication 139 on interventional staff protection.

Test Your Knowledge

Which pair correctly distinguishes the approximate ICRP 118 absorbed-dose threshold for vision-impairing lens injury from the occupational eye-lens equivalent-dose rule in Article 9 of the EU Directive?

A

Threshold 2.0 Gy; annual occupational limit 150 mSv/year

B

Threshold 5.0 Gy; annual occupational limit 50 mSv/year

C

Threshold 1.0 Gy; annual occupational limit 100 mSv/year

D

About 0.5 Gy absorbed dose. The legal limit is 20 mSv in one year or 100 mSv over five years with a 50 mSv single-year maximum, as national law specifies.

Test Your Knowledge

Under the EU Directive, four consecutive years of occupational effective dose total 59 mSv. Why would another 42 mSv fail the numerical five-year condition for the authorised exception?

A

The five-year total would be 101 mSv, averaging 20.2 mSv per year.

B

The 42 mSv value exceeds the exceptional single-year maximum of 50 mSv.

C

The 42 mSv value is an eye-lens dose and must be added twice.

D

No averaging condition applies when the annual value is below 50 mSv.