2.3 Magnetic Resonance Imaging (MRI) Physics, Sequences & Artefacts
Key Takeaways
Hydrogen protons precess at the Larmor frequency (omega_0 = gamma B_0), corresponding to ~63.9 MHz at 1.5 T and ~127.7 MHz at 3.0 T
Spin echo refocuses much of the static dephasing to show T2 decay. A gradient echo without RF refocusing retains T2* sensitivity, with contrast depending on timing and flip angle.
STIR can suppress tissue whose T1 is similar to fat, including enhancing tissue. Use suitable T1-weighted spectral fat suppression or Dixon when assessing post-gadolinium enhancement.
True restricted diffusion is hyperintense on high-b DWI and hypointense (dark) on the ADC map, whereas T2 shine-through remains bright or isointense on the ADC map
In conventional Cartesian imaging, chemical shift displaces fat along readout. Wrap-around and ghosting often affect the phase direction, but they have different causes and remedies.
Magnetic Resonance Imaging (MRI) Physics, Sequences & Artefacts
Unlike X-ray or CT, Magnetic Resonance Imaging (MRI) uses non-ionising radiofrequency (RF) pulses within a strong static magnetic field () to interrogate mobile hydrogen nuclei (). The ESR curriculum includes hardware, relaxation, image weighting, sequence selection and artefacts. Apply these concepts when deciding whether unexpected signal represents a lesion, a change in contrast or an acquisition artefact.
Larmor Frequency, Relaxation & Sequence Weighting
Larmor Equation & Magnetisation
Placed in a static magnetic field (typically 1.5 T or 3.0 T), hydrogen protons precess around the field axis at the Larmor frequency:
For , the gyromagnetic constant , yielding resonance frequencies of ~63.9 MHz at 1.5 T and ~127.7 MHz at 3.0 T. An RF pulse at resonance tips some net longitudinal magnetisation into the transverse plane. Its precessing transverse component induces a signal in the receiver coil. Gradients make resonance frequency or phase depend on position, enabling slice selection and spatial encoding. For example, a stronger slice-selection gradient gives a thinner slice for the same excitation bandwidth.
T1, T2 & T2* Relaxation
- T1 (Longitudinal / Spin-Lattice) Relaxation: Time for to recover 63% () of the way back to equilibrium by transferring energy to the molecular lattice. Short T1 tissues recover quickly and appear bright (hyperintense) on T1-weighted (T1W) images: fat or tissue enhanced by a gadolinium-based contrast agent. Subacute blood containing methaemoglobin may also be bright on T1-weighted images. Long T1 tissues (CSF, simple cysts, oedema) appear dark on T1W. T1 generally increases with field strength, with tissue-dependent changes.
- T2 (Transverse / Spin-Spin) Relaxation: Time for to decay to 37% () of initial transverse signal via dynamic spin-spin dephasing. Long T2 tissues (CSF, fluid, oedema, most neoplasms) remain bright on T2W; Tendons generally have short T2 and low signal. Air and cortical bone also give little signal on routine sequences because they contain few mobile protons or have very short-lived signal.
- Decay: Rapid transverse dephasing driven by both intrinsic T2 interactions and static local magnetic field inhomogeneities (): An effective-rate model writes , with the extra term representing dephasing from field variation. The observed decay also depends on how the field varies across the voxel.
TR/TE Weighting & Spin Echo vs Gradient Echo
| Sequence Weighting | Repetition Time (TR) | Echo Time (TE) | Dominant Contrast & Classic Bright Structures |
|---|---|---|---|
| T1-Weighted (T1W) | Short | Short | Fat, gadolinium enhancement, subacute methaemoglobin, anatomy |
| T2-Weighted (T2W) | Long | Long | CSF, oedema, cysts, inflammation, pathology |
| Proton Density (PDW) | Long | Short | Proton density ( concentration); ideal for menisci, cartilage, ligaments |
- Spin Echo (SE) / Fast Spin Echo (FSE/TSE): A conventional spin echo uses a excitation followed by a refocusing pulse. Refocusing reverses much of the static field dephasing and retains sensitivity to intrinsic T2 decay. Fast or turbo spin echo acquires several echoes per repetition and may use variable refocusing angles. The table describes conventional spin-echo weighting, and the timing that gives a given contrast depends on the scanner, field and tissue.
- Gradient Echo (GRE): Uses gradient reversal to form an echo without a RF refocusing pulse. Many spoiled gradient-echo protocols use a small flip angle, short TR and short TE for rapid three-dimensional imaging. Longer TE in a susceptibility-sensitive GRE technique allows more dephasing around blood products or other local susceptibility sources.
Inversion Recovery, Fat Suppression & Dixon
Inversion Recovery (IR) prepends a inversion pulse followed by an Inversion Time (TI) before excitation. With ideal inversion from equilibrium and a long TR, the signal crosses zero at . For tissue with of 250 ms, the ideal null is about 173 ms. A shorter TR, incomplete inversion or different field strength can change the setting needed.
- STIR (Short Tau Inversion Recovery): Uses a short TI chosen near the null point of fat. It often gives useful fat suppression across a wide field because it relies on T1 rather than a frequency-selective saturation pulse. Tissue with a comparable T1 can also lose signal. Use a suitable T1-weighted spectral or Dixon technique to assess gadolinium enhancement, which STIR may obscure.
- FLAIR (Fluid-Attenuated Inversion Recovery): Uses a long TI to suppress cerebrospinal fluid (CSF) and a long TE for T2 weighting. This can reveal a periventricular lesion that might blend with the bright CSF on conventional T2-weighted images.
- Chemical (Spectral) Fat Saturation: Delivers a narrow-band RF pulse tuned to fat's resonant frequency (roughly 3.4–3.5 ppm below water, or about 220 Hz at 1.5 T and 440 Hz at 3.0 T), followed by a spoiler gradient. It preserves gadolinium enhancement, though field variation can make suppression uneven or inadvertently affect water signal.
- Dixon Water-Fat Separation: Acquires GRE images at In-Phase (IP) and Out-of-Phase (OP) echo times (at 1.5 T: OP ~2.3 ms, IP ~4.6 ms; at 3.0 T: OP ~1.15 ms, IP ~2.3 ms) and algebraically reconstructs water-only (useful for T1-weighted post-contrast fat suppression), fat-only, IP, and OP images. Modern Dixon methods estimate field variation and can be robust over a large field, although water–fat swaps can occur. Signal dropout on OP relative to IP suggests mixed fat and water in a voxel, as in hepatic steatosis or a lipid-rich adrenal adenoma.
Diffusion-Weighted Imaging (DWI), SWI & MRA
DWI & Apparent Diffusion Coefficient (ADC)
Diffusion-Weighted Imaging (DWI) can use paired motion-sensitive gradients (Stejskal–Tanner) around a spin-echo refocusing pulse, usually followed by an echo-planar readout in routine clinical DWI. Signal decays exponentially with b-value (diffusion sensitivity, ): . Higher -values increase diffusion weighting and typically reduce signal-to-noise ratio (SNR). For the same baseline signal, a structure with a lower ADC loses less signal at the same -value.
- True Restricted Diffusion: Seen in cytotoxic oedema (acute ischaemic stroke), viscous pus (abscess), and some hypercellular lesions. Appears bright (hyperintense) on high- DWI AND dark (hypointense) on the ADC map.
- T2 Shine-Through: Tissues with intrinsically very long T2 (for example, vasogenic oedema or a subacute infarct with risen or pseudonormalised ADC) appear bright on DWI due to T2-weighting, but remain bright or isointense on the ADC map.
Susceptibility-Weighted Imaging (SWI) & MRA
- Susceptibility-Weighted Imaging (SWI): A 3D flow-compensated GRE sequence combining magnitude and filtered phase maps. While both haemorrhage and calcification cause dark magnitude blooming, paramagnetic blood products (deoxyhaemoglobin, haemosiderin) and diamagnetic calcium induce opposite phase shifts, which can help separate blood products from calcification. Verify the phase convention for the system and use CT when the distinction remains uncertain.
- Time-of-Flight (TOF) MRA: Non-contrast GRE technique exploiting flow-related enhancement (fresh unsaturated blood entering a slice where stationary tissues are RF-saturated); vulnerable to in-plane flow saturation and turbulent dephasing (may overestimate stenosis).
- Phase-Contrast (PC) MRA: Applies bipolar velocity-encoding gradients (VENC); phase shift scales linearly with flow velocity, measuring flow direction and quantification (e.g., CSF aqueductal stroke volume), though wrapping occurs if velocity exceeds VENC.
- Contrast-Enhanced (CE) MRA: Rapid 3D T1W GRE timed to first-pass IV gadolinium; gadolinium-shortened blood T1 provides contrast with less dependence on inflow. Timing matters because venous enhancement can overlap the arteries.
Beyond routine structural sequences
- Dynamic contrast-enhanced MRI (DCE MRI) repeats a T1-weighted acquisition during and after the injection. Timing and the shape of the enhancement curve provide information about delivery and tissue exchange. Quantitative results depend on the acquisition and the kinetic model. A vessel in the ROI or patient motion can distort the curve.
- Diffusion tensor imaging (DTI) samples diffusion in several directions and models how diffusion varies by direction. It can help map white-matter orientation. A crossing fibre bundle or partial volume can make a simple tensor model misleading.
- Functional MRI using blood oxygenation level-dependent (BOLD) contrast exploits sensitivity to local deoxyhaemoglobin-related susceptibility. A task-related change is an indirect haemodynamic marker and depends on neurovascular coupling.
- Proton MR spectroscopy separates resonance peaks by frequency to estimate the relative signals of tissue metabolites. Voxel placement, field homogeneity and sequence parameters affect the spectrum. Compare the spectrum with the anatomical images and avoid diagnosing a lesion from a single peak.
See the DCE parameter standard, diffusion tensor method, and BOLD contrast experiment.
MRI Artefacts & Axis Rules
| Artefact | Mechanism & Axis | Key Diagnostic Fix |
|---|---|---|
| Chemical Shift Type 1 | Fat resonates ~3.5 ppm slower than water, shifting fat pixels along the frequency-encoding axis (bright/dark rim at kidney-fat border) | Increase receiver bandwidth per pixel or apply suitable fat suppression |
| Chemical Shift Type 2 (India ink) | Water and fat cancel in mixed voxels at opposed-phase echo times, often giving a dark outline | Compare the in-phase echo. Use the signal loss intentionally when assessing mixed fat and water. |
| Susceptibility | Local distortion near air, bone, metal, or blood causes signal dropout/geometric warp (severe on EPI/GRE) | Switch GRE to SE/TSE, shorten TE, increase readout bandwidth or use a dedicated metal artefact reduction technique if needed |
| Wrap-Around (Aliasing) | Anatomy outside the FOV along the phase-encoding axis folds onto the opposite edge | Phase oversampling (no-phase-wrap), enlarge FOV, swap phase/frequency axes |
| Gibbs (Truncation) | Finite Fourier k-space sampling causes parallel ringing lines near sharp interfaces (can mimic cervical cord syrinx) | Acquire a higher-resolution matrix if practical. Zero-filling alone does not add acquired high-frequency data. |
| Magic Angle | Collagen fibres angled at () to exhibit prolonged T2, looking spuriously bright on short-TE (T1W, PDW, GRE) | Check longer-TE T2-weighted images and tendon morphology |
| Motion / Ghosting | Periodic respiration, heartbeat, or pulsation often creates ghosts along the phase-encoding axis in Cartesian imaging | Respiratory/ECG gating, presaturation bands, swap phase axis, radial/PROPELLER k-space |
See the IAEA Diagnostic Radiology Physics handbook, chapters 14–15, Dixon’s original water–fat study, the serial stroke diffusion study and the tendon orientation study.
Following intravenous gadolinium administration for suspected cervical spine spondylodiscitis, which fat-suppression technique should be avoided because it risks obscuring true contrast enhancement?
Dixon water-only T1-weighted imaging
Short Tau Inversion Recovery (STIR)
Frequency-selective chemical fat saturation
T1-weighted Spin Echo without inversion recovery
A cortical lesion has high signal on diffusion-weighted MRI at b = 1000 s/mm², elevated ADC and high T2 signal. Which mechanism best explains the persisting brightness on the diffusion-weighted image?
Hyperacute cytotoxic oedema causing maximal water restriction
Susceptibility blooming from acute deoxyhaemoglobin
T2 shine-through arising from prolonged intrinsic T2 relaxation
Gibbs truncation ringing along the cortical ribbon
Match each MRI artefact with its characteristic physical mechanism and primary corrective strategy.
Match each item on the left with the correct item on the right
Chemical shift Type 1
Wrap-around (aliasing)
Gibbs truncation ringing
Magic angle effect (~55 degrees)