2.5 Image Processing, Registration and Quantification
Key Takeaways
Image content above the Nyquist frequency can produce aliasing. Interpolation cannot add detail from unmeasured spatial frequencies.
Smoothing, sharp kernels and deconvolution alter the resolution–noise trade-off. Assess performance at the contrast and size of the clinical target.
Compare the same anatomy and technique when measuring attenuation in a region of interest.
Check the original images when misregistration or partial volume could change the apparent result.
Trace the finding from acquisition to display
A cross-section is the product of sampling, reconstruction and display. In a case, decide whether a finding is present in the acquired data, introduced by processing, or visible only under a particular display setting. An edge seen on one sharp reconstruction can have a different apparent thickness on a smooth series even when both come from the same CT acquisition. Compare the series before treating the change as disease. The desired balance of detail and noise depends on the task. A thin cortical fracture and a small low-contrast liver lesion present different problems.
Sampling and linear systems
A detector or voxel averages information over a finite area or volume. Spatial sampling sets the spacing between recorded data points. Finer sampling can represent smaller detail if the underlying detector and reconstruction also provide that resolution. For ideal sampling at frequency , the Nyquist frequency is . Detail above that limit can alias into a false lower-frequency pattern. Examples include grid-line interference in digital radiography and wrap-around from undersampling the field of view in Cartesian MRI. Enlarging the image or interpolating more pixels leaves the original samples unchanged.
A simple linear, position-invariant model describes blur as convolution of the object with a point-spread function. The point-spread function is the image produced by a very small object. A wider function spreads its signal farther across neighbouring pixels. A Fourier transform describes an image or signal as spatial-frequency components. Slow variations give lower spatial frequencies. Edges contain higher frequencies. The system's modulation transfer function describes how much modulation is preserved at each frequency. Deconvolution attempts to reverse a known blur, but it can amplify noise where the signal was strongly attenuated. Investigate a method at clinically relevant sizes and contrast levels before relying on its apparent sharpness.
| Processing step | Reason to use it | Common error |
|---|---|---|
| Smoothing | Reduce visible noise for a low-contrast task | Small features or edge definition can be lost. |
| Edge enhancement or sharp kernels | Improve visibility of narrow high-contrast boundaries | Noise, edge overshoot or streaks can appear more conspicuous. |
| Windowing | Allocate available grey levels to a clinically relevant attenuation range | A display adjustment changes appearance while leaving the stored quantitative values unchanged. |
| Segmentation | Outline a structure for measurement or visualisation | Motion, partial volume and threshold settings can move the selected boundary. |
| Registration and fusion | Put data into a common coordinate system and view them together | Patient motion, breathing or deformation can place one image's finding over the wrong anatomy in another. |
Quantify a signal that can be compared
A region of interest (ROI) measures signal or attenuation in a chosen area or volume. A small ROI can avoid adjacent structures but may have noisier results. A large ROI gives more averaging and can mix tissues across an edge. Copy or reproduce the ROI only after checking slice position, anatomy, acquisition phase and registration. Record what was measured, which source series was used, and whether the value is a mean or a maximum. In near-uniform tissue, the standard deviation of values can help evaluate noise. In a lesion, that same statistic can reflect true heterogeneity as well as noise.
In an illustrative contrast study, a lesion measures 40 HU unenhanced, 65 HU early and 55 HU on a later phase, with comparable and correctly placed ROIs. It has increased 25 HU at the early time point and remains 15 HU above baseline later. Whether that pattern supports a specific diagnosis depends on the examination, contrast agent, organ and phase definition. Moving an ROI partly outside a lesion can manufacture apparent washout, so inspect the contours on each image. A partially imaged lung base or a single contrast phase cannot answer every question that a dedicated protocol would answer.
Time–activity curves use a measured PET or SPECT signal from a region over time. Dynamic acquisitions permit comparison of early delivery and later retention or clearance. A curve depends on scan timing, the region, motion and reconstruction. Factor analysis can describe dynamic data as combinations of a small number of time curves with spatial weights. A calculated component is a mathematical estimate. It needs anatomical or physiological validation before assigning it to a specific tissue. In dynamic CT or MRI, analogous time curves can help estimate perfusion or contrast kinetics, with estimates depending on temporal sampling, input curve selection and the analysis model.
Recognise when processing cannot undo poor input
A reconstruction can change the representation of the available data. It cannot reliably restore information that was never measured, such as through-plane detail lost in a thick acquisition or an artery that was not opacified at scan time. A projection image, for example a maximum-intensity projection of CT angiography, emphasises high attenuation along a line. A small vessel can be hidden by nearby bone or overlapping contrast. Confirm a suspected filling defect on the source slices and suitable multiplanar reformations.
See the IAEA Diagnostic Radiology Physics handbook, the DICOM windowing specification, and the IAEA nuclear medicine physics handbook. See the study of factor analysis in dynamic PET for a practical example of extracting time–activity curves.
On two properly aligned CT series, an ROI in a lesion measures 40 HU unenhanced and 65 HU early after contrast. What is the measured early increase?
15 HU
40 HU
105 HU
25 HU
A narrow CT angiography reformatted image suggests a short filling defect next to bone. Which next step best assesses whether it is real?
Review thin source slices and another plane to check for overlap or partial volume.
Enlarge the projected image so that interpolation replaces missing through-plane data.
Increase dose on the reconstruction workstation.
Use the appearance of a single projection as definitive evidence.