2.4 Ultrasound & Doppler Physics, Resolution & Artefacts
Key Takeaways
Diagnostic scanners assume about 1540 m/s in soft tissue. A large impedance mismatch reflects more sound, so coupling gel helps transmit the beam through the skin.
A short spatial pulse improves axial resolution. Beam width controls lateral resolution, and higher frequencies generally trade penetration for detail.
Mechanical and thermal indices estimate potential acoustic effects. Select the output and scanning time needed for the task, and use a contrast-specific low-output mode for routine microbubble imaging.
The Doppler shift depends on the cosine of the beam-to-flow angle. A perpendicular beam can show no shift despite patent flow.
Increasing PRF expands the measurable Doppler span. Lowering transmit frequency reduces the shift. Moving the baseline redistributes the displayed range at the same PRF.
Ultrasound & Doppler Physics, Resolution & Artefacts
Medical ultrasound images anatomy and real-time haemodynamics without ionising radiation by transmitting high-frequency longitudinal mechanical pressure waves from piezoelectric crystals. The ESR curriculum includes these physical trade-offs, Doppler sampling and contrast-enhanced ultrasound (CEUS). A good case answer links the appearance to the settings or tissue property that produced it.
Wave Propagation, Acoustic Impedance & Resolution
Propagation Speed & Acoustic Impedance
Diagnostic scanners assume an average soft-tissue propagation speed of . Wavelength follows : a 5 MHz wave has in soft tissue, compared with at 10 MHz.
Acoustic impedance () is the product of tissue density () and propagation velocity: (measured in Rayls). At normal incidence, the intensity reflection coefficient is:
- At many soft-tissue interfaces only a small fraction of incident energy is reflected, allowing a portion of the beam to travel deeper. The fraction depends on the two impedances.
- A soft tissue–air interface reflects almost all the incident beam at normal incidence. Coupling gel removes trapped air between skin and probe. Bone strongly attenuates transmitted ultrasound through reflection and absorption. Gas may cause a dirty shadow with reverberation rather than one uniformly dark shadow.
Transducer Selection & Spatial Resolution
| Resolution Axis | Governing Determinant | Depth Dependence & Optimisation |
|---|---|---|
| Axial Resolution (parallel to beam) | (half the Spatial Pulse Length, ) | A shorter pulse improves axial resolution. Higher frequency shortens wavelength, and broader transducer bandwidth can shorten the pulse by reducing its cycle count. |
| Lateral Resolution (perpendicular to beam) | Beam width within the scan plane | Best at the focal zone (narrowest waist). Improved by higher frequency, wider aperture, and placing the focal zone at target depth |
| Elevational Resolution (slice thickness) | Mechanical acoustic lens / 1.5D-2D array geometry | Governs partial-volume averaging across slice thickness (can fill small cysts with spurious low-level echoes) |
Frequency Trade-Off: Attenuation in soft tissue rises roughly linearly with frequency (the coefficient depends on the tissue). High-frequency linear probes deliver crisp axial and lateral resolution for superficial structures (thyroid, breast, scrotum, tendons, peripheral veins), whereas low-frequency curvilinear probes sacrifice fine resolution to achieve deep abdominal/pelvic penetration, and small-footprint phased-array probes fit between intercostal spaces.
Connect the display mode to the clinical task
A-mode displays echo amplitude against depth along one beam. B-mode encodes echo amplitude as brightness to build a two-dimensional slice. M-mode plots motion against time along a selected line, with high temporal sampling for structures such as a cardiac valve. Target depth, sector width and line density affect frame rate. A wider colour region can require more transmissions and reduce temporal resolution. A smaller field of view can improve frame rate when the target remains in view.
Applied voltage makes the piezoelectric element deform and generate a sound pulse. Returning pressure changes produce electrical signals in the same element or array. Centre frequency depends on element design, including the thickness of the active material. Timing offsets across array elements steer and focus the beam. For a deep target, optimise the acoustic window and transmit frequency before increasing output. An endoluminal probe, where appropriate and consented, brings the transducer closer to some structures and can support a higher frequency than a transcutaneous window. Speckle is a granular interference pattern from many small scatterers. Confirm that an apparent small lesion persists with a changed view and adequate focusing.
Gain, TGC & Acoustic Safety Indices
- Overall Gain vs Time Gain Compensation (TGC): Overall gain uniformly amplifies all returning electrical signals (changing display brightness without altering patient acoustic exposure). TGC applies depth-specific amplification to counter exponential attenuation so homogeneous liver parenchyma displays uniform echogenicity from near to far field.
- Mechanical Index (): The displayed estimate is the derated peak rarefactional pressure (in MPa) divided by the square root of centre frequency (in MHz). MI gives an indication of potential mechanical effects such as cavitation, rather than a measured patient injury risk. Choose an appropriate preset and output for the examination.
- Thermal Index (): An output estimate based on the power needed to raise temperature by about under a specified model: TIS for soft tissue, TIB with bone near the focus, TIC with bone near the probe. It does not measure the actual temperature rise. Use the lowest output and shortest dwell time consistent with the diagnostic task. In early pregnancy, use pulsed Doppler only with a clinical indication and follow gestational-age-specific ISUOG safety guidance.
Essential Ultrasound Artefacts
| Artefact | Physical Origin | Classic Clinical Example |
|---|---|---|
| Posterior Acoustic Shadowing | Extreme reflection or absorption depletes downstream beam energy | "Clean" sharp shadow behind gallstones, renal calculi, or bone; "dirty" heterogeneous shadow behind gas |
| Posterior Acoustic Enhancement | Lower attenuation through fluid leaves stronger echoes behind it than at the same depth through adjacent tissue | Bright band posterior to simple renal/hepatic cysts, gallbladder, or urinary bladder |
| Reverberation, Comet-Tail & Ring-Down | Repeated reflections create reverberation and some closely spaced echoes produce comet-tail. Gas can also generate prolonged ring-down echoes. | Horizontal pleural A-lines and short comet-tail from cholesterol crystals in gallbladder adenomyomatosis |
| Mirror Image | Specular reflection off a strong curved interface (diaphragm/pleura) | Duplicate hepatic haemangioma or ascites appearing supradiaphragmatically inside lung base |
| Edge Shadowing | Refraction/deflection at curved lateral borders of smooth cystic walls | Thin hypoechoic lines extending from lateral margins of gallbladder or simple cysts |
| Anisotropy | Specular reflection depends on beam angle relative to ordered collagen fibrils | Tendon or nerve appears falsely hypoechoic (mimicking tear/tendinopathy) when insonated obliquely. Re-angle to see whether the apparent defect disappears. |
Doppler Physics, Aliasing & CEUS
The Doppler Equation & Angle Pitfalls
Moving erythrocytes shift the reflected frequency according to the Doppler equation:
where is transmitted frequency, is blood flow velocity, , and is the angle between the ultrasound beam and the flow direction. For quantitative vascular measurements, follow the protocol. A beam-to-flow angle of or less is commonly used because small cursor errors give larger fractional velocity errors at steep angles. The bias can go either way. At , the predicted shift is zero even when blood is moving. Re-angle or steer the beam and assess calibre, compressibility and other findings before calling occlusion.
Nyquist Limit & Aliasing Correction
Pulsed-wave (PW) and colour Doppler sample echoes intermittently at the Pulse Repetition Frequency (PRF). With a centred baseline the unambiguous shift is bounded by the Nyquist limit on either side (). A shift beyond that range is displayed at an incorrect frequency. It can wrap across the baseline on the spectrum or appear as a colour reversal. A colour mosaic alone does not prove turbulence. For example, a 3 MHz beam at to flow of 1 m/s gives a shift of about 1.95 kHz. That wraps at 3.5 kHz PRF (1.75 kHz on either side) but fits at 5 kHz PRF (2.5 kHz on either side), assuming those rates are available at the sample depth.
Choose the adjustment according to what needs to be measured:
- Increase PRF (raise the velocity scale) — raises the Nyquist limit directly.
- Shift the Baseline downward (or upward) — can allocate more of the available spectral display to the known direction of flow. It leaves PRF and the total sampled frequency span unchanged.
- Lower Transducer Frequency () — decreases for the same blood velocity.
- If the depth limits the available PRF, use a shallower acoustic window if the same sample site can be reached. Preserve comparable angles for serial measurements.
- Switch to continuous-wave (CW) Doppler when appropriate, for example to measure a high-velocity cardiac jet. It avoids pulsed range-gated aliasing but combines velocities along the beam.
Contrast-Enhanced Ultrasound (CEUS)
In intravenous CEUS, stabilised microbubbles reflect ultrasound with a nonlinear response. A contrast-specific low-output mode separates bubble echoes from the tissue background and limits destruction during the examination. Sulphur hexafluoride microbubbles used for vascular imaging are blood-pool tracers, with pulmonary exhalation of the gas. That clearance is useful when renal function is impaired. Apply the approved agent’s contraindications and requirements for hypersensitivity monitoring. The mechanical index depends on the scanner, agent and application. For a liver lesion, record when it enhances and when it becomes less enhanced than the surrounding liver. The timing and degree of washout matter, along with the patient population and the contrast agent used.
See the IAEA ultrasound chapter, EMA SonoVue product information, and the ESUR Contrast Media Safety Committee guidelines.
During a lower-limb venous Doppler ultrasound, a linear transducer is placed strictly perpendicular (90 degrees) to the mid-femoral vein, and no colour flow signal is seen within the lumen despite full venous compressibility. What explains this finding?
Acoustic impedance mismatch between blood and venous endothelium reflects 99.9% of the beam
High transducer frequency causes excessive Nyquist aliasing that cancels all colour pixels
Posterior acoustic enhancement from venous blood masks the colour Doppler signal
At a 90-degree insonation angle, cos(90) equals zero, producing zero Doppler frequency shift
During pulsed-wave Doppler in a stenotic renal artery, the high-velocity peak wraps below the baseline. Which change increases the sampled frequency span or reduces the shift for the same flow?
Increase the Pulse Repetition Frequency (PRF) or switch to a lower-frequency transducer
Increase the transmitted transducer frequency (f0) to improve axial resolution
Decrease the Doppler angle from 55 degrees down toward 0 degrees
Increase overall receiver gain and raise the wall filter
A patient with stage 5 chronic kidney disease needs vascular characterisation of an indeterminate hepatic lesion. Which combination explains why intravenous sulphur hexafluoride CEUS can be used when otherwise appropriate and supported by the approved indication?
Microbubbles undergo rapid renal tubular clearance; high Mechanical Index (>1.5) is needed to maximise signal
Microbubbles diffuse freely into the tumour interstitium; continuous high-PRF spectral Doppler is mandatory
Gas leaves via the lungs. A contrast-specific low mechanical index mode limits bubble destruction during real-time vascular imaging.
Microbubbles bind selectively to hepatocytes after 20 minutes; thermal index must exceed 2.0 to activate shells