2.2 Computed Tomography (CT) Physics, Protocols & Artefacts

Key Takeaways

  • Water defines 0 HU and air is near −1000 HU. Tissue measurements vary with composition, partial volume and acquisition, so interpret attenuation with the image and the protocol.

  • Window level centres the displayed range; window width controls its span. A narrow brain window increases contrast between in-range grey and white matter without changing voxel values.

  • For a helical scan at fixed weighted CT dose index, doubling pitch halves the volume index. At fixed scan length the DLP also halves; automatic exposure control can change that relationship.

  • A reference-phantom CT dose index is not an individual organ dose. Size-specific dose estimates and approximate effective doses are different quantities used for different purposes.

  • Dual-energy CT can generate iodine maps and virtual monoenergetic images. Higher-energy images can reduce artefact from metal, while lower-energy images can increase iodine conspicuity.

Last updated: November 2026

Read the image as a map of attenuation

In multidetector computed tomography (CT), a rotating X-ray source and detector array measure transmission through the patient from many directions. Reconstruction estimates the attenuation in each voxel. In helical scanning, the table moves during rotation so a volume can be acquired in one continuous run. An artefact can be introduced during acquisition, reconstruction or display; identifying which stage is responsible helps you decide whether a different display, a new reconstruction or a repeat acquisition is necessary.

Hounsfield units (HU) express attenuation relative to water. In conventional CT, the relation is approximately:

HU=1000×μtissue−μwaterμwater\text{HU} = 1000 \times \frac{\mu_{\text{tissue}} - \mu_{\text{water}}}{\mu_{\text{water}}}

Water is the 0 HU reference and air is near −1000 HU. Fat has negative attenuation but is denser than air, and cortical bone has much higher attenuation than either. Within a patient, motion, partial volume, the energy spectrum and intravenous contrast affect the measured value. Do not diagnose a lesion from an isolated attenuation number. For example, high attenuation in a collection on a contrast study can reflect haemorrhage, iodine or a mixture; use the phases, anatomy and other findings.

Window level sets the centre of the attenuation range shown in grey, and window width sets its size. In the usual CT greyscale, values below the range appear black, values above it white. For simple mental arithmetic, the limits are approximately level minus and plus half the width. A width of 80 HU centred at +40 HU assigns the display range to approximately 0–80 HU. Cortical values of +35 and +40 HU separate better in this window than in a 2000-HU bone window. A narrower window increases display contrast for values it contains but clips more values outside it; it does not change voxel values or reduce the noise in the acquisition. In an acute head CT, review a brain preset for haemorrhage and use a narrower display as an additional look for subtle grey-white loss.

Pitch, collimation and dose

For a helical acquisition, pitch is the table travel in one complete rotation divided by the total nominal beam collimation, not the reconstructed slice thickness. For example, 24 mm per rotation with 30 mm total nominal collimation gives 0.8. When pitch is below 1, successive beam footprints overlap along the table axis. Higher pitch advances the table faster at fixed rotation time. Do not infer a gap in the reconstructed slices from pitch alone: interpolation and reconstruction can supply a contiguous image series. Nor does thinner acquisition collimation by itself guarantee isotropic voxels. For a near-isotropic reconstruction, also choose an in-plane pixel size, reconstructed thickness and interval suitable for the task. The value of thin sections is clearer multiplanar detail; their trade-off is higher noise unless the technique or reconstruction compensates.

Dose indicatorWhat it meansInterpretation trap
Volume CT dose index (CTDIvol\text{CTDI}_{\text{vol}}, mGy)Scanner-output index referenced to a standard CT dosimetry phantom, typically 16 or 32 cm in diameterIt is not the individual patient's mean or organ dose. The phantom size matters when comparing reports.
Dose–length product (DLP, mGy·cm)The volume index multiplied by scan length for a series; add series DLPs to obtain the examination totalIt is not energy in joules. A longer scan or added phase can raise total DLP without changing the per-series volume index.
Size-specific dose estimate (SSDE, mGy)Size adjustment to the phantom-referenced volume indexIt is not the same as an organ-dose estimate and should not be substituted for the scanner index in a DLP-based effective dose calculation.
Estimated effective dose (mSv)DLP multiplied by a suitable approximate region-specific coefficient for a reference populationUseful for broad comparisons; it does not quantify the risk to a particular patient.

Worked example: change pitch, not every other variable

Consider an illustrative chest acquisition with a weighted CT dose index of 12 mGy and length of 30 cm. Hold kVp, tube current, rotation time, collimation and scan length fixed. In this model CTDIvol=CTDIw/pitch\text{CTDI}_{\text{vol}} = \text{CTDI}_w/\text{pitch}. At pitch 0.8, the volume index is 15 mGy and DLP is 450 mGy·cm. At pitch 1.6, the results are 7.5 mGy and 225 mGy·cm. If using the same illustrative adult chest coefficient of 0.014 mSv per mGy·cm, the effective-dose estimates are 6.3 and 3.15 mSv. The arithmetic shows a relative change, not a recommendation to use either protocol.

In clinical practice, automatic exposure control can raise the tube current when pitch rises to maintain a target image quality, so do not predict a 50% dose saving without checking the actual settings. Tube-current modulation can adapt output along the body and around the patient according to attenuation. An off-centre patient changes how the localiser represents their size and how the bow-tie filter places radiation across the body. Depending on localiser direction and scanner design, exposure can be higher or lower than intended; positioning at isocentre helps avoid both errors.

Reconstruction and spectral CT

  • Filtered back projection (FBP) filters the measured projection data before back-projecting them into an image. Changing the reconstruction kernel changes the noise–sharpness balance: a sharp kernel can show fine lung or bone detail but makes noise more visible in soft tissue.
  • Iterative reconstruction starts with an estimate and updates it using the measured data and, depending on the method, a scanner or noise model. It can suppress noise and improve image quality in a lower-exposure protocol. High settings can change the appearance of texture, so a cleaner-looking image is not by itself proof that small low-contrast findings are still detectable.
  • Deep-learning reconstruction uses a trained model to estimate a less noisy image or cleaner data. Its performance and achievable dose depend on the scanner, algorithm, protocol and diagnostic task. Dose changes need validation with the lesions and clinical questions the scan must answer.
  • Dual-energy CT (DECT) uses measurements from two X-ray spectra. The system can distinguish materials by differences in energy-dependent attenuation. The spectra may come from two sources, rapid tube-voltage switching, filtration or a detector that separates energy bands. A virtual non-contrast image estimates the appearance after removing iodine signal; do not assume it can replace actual unenhanced images when residual iodine, calcification or a quantitative measurement could change management. Iodine maps depict iodine distribution at the acquisition time, not real-time perfusion alone.
  • A virtual monoenergetic image models attenuation at a chosen photon energy. Low energies can increase iodine conspicuity but may amplify noise and artefacts. Higher energies can reduce the beam-hardening component of metal streaks; they cannot recover photons that never reached the detector. Compare high-energy images with conventional reconstructions. DECT can also distinguish uric acid from calcium-containing stones, a distinction relevant to whether selected uric acid stones can be treated by oral dissolution.

Photon-counting detectors and choosing a vascular protocol

Energy-integrating detectors sum signal deposited by incident photons over a reading interval. Photon-counting detectors register individual events and sort them by energy thresholds. This can provide spectral information and permit smaller detector elements. It does not make spatial resolution or dose independent of the protocol. A dual-source scanner uses two source–detector pairs and can operate them with different spectra for dual-energy imaging, depending on the mode.

CT angiography (CTA) times an iodine bolus to opacify the vessel of interest. Bolus tracking or a test bolus helps select timing. Review thin source sections and appropriate reformations. Perfusion CT instead uses repeated acquisitions across a bolus to estimate tissue haemodynamics from time–attenuation curves. An arterial input curve, the analysis model and coverage influence derived maps. In suspected stroke, motion or a poorly captured bolus can invalidate a perfusion estimate, so compare it with the unenhanced CT, angiography and clinical findings.

See the first clinical photon-counting detector evaluation and the stroke imaging roadmap.

Recognise an artefact before recommending a repeat

AppearanceLikely mechanismWhat to try first
Dark bands between dense structures or a centre that appears abnormally low in a uniform objectBeam hardening: low-energy photons are selectively absorbed along long or dense pathsCheck adjacent reconstructions and use appropriate beam-hardening correction. For prostheses, compare high-energy images or metal-artefact reduction if available.
Border becomes indistinct or a tiny structure seems to have intermediate attenuationPartial volume: the voxel contains more than one materialInspect thin contiguous sections and multiplanar images if the acquisition supports them.
Doubled edges, inconsistent anatomy or streaks between viewsPatient, respiratory or cardiac motion during acquisitionUse the best existing reconstructions first; for planned reacquisition, target the cause with coaching, faster acquisition or cardiac gating as appropriate.
Concentric circles centred on the rotation axisDetector response or calibration errorIf the pattern recurs, have the system checked or recalibrated rather than attributing it to the patient.
Very noisy bright and dark streaks along highly attenuating pathsToo few transmitted photons, often compounded by beam hardening near metalCompare reconstructions, assess obscured structures and report a limitation if the clinical question remains unanswered.
Repeating propeller-like streaks near high-contrast edgesHelical sampling and interpolation (windmill artefact); not synonymous with every cone-beam artefactCompare an alternative reconstruction, including thicker sections if the diagnostic task allows, and optimise longitudinal sampling and reconstruction for future scans.

Review the IAEA diagnostic radiology physics handbook, the DICOM windowing specification, AAPM CT dosimetry and size-specific dose estimates. For the spectral examples, see the EAU urolithiasis guideline and the hip prosthesis study.

Worked example: scanner output index (mGy) with all other settings held fixed
Test Your Knowledge

On unenhanced head CT for sudden right hemiparesis, which listed display is most suitable for closely comparing adjacent grey and white matter at +40 HU and +35 HU?

A

Wide window width (2000 HU) and high window level (+500 HU)

B

Wide window width (1500 HU) and negative window level (-600 HU)

C

Narrow window width (40-80 HU) and window level near +35 to +40 HU

D

Narrow window width (40 HU) and window level at -100 HU

Test Your Knowledge

During a CT pulmonary angiogram with constant tube potential, fixed mA, rotation time, and z-axis scan length, the radiographer increases helical pitch from 0.8 to 1.6 to shorten the breath-hold. How do CTDIvol, DLP, and estimated effective dose change?

A

CTDIvol doubles while DLP and effective dose remain unchanged

B

CTDIvol halves while DLP doubles because of faster table motion

C

CTDIvol remains unchanged while DLP and effective dose halve

D

CTDIvol, DLP, and effective dose all decrease by 50%

Test Your Knowledge

A dual-energy CT study of the urinary tract shows a 7 mm calculus in the proximal ureter. How does dual-energy material decomposition characterise this stone, and which post-processing image helps suppress streak artefact from an adjacent hip prosthesis?

A

Dual-energy material decomposition can distinguish uric acid from calcium-containing stones; high-keV virtual monoenergetic images can reduce metal streaks.

B

Material decomposition measures stone volume via virtual non-contrast; low-keV (40 keV) monoenergetic images eliminate metal streaks

C

Material decomposition differentiates uric acid from calcium using T2* phase; Filtered Back Projection removes metal streaks

D

Material decomposition isolates iodine inside the calculus; narrow soft-tissue windowing removes photon starvation