4.3 Beam Restriction, Filtration & Scatter Control Grids

Key Takeaways

  • Collimation restricts irradiated anatomy and reduces scatter. Filtration preferentially removes lower-energy photons.

  • AERB Table 1B sets filtration by maximum rated kVp for the equipment within its scope. The general-radiography QA form applies 2% of the test distance to each axis sum.

  • Grid ratio is strip height divided by interspace width. The exposure adjustment for a grid depends on total transmission for the protocol.

  • An inverted focused grid or a wrong focus distance reduces peripheral exposure. Tilt across the strips can reduce exposure broadly.

  • An air gap reduces detected scatter and increases object-to-image distance. Account for magnification and exposure when using the technique.

Last updated: November 2026

Collimation limits the size and shape of the primary beam. Filtration changes its energy spectrum. Grids and air gaps reduce scatter reaching the receptor. Each affects the information recorded and the exposure needed for an image.

Beam Restriction

Collimate to the anatomy needed for the projection. This limits irradiation of adjacent tissue and reduces the volume generating scatter. With less scatter reaching the detector, subject contrast is better preserved.

  • A fixed aperture diaphragm uses an opening in a metal plate. A cone or extension cylinder can define a circular field. A distal aperture helps intercept some off-focus rays.
  • A light-beam collimator commonly uses movable, perpendicular sets of shutters and a lamp with a mirror to show the approximate radiation field on the patient. Use the light to check coverage and centre the beam. Permanent or adjustable shutters can also reduce off-focus exposure.
  • Positive beam limitation is an automatic field-sizing feature found on some units. It limits the radiation field for the selected receptor geometry. Confirm coverage and collimate further when the projection requires a smaller field.

The AERB general-radiography QA form compares radiation and optical field edges along each axis. Add the absolute deviations of the two opposing edges separately on each axis. The tolerance is 2% of the focus-to-receptor distance used in the test. At 100 cm, the tolerance is therefore 2 cm per axis. For example, shifts of 0.8 cm and 1.0 cm on opposite edges have a sum of 1.8 cm, or 1.8% at that distance.

Filtration and Beam Hardening

Inherent filtration comes from components in the beam path such as the tube window, oil, housing window and light-localiser mirror, depending on construction. Added aluminium or copper filters remove a greater fraction of lower-energy photons. The transmitted beam has reduced fluence and increased mean energy and half-value layer. For a specified image-receptor exposure, removing photons unlikely to reach the detector can reduce superficial dose. The technique, required contrast and tube loading still matter.

Under AERB/RF-MED/SC-3 (Rev. 2) Appendix I §I.2.1, the minimum total filtration for constant-potential equipment covered by that provision is 2.5 mm Al equivalent. Table 1A lists the associated first HVL values. Table 1B specifies separate minimums for the other equipment in its scope:

Maximum rated tube potentialTable 1B minimum total filtration
Below 70 kVp1.5 mm Al equivalent
70 through 100 kVp2.0 mm Al equivalent
Above 100 kVp2.5 mm Al equivalent

A compensating filter varies thickness across the beam to balance exposure when part thickness varies sharply. A wedge can attenuate more of the beam over the thinner end of an extremity or spine projection. Align the filter to the required anatomy and protocol.

Grids and Scatter Rejection

An anti-scatter grid uses absorbing strips, usually lead, and relatively radiolucent interspaces. The strips preferentially absorb oblique scatter, but also absorb some primary photons. The grid ratio is strip height divided by interspace width, r=h/Dr = h/D. Grid frequency counts strips per unit length. Higher ratios generally improve scatter rejection at the cost of stricter alignment tolerances and greater exposure compensation when matching detector exposure.

The Bucky factor compares the incident exposure needed with and without a grid to deliver the same receptor exposure, for the stated geometry and beam. It is the reciprocal of the grid's total transmission under those conditions. If transmission is 25%, the factor is 1/0.25=41/0.25 = 4. The value depends on scatter, beam quality and grid design. The contrast improvement factor is the ratio of image contrast with the grid to image contrast without it under the specified comparison.

A linear grid allows tube angulation along its strips if the geometry stays within its ratings. A crossed grid restricts angulation along both axes. A focused grid angles the strips to match beam divergence over a marked distance range. A parallel grid has parallel strips and may lose primary exposure toward the margins, particularly with a large field or short distance. A moving Bucky oscillates the grid to reduce visible strip shadows.

Grid errorTypical effect on primary exposure
Focused grid upside downMarked cutoff at both lateral edges, with exposure better preserved near the centre.
Off levelTilting across the strip direction reduces receptor exposure broadly.
Off centreShifting the source across the grid strips away from the focus line reduces primary transmission.
Off focusA source distance outside a focused grid's focal range produces increasing cutoff toward the periphery.

Digital rescaling may hide the change in displayed brightness. Check detector exposure, noise and the pattern of cutoff before deciding whether another image is necessary.

Air Gap

Increasing the patient-to-detector distance lets some of the scattered photons miss the receptor. An air gap can be an option when the protocol supports its magnification, resolution and detector-exposure trade-offs. A longer source-to-image distance can help control magnification. Choose grid, air gap or other scatter control for the patient, projection and detector.

Test Your Knowledge

A general-radiography unit within AERB Rev. 2 Table 1B has a maximum rated tube potential of 90 kVp. What minimum total filtration and maximum per-axis light-field deviation apply under Table 1B and the AERB general-radiography QA form, respectively?

A

1.5 mm Al total filtration, ≤5% of SID

B

2.0 mm Al total filtration, ≤2% of SID

C

2.5 mm Al total filtration, ≤2% of SID

D

2.0 mm Al total filtration, ≤1% of SID

Test Your Knowledge

A grid has lead strips 3.6 mm high separated by 0.3 mm wide aluminum interspaces. What is the grid ratio, and what happens if a focused grid of this type is accidentally placed upside down?

A

10:1, uniform exposure loss across the entire film

B

12:1, cutoff along one lateral edge only

C

12:1, severe bilateral peripheral cutoff leaving only a narrow central band exposed

D

8:1, increased magnification without grid cutoff

Test Your Knowledge

A correctly centred linear focused grid is tilted across the direction of its strips for a mobile chest image. Which change in primary exposure at the detector is typical of this off-level error?

A

Broadly decreased primary exposure across the detector

B

Overexposure along the central axis with dark peripheral bands

C

Sharp cutoff limited strictly to the upper and lower borders

D

Increased contrast with prominent quantum mottle at the center