10.1 Radiation Protection Principles: ALARA, Time, Distance & Shielding

Key Takeaways

  • For patients, justify the examination and optimise the dose for adequate diagnostic images. The occupational and public dose limits do not cap a patient’s own medical exposure.

  • Reducing exposure time lowers dose at a fixed dose rate. The inverse-square law models distance effects for a point source, and the half-value layer describes attenuation.

  • AERB requires protective aprons during operation of mobile and portable X-ray equipment and a minimum 2 m between the operator and mobile equipment.

  • Primary barriers intercept the useful beam. Secondary barriers protect against scattered and leakage radiation. Workload, beam direction, occupancy and distance inform shielding design.

  • AERB’s medical diagnostic X-ray code lists 0.25 mm minimum lead equivalence for protective aprons, gloves, thyroid shields and goggles. Verify protective devices periodically.

Last updated: November 2026

Justification, Optimisation and Dose Limitation

Justification asks whether the expected benefit of an examination outweighs the radiation detriment. Review the clinical question, previous images and alternative tests. For example, images already available from an earlier examination may answer the question without another exposure. A repeat to correct a positioning error is justified only if the original image cannot answer the clinical question and the additional image is needed.

Optimisation means keeping radiation exposure as low as reasonably achievable (ALARA) while meeting the diagnostic objective. In a chest radiograph, position and collimate before exposure, select appropriate technique factors and keep exposure time short enough to limit motion. Deliberately lowering dose until an image becomes nondiagnostic can result in a repeat and greater total dose.

Dose limitation applies to occupational and public exposures. A patient's own medical exposure is managed through justification and optimisation. Diagnostic reference levels (DRLs) help identify protocols that warrant investigation when typical doses are unusually high or low. Use DRLs to compare groups of comparable examinations. A DRL is not a maximum dose for an individual patient. Voluntary carers and comforters are a separate medical-exposure situation for which dose constraints can guide protection.

Time, Distance and Shielding

  • Time. At a constant dose rate D˙\dot D, the dose after time tt is D=D˙tD = \dot D t. Halving exposure time halves dose in that model. In fluoroscopy, use the lowest pulse rate that provides the needed temporal information and review a saved image when it can answer the question without further exposure. Compare settings as well as beam-on time when dose rate changes.
  • Distance. For an unobstructed point source in the same direction, I2=I1(d1/d2)2I_2 = I_1(d_1/d_2)^2. Doubling distance gives one quarter of the initial intensity. If a rate is 4 mGy/h at 1 m in an idealised example, the model predicts 1 mGy/h at 2 m. Real patient scatter is directional, and the patient is an extended source, so use the assumptions provided in exam calculations.
  • Shielding. For a monoenergetic narrow beam passing through a uniform material, I=I0e−μxI = I_0 e^{-\mu x}. A half-value layer (HVL) transmits one half and a tenth-value layer (TVL) transmits one tenth. With constant μ\mu, HVL=ln⁡2/μ\text{HVL} = \ln 2/\mu and TVL=ln⁡10/μ≈3.32 HVL\text{TVL} = \ln 10/\mu \approx 3.32\,\text{HVL}. Three identical HVLs transmit 12.5%. A TVL plus an HVL transmits 5%. A polychromatic beam can harden as it traverses an absorber, so do not assume identical successive HVLs for a clinical beam.

Structural Barriers and Protective Equipment

InputWhat it describes
Primary barrierStructure designed to attenuate the useful beam where it may be directed.
Secondary barrierStructure designed to attenuate patient scatter and tube leakage.
WorkloadExpected use of the equipment. A conventional radiography calculation can express it in mA-min per week.
Use factorFraction of the primary workload directed toward a surface. A conventional secondary-barrier calculation uses unity for scatter and leakage.
Occupancy and distanceHow much time people spend in the adjacent area and how far the occupied point is from the relevant source.

A control booth wall that cannot be struck by the useful beam is a secondary barrier. An adjoining office with high occupancy may need more protection than a less occupied space at the same distance. Actual material and thickness depend on assessment of the installation and applicable requirements.

In India, AERB’s medical diagnostic X-ray code specifies 0.25 mm minimum lead equivalence for protective aprons, gloves, thyroid shields and goggles. It requires periodic verification of shielding adequacy. During interventional imaging, use the equipment’s suspended screen and table curtains where provided. Keep hands out of the direct beam. A glove provides scatter protection and can cause an automatic exposure control system to raise output if introduced into the primary beam.

Mobile Imaging and Pregnancy Precautions

For mobile imaging, the AERB code requires a protective apron and at least 2 m between the operator and the mobile equipment. Work from behind a suitable protective barrier and stay out of the useful beam. When a child or infirm patient needs an accompanying person to assist, provide that person with a protective apron and follow the facility’s instructions.

Ask about possible pregnancy and follow the department’s screening process. Escalate uncertainties to the responsible practitioner before exposure so the examination can be justified and the technique adapted. AERB’s code calls for beam limitation, abdominal shielding and low-dose protocols for radiographs of areas away from the fetus in patients who are pregnant or likely to be pregnant. For children, use a size-appropriate technique and collimate to the area of interest. When shielding is used, avoid obscuring the anatomy or covering automatic exposure control sensors.

Test Your Knowledge

Which principle applies to a patient’s own diagnostic X-ray exposure?

A

Justify the examination and optimise the exposure. Use diagnostic reference levels to review typical doses for comparable examinations.

B

Diagnostic X-rays below 120 kVp produce zero stochastic biological risk in adult patients

C

Patient doses are already included inside the radiographer's annual occupational dose record

D

Lead aprons worn by patients eliminate 100% of absorbed organ dose during radiography

Test Your Knowledge

In an idealised mobile radiography example, treat the patient as a point source of scatter and compare positions in the same direction. The measured air-kerma rate at 1 m is 3.6 mGy/h. What rate does the inverse-square model predict at 3 m?

A

1.2 mGy/h

B

0.9 mGy/h

C

0.4 mGy/h

D

0.1 mGy/h

Test Your Knowledge

In a conventional shielding calculation for a secondary barrier, what use factor (U) is applied to scatter and leakage? For a separate monoenergetic narrow-beam calculation in a uniform absorber, how does TVL relate to HVL?

A

U = 0.25; TVL ≈ 2.0 × HVL

B

U = 1; TVL ≈ 3.32 × HVL

C

U = 0.1; TVL ≈ 10.0 × HVL

D

U = 0.5; TVL ≈ 1.44 × HVL