9.2 Magnetic Resonance Imaging (MRI): Physics, Weighting & Safety Zones
Key Takeaways
Use the proton gyromagnetic ratio in frequency units, about 42.58 MHz/T. The corresponding Larmor frequencies are about 63.87 MHz at 1.5 T and 127.74 MHz at 3.0 T.
In conventional spin echo, short TR and short TE favour T1 weighting. Long TR with long TE favours T2 weighting, and long TR with short TE favours proton-density weighting.
FLAIR nulls free fluid such as CSF. STIR suppresses fat. Increased high-b-value DWI signal with decreased ADC supports restricted diffusion.
In the ACR four-zone model, Zone III is the controlled interface and Zone IV is the scanner room. Verify device conditions and complete a final safety check before Zone IV entry.
Prevent RF burns and protect hearing. If a quench vents into the room, cryogenic gas can cause cold injury and oxygen displacement. Gadolinium-related NSF risk depends strongly on the agent and renal status.
MRI uses a strong static magnetic field, magnetic field gradients and radiofrequency (RF) pulses. The usual signal comes from hydrogen nuclei, abundant in water and fat. The static field creates a small population imbalance between nuclear spin energy states, producing net magnetisation along the field.
Precession and Relaxation
The proton magnetisation precesses at the Larmor frequency. Keep angular frequency and ordinary frequency distinct:
For hydrogen, is approximately . At this gives . At it gives . The transmitting coil applies an RF field near resonance to tip net magnetisation. Transverse magnetisation induces a signal in a receiving coil.
- T1 describes recovery toward longitudinal equilibrium. Following a 90° excitation from equilibrium, approximately 63% of equilibrium longitudinal magnetisation has recovered at one T1.
- T2 describes loss of coherent transverse magnetisation from spin interactions. After one T2, about 37% of the initial transverse magnetisation remains in the ideal exponential model.
- T2-star adds dephasing from static field differences and susceptibility to the irreversible T2 processes. A gradient echo can reveal these additional differences.
Weighting, Inversion and Diffusion
Repetition time (TR) is the interval between successive excitations of the same slice or volume in the conventional sequence. Echo time (TE) is the interval between excitation and the echo. Other sequence designs can use additional controls, including flip angle and inversion time (TI).
| Sequence or weighting | Typical conventional approach | What to recognise |
|---|---|---|
| T1-weighted spin echo | Short TR, short TE | Fat often appears bright and CSF dark. Useful for anatomy and T1 enhancement |
| T2-weighted spin echo | Long TR, long TE | Fluid and many oedematous regions appear bright |
| Proton-density weighting | Long TR, short TE | Reduces the relative contributions of T1 and T2 to emphasize proton density |
| Fluid-attenuated inversion recovery (FLAIR) | Inversion time chosen to null CSF, often combined with T2 weighting | CSF is dark. Abnormal high signal near the ventricles is easier to see in many settings |
| Short-TI inversion recovery (STIR) | Short inversion time chosen to null fat | Suppressed fat makes fluid-sensitive findings such as marrow oedema more conspicuous |
| Diffusion-weighted imaging (DWI) and ADC | Compare high-b-value signal with the apparent diffusion coefficient map | High DWI signal with low ADC supports reduced apparent diffusion. Use the ADC map to distinguish it from T2 shine-through |
In a conventional spin echo, a refocusing RF pulse reverses dephasing from stationary field offsets. In a gradient echo, reversing a gradient forms the echo without that RF refocusing step. Gradient-echo contrast depends on TR, TE, flip angle and the sequence design. With suitable settings it can reveal T2-star effects such as susceptibility from blood products.
Gradients encode spatial position. The sampled signal is arranged in k-space, and an inverse Fourier transform can reconstruct the image. The centre contains low spatial frequencies important for broad intensity and contrast variations. Outer samples represent finer spatial variations. Acquiring higher spatial frequencies supports finer detail when the rest of the acquisition allows.
ACR Zones and Device Screening
In the ACR zoning model, Zone I is public space. Zone II is a supervised transition where staff receive and screen patients. Zone III is a controlled area supervised by appropriately trained MR personnel, often including the operator console. Zone IV is the scanner room and includes the projectile hazard area. A static fringe field can extend outside the scanner room. The 2026 ACR manual references a 9-gauss (0.9 mT) contour in its definition of the MR Environment. Zone III access is controlled by the facility's layout and safety policy, rather than a single room-wide field value.
Identify implants and equipment by exact make, model and configuration. MR Safe indicates no known hazards from exposure to an MR environment. MR Conditional indicates safety within the conditions in the labelling. Conditions can include static field, spatial field gradient, RF exposure, scan duration, positioning and device settings. MR Unsafe indicates an unacceptable risk in the MR environment. An unverified device requires the facility's formal assessment before entry or scanning. The MR technologist completes a final check before the patient enters Zone IV.
RF, Noise, Quench and Gadolinium
Specific absorption rate (SAR) estimates RF power absorbed per unit mass in watts per kilogram. Local burns can still occur, even when system estimates meet operating limits. Use the specified padding between skin surfaces, keep conductive cables arranged under their instructions, and check implants, patches and clothing before scanning. Gradient switching generates acoustic noise. Provide and check suitable hearing protection under the local procedure.
A quench is loss of superconductivity in a superconducting magnet. At a system with a helium vent, a venting failure can release cold helium gas into the scanner room. Cold injury and oxygen displacement are immediate hazards. Follow the site's emergency plan and clear the room as quickly as safely feasible. Treat the magnetic field as present until qualified personnel verify otherwise. Routine scanner shutdown is not equivalent to removing the static field.
Gadolinium-based contrast can shorten T1 and increase signal on common postcontrast T1-weighted images. Nephrogenic systemic fibrosis (NSF) risk is concentrated in severe kidney dysfunction with higher-risk agents. Under the ACR–NKF approach, a Group II agent has very low NSF risk at a standard dose, including in advanced kidney disease. ACR group refers to the evidence for NSF risk, rather than a simple division into linear and macrocyclic chemistry. Check agent, indication, renal status as required and the department policy with the supervising clinician.
A patient with suspected acute cerebral ischaemia undergoes MRI. Which finding supports true restricted diffusion in the affected region?
T1-weighted bright and T2-weighted dark
STIR dark and Proton Density bright
FLAIR dark with bright cortical CSF
DWI hyperintense (bright) paired with ADC map hypointense (dark)
Hydrogen has a gyromagnetic ratio in ordinary frequency units of about 42.58 MHz/T. At 3.0 T, what is its Larmor frequency and which part of k-space encodes low spatial frequencies?
127.74 MHz; the centre of k-space
63.87 MHz; the centre of k-space
127.74 MHz; the outer samples of k-space
21.29 MHz; the outer samples of k-space
A helium-cooled superconducting MRI magnet quenches and a vent failure allows cold helium gas into Zone IV. What are the immediate hazards from the gas?
Acute ionizing radiation exposure and high-voltage X-ray tube arcing
Sudden conversion of MR-Safe plastics into ferromagnetic projectiles
Oxygen displacement and cold injury from cryogenic gas
Instantaneous heating of gadolinium contrast inside peripheral veins
Sections you finish are checked off in the contents.