10.3 Diagnostic X-Ray Quality Assurance Tests & AERB Tolerances

Key Takeaways

  • AERB calls for acceptance QA after installation and periodic QA at least every two years and after repairs with radiation safety implications.

  • The AERB general-radiography report checks the sum of absolute field-edge deviations in each axis against 2% of the measurement distance. Central beam tilt must be less than 1.5°.

  • The AERB report lists a ±5 kV operating potential tolerance and a 10% timer-error criterion for this equipment.

  • Output linearity requires Coefficient of Linearity CoL < 0.1, and output consistency requires Coefficient of Variation CoV < 0.05

  • Under the Rev. 2 safety code, constant-potential X-ray equipment outside mammography needs at least 2.5 mm Al-equivalent total filtration. Table 1B specifies separate tiers for other equipment in its scope.

Last updated: November 2026

A well-maintained X-ray unit produces consistent diagnostic contrast at the lowest patient dose. AERB requires acceptance quality assurance (QA) after installation and periodic QA at least every two years and after repairs that affect radiation safety. The examples below use its diagnostic X-ray QA form for general radiography and the design requirements of AERB/RF-MED/SC-3 (Rev. 2). Use the modality-specific form for other equipment, such as mammography.

Master Table of AERB Diagnostic X-Ray QA Tolerances

QA ParameterTest Tool / MethodAERB Tolerance Limit
Optical & Radiation Field CongruenceCoins/paperclips or radiopaque template at 100 cm100\text{ cm} SIDSum of absolute edge deviations along either axis ≤2%\le 2\% of SID (≤2 cm\le 2\text{ cm} at 100 cm100\text{ cm})
Central Ray PerpendicularityTwo-ball beam alignment cylinder at 100 cm100\text{ cm} measurement distanceTilt <1.5∘<1.5^\circ
Operating Potential (kVp\text{kVp}) AccuracyNon-invasive digital kVp meter±5 kV\pm 5\text{ kV}
Irradiation Time AccuracyElectronic timer / solid-state dosimeterCheck absolute percent error against 10%10\%. The form’s summary specifies <10%<10\%.
Output Linearity (mA\text{mA} / Timer)Calibrated dosimeter across mA stationsCoL<0.1\text{CoL} < 0.1
Output Consistency (Reproducibility)Five repeated exposures at fixed settings in the AERB QA formCoV<0.05\text{CoV} < 0.05
Beam qualityFirst half-value layer measured with aluminium attenuatorsCompare with the code table for the generator type and modality.
Tube Housing LeakageSurvey at 1 m1\text{ m} from the target under the code test conditionsFor the general category in Rev. 2 Appendix I.1.1, no more than 1 mGy in one hour.
Focal Spot SizeCompare measured dimensions to nominal size ffThe form allows +0.5f+0.5f for f<0.8 mmf<0.8\text{ mm}, +0.4f+0.4f for 0.8≤f≤1.5 mm0.8\le f\le1.5\text{ mm} and +0.3f+0.3f above that.

Geometric & Beam-Quality Tests

Congruence of Optical and Radiation Field

A skewed collimator mirror or displaced lamp bulb causes the light field to show one area while the X-ray beam exposes another—clipping anatomy or irradiating untargeted tissue. At 100 cm100\text{ cm} Source-to-Image Distance (SID), place four radiopaque markers (coins, paperclips, or a calibrated template) at the corners/edges of a 20×20 cm20 \times 20\text{ cm} light field and expose. Measure the mismatch between the light-field border and exposed X-ray border on opposite sides (∣ΔX1∣+∣ΔX2∣\lvert \Delta X_1 \rvert + \lvert \Delta X_2 \rvert and ∣ΔY1∣+∣ΔY2∣\lvert \Delta Y_1 \rvert + \lvert \Delta Y_2 \rvert). Under AERB rules, the sum of absolute edge deviations along either axis must not exceed 2%2\% of SID (≤2 cm\le 2\text{ cm} at 100 cm100\text{ cm} SID).

Central Ray Perpendicularity (Beam Alignment)

A beam alignment test cylinder contains two vertically separated steel spheres inside a radiolucent Lucite cylinder placed at the center of the light field. If the central ray is perpendicular to the image receptor, the two steel spheres superimpose cleanly; if tilted, the upper sphere shifts laterally. The AERB general-radiography form specifies a central beam tilt less than 1.5∘1.5^\circ.

Total Filtration & Half-Value Layer (HVL)

Inherent filtration (glass envelope, insulating oil, tube port) plus added aluminum sheets strips useless low-energy photons that would otherwise be absorbed superficially by the patient's skin. The first half-value layer (HVL) is the aluminium thickness that halves air kerma in the measurement setup. An HVL value and a total filtration specification describe different measurements. In Rev. 2 Appendix I.2.1, constant-potential equipment in scope needs at least 2.5 mm Al equivalent total filtration and meets the voltage-dependent first-HVL requirements in Table 1A. At 80 kV, the table gives 2.9 mm Al minimum first HVL.

For other equipment in the scope of Appendix I.2.1, Table 1B sets minimum total filtration of 1.5 mm Al below 70 kVp, 2.0 mm Al from 70 to 100 kVp inclusive, and 2.5 mm Al above 100 kVp. Use the maximum rated tube potential. Note that the separate general QA form lists its lowest tier up to and including 70 kV. An exam question about exactly 70 kV should identify which source applies.


Generator Performance: kVp, Timer, Linearity & Consistency

Operating Potential (kVp\text{kVp}) and Timer Accuracy

Measured non-invasively with a solid-state multifunction X-ray multimeter, kVp accuracy must stay within ±5 kV\pm 5\text{ kV} in the AERB QA form. For a setting of 80 kV and reading of 84 kV, the 4 kV difference meets that criterion. For a timer set to 100 ms with a measured 106 ms, the absolute percent error is 6%, below 10%. The form’s timer test page writes the tolerance as ±10%\pm 10\% and its summary as <10%<10\%; classify readings at the exact boundary under the specified test criterion.

Linearity of Radiation Output (CoL\text{CoL})

At fixed tube voltage and measurement geometry, compare the output per mAs\text{mAs} across loading stations. For 20 mAs, 50 mA for 0.40 s and 200 mA for 0.10 s give the same nominal mAs. Each output is expressed per mAs\text{mAs} (Xi=Air Kermai/mAsiX_i = \text{Air Kerma}_i / \text{mAs}_i). Linearity is quantified by the Coefficient of Linearity (CoL\text{CoL}):

CoL=Xmax⁡−Xmin⁡Xmax⁡+Xmin⁡<0.1\text{CoL} = \frac{X_{\max} - X_{\min}}{X_{\max} + X_{\min}} < 0.1

Worked Example: At 80 kVp80\text{ kVp} and 100 cm100\text{ cm} SID, a radiographer measures X1=0.052 mGy/mAsX_1 = 0.052\text{ mGy/mAs} at 50 mA50\text{ mA} (20 mAs20\text{ mAs}) and X2=0.058 mGy/mAsX_2 = 0.058\text{ mGy/mAs} at 200 mA200\text{ mA} (20 mAs20\text{ mAs}). Calculating CoL\text{CoL}:

CoL=0.058−0.0520.058+0.052=0.0060.110≈0.055\text{CoL} = \frac{0.058 - 0.052}{0.058 + 0.052} = \frac{0.006}{0.110} \approx 0.055

Because 0.055<0.100.055 < 0.10, the unit passes AERB linearity. If Xmax⁡X_{\max} were 0.072 mGy/mAs0.072\text{ mGy/mAs} instead, CoL=0.020/0.124≈0.161≥0.10\text{CoL} = 0.020 / 0.124 \approx 0.161 \ge 0.10, which fails the form’s linearity criterion and warrants investigation.

Consistency / Reproducibility of Output (CoV\text{CoV})

Repeating five consecutive exposures at identical kVp\text{kVp}, mAs\text{mAs}, and SID tests generator repeatability using the Coefficient of Variation (CoV\text{CoV}):

CoV=σXˉ<0.05\text{CoV} = \frac{\sigma}{\bar{X}} < 0.05

where σ\sigma is the standard deviation of the repeated readings and Xˉ\bar{X} is the mean output. Any CoV≥0.05\text{CoV} \ge 0.05 (5%5\%) fails the stated consistency check and should be escalated for evaluation.


Focal Spot, Tube Leakage & Grid Alignment Checks

  • Effective Focal Spot Size — Measured during acceptance and tube changes using a star test pattern (calculates effective focal spot size from the diameter of the unsharp blur zone), a slit camera, or a pinhole camera (direct magnified image). Focal spot blooming at high mA degrades spatial resolution.
  • Tube Housing Leakage — Rev. 2 Appendix I.1.1 limits leakage air kerma to 1 mGy in one hour at 1 m from the target under its specified maximum rated potential and current conditions. The code sets separate measurement conditions and limits for mammography and intraoral dental equipment. Use the applicable procedure and calibrated survey instruments.
  • Grid Alignment & Table/Bucky Checks — Center the X-ray tube transversely to the Bucky grid within its specified focal range, verify smooth reciprocating grid motion (absence of grid lines), check lateral decentering on homogeneous phantom exposures, and check the applicable tabletop attenuation requirement.
Test Your Knowledge

At 80 kVp and 100 cm SID, a diagnostic X-ray unit produces 0.045 mGy/mAs at the 100 mA station and 0.055 mGy/mAs at the 200 mA station. What is the Coefficient of Linearity (CoL), and does it pass the AERB diagnostic X-ray QA form’s criterion?

A

CoL = 0.20; Fails

B

CoL = 0.05; Passes

C

CoL = 0.10; Fails (must be strictly < 0.1)

D

CoL = 0.01; Passes

Test Your Knowledge

A general-radiography unit has a maximum rated tube potential of 120 kVp. Under the AERB Rev. 2 design code, what are the minimum total filtration and maximum tube housing leakage at 1 m under the specified test conditions?

A

1.5 mm Al filtration; ≤ 5 mGy/h leakage

B

2.0 mm Al filtration; ≤ 2 mGy/h leakage

C

2.5 mm Al filtration; ≤ 0.1 mGy/h leakage

D

2.5 mm Al filtration; ≤ 1 mGy/h leakage

Test Your Knowledge

During a collimator congruence QA test performed at 100 cm SID, what is the maximum permissible sum of absolute deviations between the optical light field and the X-ray radiation field along either axis?

A

2 cm (2% of SID)

B

5 cm (5% of SID)

C

1 cm (1% of SID)

D

0.5 cm (0.5% of SID)