3.3 X-Ray Interactions with Matter: Photoelectric, Compton & Attenuation
Key Takeaways
Photoelectric absorption removes a photon. Iodine and barium have K absorption edges near 33.2 and 37.4 keV.
Compton scattering releases a recoil electron and redirects a lower-energy photon. It contributes to scatter at the receptor and around the patient.
Coherent scattering changes photon direction with negligible energy change. Nuclear pair production and photodisintegration require higher energies than diagnostic X-ray beams provide.
For a narrow monoenergetic beam in a uniform material, I = I₀ exp(−μx) and HVL = ln(2)/μ.
The equal-HVL rule assumes constant μ. In a polyenergetic beam the second HVL is generally thicker than the first because of beam hardening.
As a photon crosses matter it can be absorbed, scattered or transmitted without an interaction. In diagnostic projection imaging, the photoelectric effect and Compton scattering explain much of how subject contrast forms, why scatter reaches the receptor and how energy is transferred in the patient.
Photon Interactions with Matter
| Interaction | What the photon interacts with | Outcome |
|---|---|---|
| Coherent (Rayleigh) scattering | Bound electron cloud | Direction changes with negligible energy change and no ionisation in that event. Its contribution is more relevant toward the lower end of diagnostic photon energies. |
| Photoelectric absorption | Bound electron | The photon is absorbed and a photoelectron is emitted if there is enough energy to free it. |
| Compton scattering | Electron, approximated as free and initially at rest for the usual equations | A recoil electron and a scattered photon share the incident energy. |
| Pair production | Electric field of a nucleus | A photon above the approximately 1.022 MeV threshold can create an electron and a positron. |
| Photodisintegration | Nucleus | An energetic photon can eject a nucleon. The threshold depends on the nucleus. |
Photoelectric Absorption
A bound electron ejected by photoelectric absorption has kinetic energy approximately equal to the photon energy minus its binding energy, allowing for recoil. An outer electron can fill the vacancy and emit a characteristic X-ray or transfer the energy to another electron (an Auger electron). A tightly bound shell can contribute only when the photon has enough energy to eject one of its electrons.
Between absorption edges, photoelectric interaction probability rises strongly with atomic number and falls rapidly with photon energy. The common rule describes a trend and is only approximate. At the K edge, K-shell absorption becomes possible and attenuation jumps. Iodine's K edge is about 33.2 keV, and barium's is about 37.4 keV. The difference between materials contributes to projection contrast in bone and in areas containing iodine or barium. Absorption also transfers energy to matter. The resulting patient dose depends on the full spectrum, technique and anatomy.
Compton Scattering
An incident photon scatters from an electron and transfers part of its energy to that electron. The scattered photon generally has lower energy and a new direction. For a free electron at rest, the scattered wavelength increases by:
The largest wavelength change and energy loss for a fixed incident energy occur at 180°. As the incoming energy grows large, the backscattered energy approaches 255.5 keV in the free-electron model. For example, a 100 keV photon backscattered by that model gives .
Compton attenuation per unit volume depends on electron density and photon energy. Per unit mass, electron density is similar for many biological materials, with variation from composition. Thus the Compton dependence on elemental atomic number is much weaker than in photoelectric absorption. Its relative contribution in soft tissue typically grows as photon energy increases across the diagnostic range. Compton scatter reaching the receptor reduces contrast, and the patient is a principal source of staff scatter in fluoroscopy.
Higher-Energy Interactions
Pair production near a nucleus creates the electron and positron. The threshold is set by twice the electron rest-mass energy, with a small recoil correction. After slowing down, a positron can annihilate and yield two approximately opposed 511 keV photons. In PET, the positron comes from radionuclide decay. Photodisintegration ejects a nucleon or other nuclear fragment at energies above the reaction-specific threshold. Both pair production and photodisintegration require energies above those used in diagnostic X-ray beams.
Exponential Attenuation and Half-Value Layer
For the primary component of a narrow monoenergetic beam in a uniform material:
Here is the total linear attenuation coefficient in inverse length, formed by adding the contributions from the relevant interactions. The mass attenuation coefficient in expresses attenuation per mass thickness at that energy and composition. Multiplying by material density gives for a specified density.
In diagnostic beam quality testing, the half-value layer (HVL) is measured in a specified absorber, commonly aluminium, as the thickness that halves air kerma or air-kerma rate in narrow-beam geometry. For a monoenergetic beam with a constant , each additional equal thickness halves the remaining primary intensity:
For example, in a monoenergetic approximation with a constant 3.0 mm Al HVL and initial air kerma of 2.4 mGy, , or . After 9.0 mm Al, three halvings give , which is 12.5% transmission. In an actual polyenergetic diagnostic beam, adding absorber preferentially removes lower energies. As the beam hardens, the second HVL is generally thicker than the first, and multiple copies of the measured first HVL usually give a different result from the monoenergetic calculation.
What explains the marked increase in the attenuation coefficient of iodine at about 33.2 keV?
Coherent scattering starts at that energy
Compton scattering suddenly scales as the cube of atomic number
Photons reach the binding threshold for iodine K-shell photoelectric absorption
Nuclear pair production starts in iodine
Which interaction commonly produces both scatter at the detector and staff exposure from radiation scattered by the patient during fluoroscopy?
Coherent (Rayleigh) scattering
Compton scattering
Photoelectric absorption
Photodisintegration
For a narrow monoenergetic beam with a constant aluminium HVL of 2.5 mm, what thickness reduces primary-beam intensity to 6.25%?
5.0 mm Al
6.25 mm Al
7.5 mm Al
10.0 mm Al