7.3 Trauma, Bedside Mobile, Pediatric & Geriatric Positioning
Key Takeaways
Work within the trauma team’s movement precautions. Move the tube and receptor around an injured limb if the requested projections can be obtained safely.
A horizontal-beam lateral demonstrates a knee fat–fluid level or profiles a trauma hip without a frog-leg position. Cervical radiographs, where indicated, should show the cervicothoracic junction.
For Indian mobile radiography, AERB calls for a barrier, protective apron and at least 2 m between the mobile equipment and the operator.
For children, support positioning, measure part thickness, collimate to the anatomy and use an exposure time short enough for the clinical task.
For frail adults, adapt the beam and receptor to a safe position, assist transfers and select exposure based on part thickness and the validated protocol.
Radiography in Acute Trauma
Before positioning, confirm the requested views and the trauma team's movement restrictions. Maintain cervical motion restriction when required. A collar or support should be adjusted or removed only with the treating team's direction. A longboard used for extrication is not a routine transport or prolonged immobilisation surface. Coordinate any transfer with the team so airway, circulation, lines and spine precautions are maintained.
Where two projections are required to assess a fracture, seek substantially different views without forcing an injured limb into a routine position. Support the detector and change the tube direction as needed. If pain, instability or other injuries prevent the planned view, document the limitation and ask the responsible clinician about the next view or modality.
| Requested projection | How to adapt the exam |
|---|---|
| Lateral cervical radiograph, when plain radiography is selected | Use a horizontal beam when the patient must remain supine. Check visibility throughout the cervical spine and through C7–T1. If the junction is obscured, consult the trauma team before repositioning or requesting another view |
| Horizontal-beam lateral knee | Keep the injured leg supported. A horizontal beam can display a suprapatellar fat–fluid level. In acute trauma, fat above blood in the joint supports an intra-articular fracture even if its line is hard to see |
| Cross-table lateral trauma hip | Aim the horizontal beam to show the femoral neck and proximal femur. In a Danelius–Miller-type setup, raise the uninjured leg clear of the beam if clinically safe and leave the injured limb supported. Seek an approved alternative if the other leg cannot be moved |
Current ACR and NICE pathways favour noncontrast CT as the initial study for adults with suspected blunt cervical injury when imaging is indicated by the clinical criteria. The cross-table X-ray technique describes how to obtain a film if one is requested. It does not clear the spine by itself. Avoid unsupervised arm traction or neck movement to improve an image in suspected instability. A paediatric pathway uses age, examination and injury factors to choose the imaging.
Bedside Mobile Radiography
Bedside imaging helps when the patient cannot safely move to the department. Before the exposure, confirm the patient, clinical purpose and requested view. If clinically permitted, position the patient as erect as possible for a mobile chest radiograph. Support the receptor and avoid moving lines or devices. Record the projection and position so the reader can interpret geometric differences from a standard erect PA chest image.
Older capacitor-discharge generators can lose tube voltage during an exposure as charge is depleted. A modern high-frequency mobile unit regulates the voltage electronically. Select a technique chart for the unit and detector. Measure source-to-image distance and use a technique appropriate to the chosen geometry. An AP chest examination can magnify the heart if the heart is farther from the receptor than in a PA projection. With a focused grid, use the specified focal distance, centre the beam and align it in the direction the grid permits. A tilted grid or off-centre beam can cause grid cutoff, so inspect positioning before increasing exposure.
Indian Operator Protection
AERB Safety Code AERB/RF-MED/SC-3 (Rev. 2) §7.2 calls for working behind a protective barrier and wearing an apron while operating mobile or portable equipment. For a mobile unit it specifies a minimum distance of 2 m between the equipment and the operator. Appendix II specifies at least 0.25 mm lead equivalence for protective aprons. Remain outside the useful beam and wear the monitoring device as directed by the radiological safety officer.
Tell people nearby before exposing, clear the working area of unnecessary staff and visitors, and position the barrier according to the ward's radiation-safety procedure. The operator distance rule is measured from the mobile equipment. It does not substitute for the ward's arrangements to protect other people. Limit the field to the requested anatomy.
Children and Frail Adults
Explain the exam in age-appropriate language and rehearse a simple breath or stillness instruction where possible. Choose a safe support or immobilisation aid for the child's size and the required projection. Stabilise gently, and keep the airway and circulation observable. Avoid routine manual holding. If an escort has to assist under the facility's protocol, AERB calls for a protective apron and keeping people outside the primary beam.
Select technique from the measured thickness and task, not age alone. Use a short exposure time with a suitable tube current and focal spot, while delivering the intended mAs. Tight collimation reduces irradiated tissue and scatter. Image Gently advises avoiding a grid below about 10–12 cm thickness for paediatric digital radiography. Apply the validated local technique chart for the patient and field. In ACR digital radiography guidance, routine patient gonadal shielding is no longer recommended. Follow applicable local requirements, protect the intended anatomy and keep shielding away from active AEC sensors.
For a frail patient who cannot lie flat or hold a standard posture, use radiolucent support and adapt the detector and beam. Secure brakes and ask for assistance before a transfer. Agree on a short instruction the patient can hear or understand. For tremor or breathlessness, shortening exposure time can help when the equipment and technique permit it. Select factors according to the part, thickness and protocol. Osteoporosis alone does not justify a fixed percentage reduction in kVp.
The trauma team has selected plain cervical and knee radiography in a patient who must remain supine under spinal precautions. Which acquisition strategy fits the request?
Remove the cervical collar for an AP open-mouth view and obtain a standing lateral knee
Log-roll the patient onto their side for a standard lateral C-spine and flexed lateral knee
Maintain movement precautions, use horizontal-beam lateral views, check C7–T1 and look for a knee fat–fluid level
Perform immediate flexion-extension lateral C-spine views to test ligamentous integrity
During mobile bedside radiography in a hospital ward, which combination of distance and protection follows the AERB diagnostic X-ray safety code?
Stand 1 m from the tube head in line with the primary beam wearing a 0.15 mm Pb apron
Stay at least 2 m from the mobile equipment, outside the useful beam, wearing a protective apron of at least 0.25 mm lead equivalence and using a protective barrier
Stand 1.5 m behind the detector cassette without an apron because the cassette stops all scatter
Stand at least 2 m directly behind the patient in the path of the exit beam wearing a 0.25 mm Pb apron
When performing an acute chest radiograph on an uncooperative toddler, which technical and positioning strategy minimizes motion unsharpness while limiting patient dose?
Use low mA with a 200 ms exposure time and open collimation to avoid missing lung apices
Have a radiographer hold the child with bare hands inside the primary field for speed
Use age-appropriate support, a validated short-exposure technique and collimation to the required anatomy
Increase kVp by 20% and place a contact shield over the middle mediastinum