9.1 Computed Tomography (CT): Physics, Helical Scanning & Hounsfield Units

Key Takeaways

  • Modern slip-ring CT can rotate the source and detector array continuously while the table moves. Multiple detector rows acquire data across a wider section of the patient in each rotation.

  • Helical pitch is table travel per rotation divided by total nominal collimation. A higher pitch lowers CTDIvol if tube output per rotation and other relevant factors remain fixed.

  • CT numbers compare attenuation to water. Water is 0 HU and air is approximately −1,000 HU. The display window changes the visible range without changing the CT numbers.

  • CTDIvol describes scanner output relative to a standard phantom. Dose-length product multiplies it by scan length, so scan coverage and additional phases affect the total.

  • A detector calibration fault can produce rings in a rotate–rotate scanner. Beam hardening, partial volume, motion and metal have different appearances and remedies.

Last updated: November 2026

CT measures X-ray transmission from many angles. Reconstruction estimates the spatial distribution of attenuation and displays cross-sectional anatomy. The acquisition geometry governs how the scanner samples the patient.

CT geometryHow the source and detectors move
First generationA narrow pencil beam and a detector translate across the subject and rotate between traverses
Second generationA narrow fan and multiple detectors still translate and rotate, acquiring more data per traverse
Third generationA rotating source and detector array stay opposite each other. This rotate–rotate arrangement underlies most modern CT systems
Fourth generationThe X-ray source rotates inside a stationary detector ring

Slip rings allow continuous gantry rotation. In helical CT, the table translates during rotation. In sequential axial CT, the table can move between acquisitions. A multidetector array samples multiple channels along the patient axis. The number of acquired channels and the number and thickness of reconstructed images describe different parts of the process.

A CT tube is designed for the sustained heat load of repeated rotations. A bow-tie filter shapes the beam across the field and attenuates peripheral rays more strongly than central rays. Choose the filter and scan field matched to the patient and protocol. Solid-state detector channels can convert transmitted X-rays to light in a scintillator and then to electrical signal in photodiodes.

Pitch and Reconstruction

Helical pitch is table travel in one full rotation divided by total nominal beam collimation. For 64 channels each nominally 0.625 mm wide, total collimation is 40 mm. A table advance of 60 mm per rotation gives pitch 60/40=1.560/40=1.5.

At fixed collimation and rotation time, a larger pitch covers more length per unit time. At fixed tube output per rotation, the phantom output index per unit length decreases as pitch rises. Some protocols maintain effective mAs or use automatic tube current modulation, so raising pitch alone does not establish a decrease in actual patient dose. Helical interpolation and the reconstruction method also affect section profile and artefacts.

Reconstructed voxels have in-plane dimensions set by the reconstruction field and image matrix and a through-plane dimension related to slice thickness. Thin, closely spaced sections support high-quality reformations. Isotropic means approximately equal dimensions in all three axes. Multiplanar reformations (MPR) show coronal, sagittal or oblique planes. A maximum intensity projection (MIP) selects high attenuation along a line through the volume and can highlight enhanced vessels. Volume rendering uses opacity and colour mappings to display a selected anatomy in three dimensions.

CT Numbers and Display Windows

A CT number expresses attenuation relative to water. In the common approximation where air attenuation is negligible compared with water:

HU≈1000μtissue−μwaterμwater\mathrm{HU} \approx 1000\frac{\mu_{\mathrm{tissue}}-\mu_{\mathrm{water}}}{\mu_{\mathrm{water}}}

Water has a reference value of 0 HU and air is near −1,000 HU. Fat is usually negative and cortical bone is strongly positive. The value of a tissue varies with composition, beam energy, contrast enhancement and partial volume. Acute blood on a noncontrast brain CT can be hyperattenuating, and location, morphology and clinical history help distinguish it from calcification and other causes.

Window centre selects the middle of the displayed CT-number range. Window width controls the span mapped across the greyscale. With a linear display and conventional polarity, values below the window appear dark and values above it appear bright. The nominal limits are approximately centre plus or minus half of width. A narrow window distributes more display levels across a small attenuation interval.

Example displayWidth / centre (HU)Task
Brain80 / 40Compare small attenuation differences in brain tissue
Lung1,500 / −600View aerated lungs and their structures across a wider range
BoneWide window with a higher centre than brainInspect cortical margins and bony detail

These display examples leave acquisition unchanged. A brain window with width 80 and centre 40 spans approximately 0–80 HU.

Protocol and Scanner Output

Define the clinical question before selecting coverage, acquisition phase and reconstruction. In a pathway for suspected acute stroke, a noncontrast head CT helps detect haemorrhage and assess early changes. CT angiography, if indicated, uses a timed intravenous bolus to show arterial anatomy. Bolus tracking can trigger the acquisition once enhancement reaches the configured threshold in the selected vessel. Protocol timing depends on the anatomy and scanner. Add further phases only when they answer the question.

Volume CT dose index (CTDIvol) in mGy is a scanner output index based on a standard phantom. The reference phantom is commonly 16 cm or 32 cm in diameter. In helical CT, CTDIvol=CTDIw/pitch\mathrm{CTDI}_{\mathrm{vol}}=\mathrm{CTDI}_{\mathrm{w}}/\text{pitch}. For one acquisition, the dose-length product (DLP) is CTDIvol multiplied by scan length. For example, 8 mGy over 30 cm gives 240  mGy⋅cm240\;\mathrm{mGy\cdot cm}. The sum across acquisitions accounts for additional scan phases. CTDIvol and DLP describe protocol output and coverage. A size-specific dose estimate accounts for patient size under a defined method and still differs from an individual organ dose.

Optimise the planned scan length, number of phases, tube voltage and current to the diagnostic task and patient. Automatic tube current modulation adapts current to attenuation, and iterative reconstruction can reduce noise under an appropriately validated protocol.

CT Artefacts

AppearanceLikely cause or action
Cupping or dark streaks between dense structuresLower-energy photons are preferentially removed as the beam passes through matter. Filtration and beam-hardening correction can reduce the effect
Concentric ring centred on rotation axisA rotating detector channel that repeatedly reports the wrong signal. Check calibration and compare with a uniform phantom
Mixed value at a boundary, such as bone and fluid in one thick sectionPartial volume averaging. A thinner reconstruction can help when the source data permit
Blurring or repeated edgesMotion. Use appropriate positioning, breath-hold coaching or cardiac synchronisation for the protocol
Bright and dark streaks near metalPhoton starvation and beam hardening can contribute. Check the available metal-artefact reduction reconstruction and its effect on nearby anatomy
Test Your Knowledge

In a helical CT acquisition, the total nominal beam collimation is 40 mm and the table advances 60 mm per rotation. What is pitch, and what happens to the phantom CTDIvol compared with pitch 1.0 if CTDIw stays fixed?

A

Pitch is 1.5 and CTDIvol is two-thirds as large

B

Pitch is 0.67 and CTDIvol is two-thirds as large

C

Pitch is 1.5 and CTDIvol is 1.5 times as large

D

Pitch is 20 and CTDIvol is unchanged

Test Your Knowledge

A noncontrast head CT shows a new, focally hyperattenuating parenchymal collection at +68 HU. Which option is consistent with the finding and gives a suitable display for assessing the surrounding brain?

A

Cerebrospinal fluid; bone window (WW 2500 / WL 480)

B

Lipoma (adipose tissue); lung window (WW 1500 / WL -600)

C

Acute hemorrhage; brain window (WW 80 / WL 40)

D

Cortical bone; wide soft-tissue window (WW 1000 / WL 200)

Test Your Knowledge

A radiographer observes a sharp concentric circular band centered at the image isocenter on all axial slices from a 64-slice third-generation CT scanner. What is the most likely cause?

A

Beam hardening across the petrous temporal bones

B

A defective or miscalibrated single detector element in the rotating detector array

C

Partial volume averaging from selecting 0.625 mm slice thickness

D

Patient respiratory motion during a single breath-hold