10.2 AERB Regulatory Framework, Dose Limits & BARC TLD Dosimetry

Key Takeaways

  • Indian diagnostic X-ray facilities apply the Atomic Energy (Radiation Protection) Rules, 2004 and AERB medical diagnostic X-ray Safety Code AERB/RF-MED/SC-3 (Rev. 2).

  • The facility obtains a licence to operate medical diagnostic X-ray equipment through eLORA. AERB guidance calls for periodic quality assurance at least once every two years and after repairs that affect radiation safety.

  • AERB limits occupational effective dose to 20 mSv per year averaged over five consecutive calendar years and 30 mSv in any one calendar year. The eye-lens equivalent dose limit is 150 mSv in a year.

  • Once pregnancy is declared, AERB limits the equivalent dose to the embryo or fetus to 1 mSv for the remainder of the pregnancy.

  • The BARC thermoluminescent badge uses dysprosium-doped calcium sulfate. AERB advises wearing a single badge at chest level beneath a lead apron when one is used.

Last updated: November 2026

The Atomic Energy Regulatory Board (AERB) sets requirements for Indian medical diagnostic X-ray facilities. A question that asks for an Indian dose limit calls for the AERB directive. A question about an international recommendation may give a different value for the same quantity.

Licensing and Responsibility

The Atomic Energy (Radiation Protection) Rules, 2004, issued under the Atomic Energy Act, 1962, provide the regulatory framework. The practice-specific code is AERB/RF-MED/SC-3 (Rev. 2), Radiation Safety in Manufacture, Supply and Use of Medical Diagnostic X-ray Equipment. It covers equipment, room shielding, worker monitoring, protective equipment and quality assurance.

The e-Licensing of Radiation Applications (eLORA) portal handles applications for regulatory consent. For a new installation, the employer obtains procurement permission and uses a type-approved or NOC-validated model from an authorised supplier. The installation needs suitable shielding and an approved Radiological Safety Officer (RSO). Acceptance testing checks the equipment after installation. The facility must obtain a licence to operate before imaging patients.

The employer has overall responsibility for radiation safety. The licensee arranges personnel monitoring, maintains dose records and investigates overexposures in consultation with the RSO. The RSO checks safety and monitoring systems, investigates potential exposures, and advises on corrective measures. Operators use the prescribed monitoring and protective devices and promptly report hazardous situations.

An AERB-authorised agency performs periodic quality assurance (QA) at least once every two years and after repairs with radiation safety implications. Periodic safety reports follow the format and interval specified by the regulator. An authorised agency handles decommissioning after prior intimation to the competent authority.

AERB Directive No. 01/2011 Dose Limits

Effective dose accounts for the contributions of different tissues. Equivalent dose here refers to the specified tissue, such as the lens of the eye. AERB uses calendar years. The effective-dose limits exclude natural background and the worker's or public member's own medical exposures.

ExposureAERB Directive No. 01/2011
Adult worker, effective dose20 mSv per year averaged over five consecutive years on a sliding scale, with a separate maximum of 30 mSv in any year
Adult worker, equivalent doseLens of eye 150 mSv in a year. Skin 500 mSv in a year. Extremities (hands and feet) 500 mSv in a year.
Declared pregnancy in a radiation workerEquivalent dose to the embryo or fetus limited to 1 mSv for the remainder of pregnancy
Apprentices and trainees aged 16–18Effective dose 6 mSv in a year. Lens equivalent dose 50 mSv. Skin and extremities 150 mSv each.
PublicEffective dose 1 mSv in a year. Lens equivalent dose 15 mSv. Skin equivalent dose 50 mSv.

Worked example. An adult worker’s five consecutive calendar-year effective doses in a hypothetical record are 14, 24, 19, 31 and 12 mSv. The sum is 100 mSv, so the average is 20 mSv per year. The 31 mSv year still exceeds the separate 30 mSv ceiling. Compare both numbers before deciding whether the record satisfies the directive.

The International Commission on Radiological Protection (ICRP) recommends a 50 mSv single-year effective-dose ceiling within its five-year average, and a separate eye-lens recommendation of 20 mSv per year averaged over five years with a 50 mSv single-year ceiling. Keep the ICRP eye-lens figure separate from AERB’s published 150 mSv annual limit. Optimise exposures well below the applicable limits.

How Personnel Dosimeters Work

The Bhabha Atomic Research Centre (BARC) developed a thermoluminescent dosimeter (TLD) badge using dysprosium-doped calcium sulfate, CaSO₄:Dy, in Teflon discs. Radiation creates trapped charge in the sensitive material. Heating the disc in a reader releases light that is measured and calibrated to estimate dose. The badge measures exposure and adds no shielding.

DeviceWhat the reading tells you
Passive TLDA laboratory measures the dose accumulated during the assigned monitoring period.
Optically stimulated luminescence dosimeter (OSLD)Controlled light stimulates the stored signal. The readout can often be repeated.
Electronic personal dosimeter (EPD)The device can display accumulated dose as work proceeds. Its use and any alarm thresholds depend on the monitoring programme.

Wearing and Handling a TLD Badge

  • Wear your assigned badge in its cassette at the location specified by the monitoring programme. In the AERB diagnostic X-ray example, a single badge goes at chest level under the lead apron when an apron is used.
  • Change and return the card at the assigned interval. AERB materials give quarterly monitoring as an example. The monitoring schedule depends on the facility.
  • Store unused staff badges away from radiation areas, alongside the designated control badge. Keep the control badge out of clinical exposures so its reading can account for background.
  • If a badge is accidentally left in the room during exposure, report it promptly to the RSO or licensee. Give the time, location and circumstances so the monitoring service can assess the reading against the worker’s actual exposure.
Loading diagram...
AERB licensing and personnel monitoring
Test Your Knowledge

What is the maximum effective dose an occupational radiation worker in India is permitted to receive in any single year under AERB regulations?

A

15 mSv

B

20 mSv

C

50 mSv

D

30 mSv

Test Your Knowledge

Which phosphor does the BARC personnel-monitoring TLD badge use, and where does AERB diagnostic X-ray guidance place a single badge when a lead apron is worn?

A

LiF:Mg,Ti; worn over the lead apron at the collar

B

CaSO4:Dy; worn under the lead apron at chest level

C

Al2O3:C; worn on the wrist outside the lead apron

D

CaF2:Mn; kept inside the X-ray control console drawer

Test Your Knowledge

Under AERB Directive No. 01/2011, what is the equivalent-dose limit for the embryo or fetus for the remainder of pregnancy once a radiation worker declares pregnancy?

A

1 mSv

B

5 mSv

C

6 mSv

D

20 mSv