3.2 Diagnostic X-Ray Production: Bremsstrahlung & Characteristic Spectra

Key Takeaways

  • X-ray production turns only a small fraction of incident electron energy into X-rays. The target must dissipate most of the energy as heat.

  • Bremsstrahlung forms a continuum with an endpoint set by peak tube voltage. Filtration reduces the low-energy part of the spectrum.

  • Tungsten K lines require electron energies above the roughly 69.5 keV K-shell threshold. The 59.3 keV line is an emission energy.

  • At stable kVp and waveform, mAs mainly changes photon fluence. Increasing kVp also raises the endpoint and generally hardens the beam.

  • Added aluminium filtration reduces photon fluence and raises mean energy. Apply the AERB minimum filtration for the equipment category.

Last updated: November 2026

Two mechanisms shape the spectrum from a diagnostic tube. Electrons slowing in the target produce a continuous range of X-ray energies. Electrons filling vacancies in target atoms produce discrete energies characteristic of the target.

Energy Transfer at the Target

A heated cathode supplies electrons by thermionic emission. They gain kinetic energy across the tube potential. One electron traversing 80 kV gains 80 keV, so in an 80 kVp exposure the fastest electrons can reach about 80 keV. A lower instantaneous voltage in a rippled waveform produces lower-energy electrons.

Most of the electron energy contributes to target heating. A rough thick-target model gives the radiative fraction as:

η≈0.9×10−9ZV\eta \approx 0.9 \times 10^{-9} Z V

Here ZZ is the target atomic number and VV is the accelerating voltage in volts for a constant-potential approximation. For tungsten at 100 kV, η≈0.9×10−9×74×105=0.00666\eta \approx 0.9 \times 10^{-9} \times 74 \times 10^5 = 0.00666, or about 0.67%. The fraction varies with voltage and the target. Tungsten's high atomic number and its ability to withstand high temperatures suit general radiography.

Bremsstrahlung and the Continuous Spectrum

In bremsstrahlung, the electric field near a target nucleus changes the velocity of a projectile electron. The electron can lose part or nearly all of its kinetic energy by emitting an X-ray photon. The range of energy transfers produces a continuum. Its upper endpoint corresponds to an electron transferring its full available kinetic energy to one photon. An 80 kVp exposure therefore has an endpoint near 80 keV.

Low-energy photons are preferentially attenuated in the target, tube assembly and added filter, leaving a harder emergent spectrum. A filter transmits fractions that depend on energy, so the spectrum has a reduced low-energy tail. Mean photon energy depends on the full spectrum. Effective energy is the monoenergetic beam energy giving the same specified half-value layer (HVL) and is a separate measure of beam quality.

Characteristic Lines and the Tungsten Threshold

A projectile electron with sufficient energy can eject an inner-shell electron from a target atom. An electron falling from a higher level into the vacancy may emit a characteristic photon with energy approximately equal to the difference in binding energies:

Echar=Eb,inner−Eb,outerE_{\text{char}} = E_{b,\text{inner}} - E_{b,\text{outer}}

To make a tungsten K-shell vacancy, the projectile must reach the approximately 69.5 keV K binding energy. Below that threshold the tube produces no tungsten K lines. Lower-shell lines may be generated and are strongly attenuated by the tube assembly and filter. Above threshold, tungsten has KαK_\alpha lines near 58.0 and 59.3 keV and strong KβK_\beta lines near 67 keV. The lines appear at those energies even as peak voltage and the bremsstrahlung endpoint move higher.

Target selection changes characteristic line energies. Molybdenum, used in some mammography systems, has a K edge near 20.0 keV and prominent lines near 17.5 and 19.6 keV. The following chart uses illustrative relative values to show the tungsten endpoint and line locations for two tube potentials. It simplifies the close lines within each K series into one feature.

Controls that Change the Spectrum

Change with other settings fixedMain spectral effect
Increase mAsRaises the number of photons approximately in proportion to the current–time product if the voltage waveform and filtration stay stable.
Increase kVpRaises the endpoint and generally increases both output and mean photon energy. Characteristic line energies remain fixed for a given target, while the line yield can change.
Change targetChanges the characteristic line energies and the efficiency of bremsstrahlung production.
Add aluminium filtrationReduces the photon fluence and preferentially removes low-energy photons. Mean energy rises, with the same tube energy endpoint.
Reduce voltage ripple at the same peak and mAsIn a comparison such as single-phase versus a low-ripple generator, more electrons reach energies near the peak. The yield and mean photon energy rise.

Worked Output Estimate

For an exercise assuming tube output proportional to mAs×kVp2\text{mAs} \times \text{kVp}^2, with the same geometry, generator and filtration, compare 60 kVp at 20 mAs with 90 kVp at 10 mAs:

I2I1=1020(9060)2=0.5×2.25=1.125\frac{I_2}{I_1} = \frac{10}{20}\left(\frac{90}{60}\right)^2 = 0.5 \times 2.25 = 1.125

The model predicts output 112.5% of baseline, a 12.5% increase. Use a measured output curve when accuracy matters. Patient attenuation and detector exposure can change when the spectrum hardens.

Filtration in Indian Equipment Requirements

Inherent filtration includes material in the tube and housing through which the beam travels. Added filtration further hardens the beam. AERB's Rev. 2 medical diagnostic X-ray safety code lists a 2.5 mm Al-equivalent minimum for constant-potential equipment in Appendix I §I.2.1. Table 1B gives separate minimum values by maximum rated tube potential for other equipment within that provision.

Illustrative Tungsten Spectra at 80 and 100 kVp (Relative Values)
Test Your Knowledge

What approximate minimum peak tube potential lets projectile electrons reach tungsten’s K-shell ionisation threshold?

A

33.2 kVp

B

59.3 kVp

C

67.0 kVp

D

69.5 kVp

Test Your Knowledge

At fixed tube potential and mAs, what does an added aluminium filter do to the beam transmitted through it?

A

Total beam quantity increases and maximum photon energy increases

B

Total beam quantity decreases and average photon energy increases

C

Total beam quantity decreases and maximum photon energy decreases

D

Total beam quantity remains constant and characteristic peaks shift rightward

Test Your Knowledge

In the linear operating range, what happens when mAs is doubled while peak voltage, voltage waveform, target and filtration stay fixed?

A

Approximately doubles spectral amplitude and area with the same endpoint and mean energy

B

Quadruples the height of the spectrum and shifts E_max rightward

C

Doubles E_avg while leaving total photon count unchanged

D

Eliminates K-characteristic spikes while preserving Bremsstrahlung output