5.3 Computed Radiography (CR), Digital Radiography (DR) & PACS
Key Takeaways
Computed radiography stores the latent image in traps in a photostimulable phosphor plate. A laser releases the stored signal, and an erasure stage prepares the plate for reuse.
Indirect flat-panel DR uses a scintillator and photodiodes. Direct-conversion DR uses a photoconductor to generate charge without an intermediate light stage.
Pixel pitch defines a Nyquist sampling limit. MTF and DQE describe different aspects of image quality and both depend on spatial frequency.
Exposure Index estimates exposure at the detector. Use the target and local protocol when interpreting Deviation Index, and review trends for dose creep.
DICOM defines image objects, file formats and communication services. PACS archives and retrieves imaging studies, with orders and demographics supported by the RIS and hospital system.
Computed Radiography
Computed radiography (CR) uses a reusable photostimulable phosphor (PSP) imaging plate in a light-tight cassette. Europium-activated barium fluorohalides, such as , are examples of storage phosphors.
- Store the image. X-ray absorption creates charge carriers. Some electrons become trapped at sites such as halide vacancies, called F-centres. The distribution of stored signal encodes the exposure pattern.
- Read the image. The reader sweeps a red laser across the plate. Stimulation releases trapped electrons and produces shorter-wavelength photostimulated luminescence at luminescence centres. For a studied plate, red stimulation at 633 nm produced an emission peak near 390 nm in the violet range.
- Digitise and erase. An optical filter separates emitted light from scattered laser light. A light guide feeds a photomultiplier in a typical reader, and an analogue-to-digital converter digitises the signal. Intense light then erases the residual signal before reuse.
The stored signal fades with time. Follow the reader and plate instructions for prompt readout, erasure and handling. A faint image from a preceding examination can indicate incomplete erasure and calls for a plate or reader check.
Direct and Indirect Flat Panels
| Detector pathway | X-ray converter | What reaches the readout |
|---|---|---|
| CR | Storage phosphor plate | A separate reader converts emitted light to electrical signals and digitises them |
| Indirect-conversion flat-panel DR | A scintillator, such as thallium-activated caesium iodide or gadolinium oxysulfide | Light reaches photodiodes. A common design uses amorphous silicon photodiodes with thin-film transistor (TFT) readout |
| Direct-conversion flat-panel DR | A photoconductor, such as amorphous selenium | An electric field drives generated charge toward pixel electrodes. Storage capacitors and TFT switching support readout |
A columnar caesium iodide scintillator guides much of its light toward the photodiodes, reducing lateral light spread. Direct conversion avoids scintillator light spread altogether. Actual sharpness depends on the converter, thickness, pixel pitch and other detector properties. Compare measured performance at a specified beam quality and spatial frequency.
Sampling, Signal and Exposure
Pixel pitch is the spacing between detector samples. For pitch in millimetres, the Nyquist frequency is line pairs per millimetre. A pitch of 0.14 mm gives approximately 3.57 lp/mm. That is a sampling limit for unaliased detail. Contrast can already be reduced at lower frequencies. The modulation transfer function (MTF) expresses the retained modulation at each spatial frequency.
Detective quantum efficiency (DQE) compares the squared signal-to-noise ratio (SNR) at a detector's output with that of the incident beam at the same spatial frequency :
A more efficient system can obtain the needed detector SNR with fewer incident quanta, under comparable conditions. DQE by itself does not determine a patient's dose. Beam quality, detector exposure and the imaging task matter when comparing systems.
Digital detectors operate over a wide exposure range, and processing can keep image brightness similar across different exposures. Noise can still limit underexposed images, and very high exposure can saturate the detector. When a team routinely increases exposure to suppress visible noise, acceptable-looking brightness can conceal dose creep.
Exposure Index (EI) estimates detector exposure using the relevant image region and the system's calibration. Legacy indices can behave differently. The Deviation Index (DI) relates a standardised EI to the target for the examination or projection:
DI near +3 means EI is about twice the target, and DI near −3 means it is about half. Inspect the exposure factors, field recognition, positioning, target selection and image quality when a result is unexpected. Compare series of similar examinations to detect drift. AAPM Task Group 232 supports evaluating action levels with local data. A single DI does not establish patient dose or a clinical need to repeat.
Processing and Viewing
Detector corrections can compensate for gain variation and some defective pixels. Field recognition and histogram analysis help select the signals used for rescaling. Incorrect collimation or a large prosthesis can defeat segmentation and distort the displayed image. Check the raw or for-processing image and the selected exam when available. At the workstation, window centre selects the centre of the displayed value range. A narrower window width maps a smaller input range across the output greyscale and can increase displayed contrast within that range. Strong edge enhancement can produce artificial borders or halos.
Image Communication and Artefacts
Digital Imaging and Communications in Medicine (DICOM) defines image objects, attribute tags, file formats and services. Examples of attributes include the Study Instance UID, patient identifiers and, when present, exposure indices. DICOM Modality Worklist lets the modality retrieve scheduled procedure and patient details. HL7 messaging can carry orders and results between information systems. The radiology information system (RIS) coordinates radiology workflow, and the picture archiving and communication system (PACS) stores, retrieves and displays imaging studies. Local implementations determine which system manages each step.
| Appearance | Check first |
|---|---|
| Periodic bands from a stationary grid | A mismatch between projected grid lines and detector sampling can alias into coarse bands. Review grid compatibility and orientation |
| Faint prior anatomy in CR | Check residual plate signal and the erasure cycle |
| Repeated dots or lines in fixed positions | Check detector defects, calibration or readout. Compare uniform test images |
| Unexpected greyscale after digital processing | Check field recognition and whether the correct anatomy and processing were selected |
For the barium fluorobromide CR plate described in the lesson, which sequence gives the stimulating light and emitted signal?
Ultraviolet laser stimulation emitting green PSL
Red laser stimulation emitting blue-violet PSL
Blue laser stimulation emitting infrared PSL
Green laser stimulation emitting amber PSL
Which property of direct-conversion amorphous selenium removes one source of blur found in an indirect scintillator detector?
Amorphous selenium stores F-center electrons longer than BaFBr plates
Amorphous selenium uses thicker phosphor crystals that scatter light laterally
Amorphous selenium converts X-rays directly into electrical charge guided straight by a bias field, eliminating lateral light spread
Amorphous selenium amplifies light inside an external photomultiplier tube
A chest radiograph acquired on a calibrated DR system displays a Deviation Index (DI) of +3.0 relative to the departmental target Exposure Index. What does this value indicate?
The measured Exposure Index is about twice the target Exposure Index
The measured Exposure Index is about half the target Exposure Index
The patient received exactly double the target effective dose
The receptor is at the target because the image brightness is unchanged
Sections you finish are checked off in the contents.