10.4 Processing & Digital QC, Viewing Standards & Reject Analysis

Key Takeaways

  • Daily sensitometry checks base plus fog, the speed index and the contrast index against a stable processor baseline.

  • A coin test reveals darkroom or safelight fog. A mesh or contact tool tests for local loss of sharpness from poor film–screen contact.

  • Deviation index measures the offset from the target detector exposure index. DI of 0 means the detector exposure index equals its target.

  • DICOM GSDF describes the relationship between grayscale values and luminance. A display test pattern checks whether the 5% and 95% inset patches remain visible.

  • For AAPM digital reject analysis, divide rejected acquisition events by all eligible acquisition events. Count one event per exposure-button press, even if it creates multiple source images.

Last updated: November 2026

Image-quality checks cover the entire path from acquisition through viewing. Use the AERB form applicable to the equipment for regulatory quality assurance (QA). Daily or periodic constancy checks in the department answer a different question: has the processor, receptor or display changed since its baseline?

Film Processing and Darkroom QC

Check developer temperature against the manufacturer’s setting before daily sensitometry. A temperature drift changes film response. Inspect and clean rollers and racks according to the processor’s maintenance instructions, and investigate recurring marks before changing the exposure technique. Recheck the control strip after corrective work.

A sensitometer exposes a step pattern on a control film. Use the light source appropriate for the film and a densitometer to measure the resulting optical density (OD) after processing. Establish the baseline on a stable processor using consistent film and chemistry. An International Atomic Energy Agency (IAEA) constancy method uses these three measurements:

MetricWhat to measureWhat a change can indicate
Base plus fogOD on an unexposed area of the control filmIncreased background density from film, chemistry or unwanted light
Speed indexOD of the fixed step chosen near 1.0 OD above base plus fog at baselineA shift in overall film response under the same exposure
Contrast indexDifference between the two fixed steps chosen near 2.0 and 0.25 OD above base plus fog at baselineChange in separation between light and dark tones

A rise in base plus fog of 0.03 OD is the IAEA handbook's suggested limit relative to baseline. For the speed and contrast indices the handbook gives ±0.15 OD from baseline as the general-radiography suspension level. For example, if base plus fog was 0.18 OD and is now 0.23, the rise is 0.05 OD. Investigate the processor or darkroom even though the observed reading alone might look plausible. Apply the specified protocol when another film type or local QA programme uses different criteria.

A coin test detects darkroom or safelight fog. Compare protected areas and areas exposed to the darkroom lighting for different durations on a sensitised test film. Visible coin outlines indicate unintended exposure. Check for white-light leaks, a damaged or unsuitable filter, or an incorrect bulb. A film–screen contact test uses a mesh or contact test tool to reveal local unsharpness. Check screen cleanliness and cassette pressure if a region stays blurred.

CR, DR and Display Checks

Computed radiography (CR) plates can develop ghosting if the previous latent image is incompletely erased. Check for traces from a previous exposure on a subsequent uniform test image. Follow the manufacturer’s routine erasure instructions. In digital radiography (DR), a uniform exposure can reveal detector or calibration artefacts that recur at fixed positions. Review test images consistently and investigate repeatable artefacts before changing technique or repeating a patient image.

Digital brightness processing can conceal gradual increases in detector exposure. The exposure index (EI) is a detector exposure indicator. The deviation index (DI) compares it with the protocol’s target exposure index EIT\text{EI}_{\mathrm{T}}:

DI=10log⁡10(EIEIT)\text{DI}=10\log_{10}\left(\frac{\text{EI}}{\text{EI}_{\mathrm{T}}}\right)

If EI equals its target, DI is 0. An EI about twice the target gives DI near +3. This comparison describes detector exposure relative to target. Patient dose and need for a repeat require separate assessment. Review distribution shifts by protocol. AAPM Task Group 232 recommends evaluating action limits using the site's own DI data.

The DICOM Grayscale Standard Display Function (GSDF) maps grayscale differences to luminance changes using a model based on just-noticeable differences. Use a Society of Motion Picture and Television Engineers (SMPTE) or AAPM TG18-QC pattern to check the 5% inset in the 0% patch and the 95% inset in the 100% patch. Review the display under its normal controlled room lighting and investigate glare or unreadable patches. Maximum luminance and room-light criteria depend on the display's role and the QA protocol.

Digital Reject Analysis

Count eligible acquisition events by exposure-button press. A single press may produce multiple source images in a stitched or dual-energy examination. A rejected event yields no diagnostic image for that acquisition. Exclude test exposures and practitioner-directed acquisitions designated outside the reject analysis. Record the reject reason and preserve the reject information for authorised quality review. Distinguish an input image within a combined examination from a separate patient exposure event.

Reject rate=Rejected eligible acquisition eventsAll eligible acquisition events×100%\text{Reject rate}=\frac{\text{Rejected eligible acquisition events}}{\text{All eligible acquisition events}}\times 100\%

Worked example. A department logs 4,000 eligible DR acquisition events and 240 rejects, giving 6.0%. Of the rejects, 108 are coded as positioning or anatomy cutoff, so this category represents 45% of rejects and 2.7% of the eligible acquisitions. The chart uses hypothetical counts from this example. Review the coded reasons and representative images together. A spike in one projection can point to a positioning issue, while an artefact recurring in one room can point to equipment. Choose a corrective action and remeasure using the same inclusion criteria.

Hypothetical reject counts by reason (240 events)
Test Your Knowledge

A general-radiography processor has a base-plus-fog baseline of 0.18 OD. Today it measures 0.23 OD. Using the IAEA handbook’s suggested +0.03 OD limit from baseline, what is the result?

A

The rise is 0.03 OD and meets the limit

B

The rise is 0.05 OD and exceeds the limit

C

The change is 0.05 OD lower than baseline

D

The baseline is irrelevant because 0.23 OD always passes

Test Your Knowledge

What does grayscale calibration address, and what should be visible on a TG18-QC grayscale test pattern?

A

To automatically adjust patient mAs on the X-ray generator during exposure

B

To eliminate the need for quarterly CR imaging plate erasure

C

To maximize ambient room lighting above 500 lux during image reporting

D

To ensure perceptually linear grayscale steps (JNDs) and verify visibility of 5% and 95% low-contrast patches

Test Your Knowledge

A department logs 4,000 eligible DR acquisition events and 240 rejects, of which 108 are due to positioning. What is the overall reject rate and the share of rejects due to positioning?

A

2.7% overall and 45% due to positioning

B

10.8% overall and 45% due to positioning

C

6.0% overall and 45% due to positioning

D

6.0% overall and 2.7% due to positioning