9.2 Reagent Suitability Assessment, Storage, Expiry & Quality Logs
Key Takeaways
Check the lot, expiry, condition and storage history before use.
Keep lots traceable and follow the method’s approved verification before patient use.
Use the product’s instructions and hazard compatibility information when storing or segregating chemicals.
Rotate dated stock by earliest expiry among accepted lots and quarantine suspect material.
Use the approved chemical waste stream and report unknown or unlabelled material.
Suitability can change during shipping, storage or use. Check the unopened expiry, and where applicable the open-vial, reconstituted or on-board limit. Check the label, storage history and the condition of the container and material against its instructions. Appearance may flag a problem, but acceptable appearance alone cannot demonstrate suitable performance.
Inspect and Investigate
A leaking seal, damaged bottle, unexpected colour or cloudiness, or unaccounted temperature excursion requires attention before use. Some reagents can form crystals in storage that may be dissolved only under a specific instruction. Ask the designated reviewer before changing temperature or attempting a recovery. For prepared media, inspect the label, expiry, package and appearance against the established acceptance criteria. Keep a damaged or unexpectedly changed supply out of service. Follow the receiving bench’s quality process.
If two reagent bottles have the same catalogue number but different lots, preserve the separate identifiers and expiry information. Do not top up an old container with contents from a fresh one. At installation, check the instrument’s acceptance or lot registration procedure and the requirements for calibration, controls and any comparison. A passing control may still leave patient sample comparison pending. For applicable quantitative assays, CLSI EP26 describes comparing patient results between lots to detect clinically important differences. Other methods may require a different approach and the scope of EP26 is limited.
Suppose the new reagent lot passes both routine control levels but the method also requires a patient-sample comparison. Leave the new lot pending until the comparison is complete and authorized. If the comparison fails, record the results and affected lot and refer it for review. This avoids interpreting a control result as final acceptance of every performance requirement.
A method can also require verification when a shipment arrives even though the lot number matches existing stock. For example, Shared Health Manitoba’s published urinalysis verification procedure distinguishes material that has already been in service from the incoming shipment. Retrieve the local acceptance record before making a shipment available. Record its receipt date, expiry, lot and verification disposition so it can be distinguished from material already cleared.
Store and Rotate by the Applicable Conditions
Read the manufacturer’s storage requirements for the actual product and condition. A new unopened container may have different limits after opening or loading on the instrument. Identify the applicable temperature, light protection, storage position and monitoring. Record actual conditions and follow the site’s excursion procedure when a limit is exceeded.
For dated stock, first expiry, first out means selecting the earliest-expiring acceptable stock, even if it was delivered later. If an approved lot received September 2 expires December 31 and a lot received September 16 expires November 30, use the November lot first if both are cleared and suitable. An expired, recalled or pending lot stays out of active inventory. Date of receipt is relevant to traceability even when expiry determines rotation.
| Finding | Appropriate next action |
|---|---|
| Expired or suspect lot | Identify and quarantine it, record the lot and reason, and notify the designated person. |
| Brief refrigeration alarm | Preserve the affected material under the approved conditions, capture the event and ask for a product-specific disposition. |
| Unclear handling on a hazardous product | Match the container identity to the safety data sheet and use the site’s storage and compatibility procedure. |
Segregate and Dispose Safely
Select storage cabinets and secondary containment for the actual hazards and jurisdiction. Separate materials that can react dangerously even if they share another hazard classification. For example, an acid and a strong base should have segregated containment under the site’s chemical compatibility plan. Read the safe storage and incompatibility sections of the safety data sheet. Use the disposal information together with the facility and applicable jurisdictional requirements.
An unknown clear liquid cannot be classified by sight. Keep people from handling it, alert the designated person and use the site’s unknown-material process. Do not combine chemical waste streams unless the approved procedure establishes compatibility and permits it. Alberta Health Services’ hazardous chemical waste procedure lists separate waste pathways and incompatible chemicals. Use approved, correctly labelled containers for the identified streams.
Maintain the receipt, lot, opening or preparation date, assigned expiry, storage history where required, verification and final disposition. Record what was held or moved, whom you notified and which decision allows work to continue. The record helps the laboratory identify tests or materials affected by a recall or a failed comparison.
See CSMLS reagent and inventory competencies, CLSI EP26 quantitative reagent lot evaluation, WHO expiry-based stock rotation, CCOHS hazardous chemical storage cabinets, CCOHS safety data sheets, CCOHS hazardous waste management, AHS hazardous chemical waste procedure, Shared Health published urinalysis lot and shipment procedure.
A new quantitative reagent lot passes both routine QC levels. The method also requires a patient-sample comparison before lot acceptance. What should the MLA do?
Pour the new lot into the old bottle to preserve the old calibration
Start reporting because the controls passed
Keep the new lot separate and on hold until the required comparison and authorization are complete
Skip the comparison because the manufacturer checked the lot
Two suitable cleared lots are available. The lot received September 2 expires December 31. The lot received September 16 expires November 30. Which should be used first under first expiry, first out?
The lot expiring November 30
The lot received September 2 because it arrived earlier
A mixture of the lots
Either lot because both were received in September
The MLA has two identified incompatible chemical wastes and an unlabelled beaker containing a clear liquid. What is the appropriate response?
Pour them down the sink with running water
Follow the approved separate streams for the identified wastes and alert designated staff about the unknown liquid
Mix all three materials in the same waste container
Treat the unlabelled liquid as water because it is clear