14.1 Quality System Essentials (QSE), Internal/External QC, Calibration & Proficiency Testing
Key Takeaways
CLSI QMS01 describes 12 quality system essentials. ISO 15189:2022 specifies quality and competence requirements.
Distinguish a warning, a rejection, a shift and a trend using the applicable control plan.
When QC fails, suspend affected testing and hold results. Record the failure and notify the designated lead.
Use routine staff and methods for external assessments, following challenge-specific instructions.
How the quality system supports patient testing
A quality management system (QMS) coordinates people, procedures and records for safe and reliable results. CLSI QMS01, 5th edition, describes 12 quality system essentials (QSEs). ISO 15189:2022 sets requirements for quality and competence in medical laboratories and point-of-care testing (POCT). Accreditation schemes vary. For example, BC's Diagnostic Accreditation Program (DAP) has a scheme that uses both ISO 15189:2022 and its own laboratory medicine standards.
- Organization: Clear leadership structure, roles, authority, and accountability.
- Customer Focus: Meeting the clinical needs of patients, ordering clinicians, and care teams.
- Facilities and Safety: Biosafety containment, ergonomics, ventilation, spill readiness, and temperature-controlled workspaces.
- Personnel: Qualification verification, orientation, ongoing competency assessments, and continuing education.
- Purchasing and Inventory: Vetted suppliers, cold-chain receiving checks, stock rotation, and recall tracking.
- Equipment: Preventive maintenance, calibration, temperature monitoring, and service logs for centrifuges, refrigerators, and analyzers.
- Process Management: Validated workflows across pre-analytical, analytical, and post-analytical stages, including Quality Control (QC).
- Documents and Records: Controlled Standard Operating Procedures (SOPs), specimen logs, and secure record retention.
- Information Management: Accurate Laboratory Information System (LIS) entry, barcoding, privacy compliance, and downtime contingencies.
- Nonconforming Event Management: Detecting, containing, investigating, and reporting errors or near-misses.
- Assessments: Internal audits, External Quality Assessment (EQA) surveys, and accreditation inspections.
- Continual Improvement: Using quality indicators and corrective actions to strengthen performance over time.
Quality assurance (QA) covers the full testing pathway. Monitoring specimen rejection rates, auditing patient identification and checking that corrected results reach the right clinician are QA activities. Internal quality control (IQC) monitors a measurement procedure during use. For example, a glucose control at a known level can help detect a change in meter or strip performance. The lab selects control levels, materials, frequency and decision rules for each method.
Read a Levey–Jennings chart
For a quantitative procedure that uses statistical QC, plot each control level against its own mean. The distance from the mean is measured in standard deviations (SD). A shift begins with an abrupt change followed by values that stay on one side of the mean. A trend is a sequence drifting in one direction. A shift can follow a reagent lot change, and a trend can prompt a check for reagent degradation. Either pattern is a clue for investigation. The lab's control plan defines the rule and sequence length.
| Example Westgard rule | Pattern | Significance when the rule applies |
|---|---|---|
| One point above +2 SD or below −2 SD | A warning in the classic multirule procedure. Check the other rules. | |
| One point above +3 SD or below −3 SD | Reject. This rule is sensitive to large errors, including random errors. | |
| One point above +2 SD and another below −2 SD in the same run | Reject. Opposing extremes suggest random error. | |
| Two consecutive points above +2 SD or two below −2 SD | Reject. A consistent direction suggests systematic error. | |
| Ten consecutive points on the same side of the mean | Reject. This pattern suggests systematic error. |
For example, +2.2 SD and −2.1 SD span 4.3 SD. With both results in the same run, they satisfy the opposing limits for the rule. A device that evaluates QC against a lot-specific range may give a failed QC indication without producing a chart.
Respond to failed QC
- Suspend affected patient testing and hold affected results. Route urgent tests to an approved alternative through the designated lead.
- Record the original failure, device and control/reagent lots. Check the procedure and only the factors you are authorized to troubleshoot, such as correct control level and lot, expiry, storage or mixing. Follow any permitted repeat procedure and record each attempt.
- Notify the supervising technologist or designated lead. The authorized person investigates affected patient results, any needed corrections and conditions for resuming testing.
Calibration and external assessment
Calibration establishes the relationship between the device response and a known reference value. A method may require checks across its measuring interval to verify the calibration. Use the device procedure and authorized roles. A routine QC vial serves the separate function of monitoring how the calibrated measurement procedure is performing.
When authorized to perform a device calibration or verification, confirm the method, status, materials and expiry first. Document the result and compare it with the approved acceptance criteria. For an illustrative bench check, suppose a 10.0 mmol/L verification material must read from 9.5 to 10.5 mmol/L. A reading of 10.6 mmol/L fails that check. Hold affected work and report the result under the procedure. For a new reagent or media lot, verify the conditions and QC or release criteria appropriate to that material before use, as described in Chapters 9 and 12.
External quality assessment (EQA), often called proficiency testing (PT), provides an independent comparison against a reference or peer target. Use challenge-specific instructions for preparation, reporting and any referral restrictions. Routine testing personnel should handle the challenge using the normal analytical workflow. Avoid extra determinations or informal comparisons with other participants before submission. Escalate an unacceptable result for investigation and document corrective action under the program and accreditation rules.
See CLSI QMS01, CLSI C24, CLSI EP23, Westgard multirules, BC DAP laboratory standards, BC DAP external assessment, CAP program terms.
A normal chemistry control drifts downward over seven consecutive measurements without an abrupt jump. Which interpretation fits the pattern?
A shift from an abrupt change in the method
A trend that warrants checks for causes of gradual drift
A wide spread in opposite directions within one run
An automatic failure under every laboratory QC plan
A BC laboratory receives an EQA challenge with instructions for preparing its sample. Who should perform the examination and how?
A visiting specialist should choose a method that is more sensitive than the one routinely used
Two laboratories should compare their values before submitting
The most senior staff member should make several extra measurements and select the closest match
Personnel who routinely perform the test should use the routine method and follow challenge instructions
A glucose meter flags a failed liquid control during morning POCT checks. What should the MLA do first?
Stop patient testing with the affected meter, hold affected results and follow the QC failure procedure
Proceed with patient testing and flag each result with a comment
Assign the control target value as a new calibration for the meter
Repeat the control and record only the first passing result