14.2 Risk Management, Incident Reporting, Document Control & Continuous Quality Improvement

Key Takeaways

  • Contain patient and staff risk and document the observable facts.

  • Incident analysis investigates what happened. FMEA anticipates what could fail.

  • Record patient care and disclosure in the appropriate clinical record. Use the approved system for internal event reports.

  • Raise discrepancies in controlled procedures and monitor the effect of authorized changes.

Last updated: November 2026

Manage risk before and after an event

Check likelihood and severity together. A sample identification error may be infrequent yet carry serious consequences. Barcoded positive patient identification and a defined mismatch hold are examples of controls. Follow up to see whether those controls work in practice.

ToolWhen to use itExample
Incident analysis, including root cause analysis (RCA)After an incident or a near missTrace how two specimens received similar identifiers. Examine the workstation layout, software design, interruptions and cues at the time.
Failure modes and effects analysis (FMEA)Before a proposed change or during proactive review of an existing processMap where a new pneumatic transport process might leak, break a tube or fail to route a specimen. Rank the hazards and design controls.

A just culture supports fair evaluation and speaking up. An incident analysis explores contributing conditions and assigns actions to reduce risk. A deliberate decision to bypass an identification safeguard may require a separate accountability response.

Contain, report and document

Protect the patient and staff first. For example, hold a label mismatch and contact the lead. For an exposure or an equipment fault, prevent access as appropriate and activate the relevant emergency process. Report promptly through the site's approved channels. An event report should state what was observed, when it happened, which specimen or equipment was affected, actions taken and who was notified. Avoid speculation and include only the identifiers required by the secure reporting system.

The internal quality report and the clinical record serve different purposes. Record the appropriate facts, patient impact, notifications, care actions and authorized disclosure in the clinical record as required by policy and law. For example, the care record may state that recollection was requested at 09:15 after the laboratory found a specimen integrity problem. Use the designated incident system for the internal analysis. Follow applicable disclosure rules. For example, Ontario hospital rules require specified information about qualifying critical incidents and disclosure in the patient record.

Work from controlled documents

Use the current approved SOP and related forms or job aids. Controls typically include a title and unique identifier, approval, revision or effective date and authorized distribution. Review documents on the required schedule and remove superseded copies from use. For a bench procedure that differs from the approved SOP, hold the affected step if needed, protect specimens and notify the lead. A change must be authorized, incorporated into the controlled procedure and communicated to staff before routine use.

Example: At accessioning, the MLA finds a barcode from another patient attached to a urine container. Stop processing and segregate the affected specimen. Follow the identity mismatch policy and notify the MLT. The MLA reports the event objectively and documents the care-relevant facts in the designated record. Recollection or a documented exception is decided under the identity SOP.

Check whether a change helps

Use a Plan–Do–Study–Act (PDSA) cycle to test a change on a small scale. Plan a daily tally of hemolyzed specimens and a pilot change in collection workflow. Try it on one unit, compare the post-pilot proportion with its baseline and look for unintended collection delays. Revise and test again, or seek approval to implement the change and monitor it over time. Keep the denominators and time window comparable. Other useful indicators include rejected samples, quantity not sufficient, identification errors, corrected results and STAT turnaround time.

An internal audit compares evidence such as accession records or current bench SOPs against defined requirements. Track the findings, corrective actions and effectiveness checks.

Worked quality indicator

A unit sends 240 eligible specimens one week. An accessioning tally shows 12 that met the laboratory's hemolysis reporting criterion. The baseline is 12 ÷ 240 × 100 = 5%. During the pilot it sends 200 eligible specimens and six meet the same criterion, giving 6 ÷ 200 × 100 = 3%. This is a decrease of two percentage points. Keep the inclusion criteria the same and check whether the mix of specimens changed. A small sample or changes in the unit's workload can affect the comparison. Check a balancing measure such as collection-to-receipt time as well. Report the result so the team can decide whether to extend the pilot.

If rejection declines while turnaround time worsens, investigate both effects before proposing wider adoption. A supervisor can authorize another pilot using the same definitions and a different shift. This helps distinguish a sustained improvement from an unusually quiet week.

See BC DAP laboratory standards, Healthcare Excellence Canada incident management toolkit, Ontario Hospital Management regulation, Institute for Healthcare Improvement PDSA, Institute for Healthcare Improvement FMEA.

Loading diagram...
From a near miss to a safer process
Test Your Knowledge

After a centrifuge tube breaks, the bench has been secured and the supervisor notified. How should the MLA document the safety event?

A

Use an informal bench note as the only record

B

Summarize it from memory at the next staff huddle

C

Use the approved incident system and record the facts, containment and notifications

D

Enter a speculative cause in the patient chart

Test Your Knowledge

A controlled receiving SOP specifies one swab, but stock and bench practice changed. What should the MLA do?

A

Escalate the discrepancy through the quality and document control system for an authorized resolution

B

Cross out the obsolete entry in the controlled SOP and initial it

C

Keep using the bench substitute without reporting the mismatch

D

Post an unofficial list of substitute swabs over the SOP

Test Your Knowledge

Before introducing pneumatic transport, a team maps possible failures and ranks their consequences so it can design safeguards. Which tool is it using?

A

An internal compliance audit

B

Retrospective root cause analysis

C

A Levey–Jennings chart

D

Failure modes and effects analysis