12.3 Basic Testing Principles, Troubleshooting & Disease-State Correlation
Key Takeaways
Distinguish the signal from the clinical meaning assigned by the responsible professional.
Recognize optical interference and electrochemical measurements as different principles.
Recognize red-cell clumping and method-specific immunoassay interference as possible analytical problems.
Understand what the erythrocyte sedimentation rate measures and check the method before proceeding.
Document flags, observed defects and authorized corrective actions without changing patient values.
A laboratory result reflects a specific measurement method applied to a particular specimen. A number may be unexpected because of a real patient condition, a specimen problem or interference in the method. The MLA identifies relevant evidence, preserves the record and follows the escalation procedure so the designated reviewer can investigate.
Identify Which Signal the Device Uses
A photometric method measures absorption at selected wavelengths. Turbidity or coloured material in a specimen can affect optical methods, and the effect can vary between measurements and platforms. Cell rupture can also add substances to the liquid portion of blood. Potassium, lactate dehydrogenase and aspartate aminotransferase can increase with in vitro hemolysis. Follow the method’s specimen and interference criteria rather than assigning a correction to the measured result.
Ion-selective electrodes measure an electrical response related to the activity of the target ion. A direct method uses an undiluted specimen. Some core chemistry methods use an indirect measurement that dilutes a defined volume. In pronounced lipemia, lipids occupy more space in that volume, leaving a smaller water fraction. An indirect method may then report falsely low sodium. In a case with a lipemic specimen and discordant readings, verify the matrix, instruments and collection circumstances and refer the results. The worker cannot select a patient’s reportable result simply by preferring the higher number.
Immunoassays use binding between an antigen and antibody. Health Canada cautions that high biotin intake can interfere with assays using biotin technology. Whether results are biased upward or downward depends on the assay. Follow the manufacturer and laboratory instructions. Some sandwich methods can show a high-dose hook effect, in which excess target can cause an inappropriately low signal. Refer an apparently inconsistent result through the site’s procedure. Any repeat by dilution needs an authorized method and proper records.
Connect a Cell Count to the Observed Specimen
An impedance method registers changes in electrical resistance as particles cross the sensing area. Pulse counts and sizes inform cell counts and volumes under the instrument’s method. Some analysers also use optical scatter and fluorescence to distinguish cells. A clump can behave differently from a single cell passing the detector.
Cold-agglutinin interference can cause an underestimated red blood cell count and altered red cell indices. Other problems can also give discordant indices. The analyser flag and specimen appearance call for review, rather than proving the cause or authorizing warming. Verify the sample and enter the observed finding without changing or manually recalculating the reported values.
Understand Erythrocyte Sedimentation
The erythrocyte sedimentation rate (ESR) relates to the settling of red cells in anticoagulated blood. The Westergren reference method reads the millimetres of settling after one hour, reported in mm/h. Modified or alternative instruments can use different measurement principles and run times. Use the approved method for the site, including its specimen, stability, level or positioning and quality checks. Some APL sites use an approved C-reactive protein substitution policy for certain requests, so confirm the actual test to be performed.
In a classroom example of the reference method, the clear plasma column above the settled red cells measures 18 mm at one hour. The corresponding reading is 18 mm/h. The number by itself does not identify the cause of an inflammatory response. If the rack is tilted or the timer was set incorrectly, document the deviation and have the bench decide the next step under the approved procedure.
Compare Against the Applicable Decision Limit
The reporting laboratory’s reference interval describes results in its reference population. A critical notification limit is a separate threshold used to trigger communication. APL’s February 2026 Calgary biochemistry list, for example, uses different high-potassium notification limits for newborns, children and adults. Do not substitute a generic cutoff or assume a specimen flag explains away an abnormal result. Check age group, specimen condition, method, and the authorized urgency pathway.
Preserve the Evidence and Get Review
- Determine whether the problem is limited to one sample, a supply lot or an analyser.
- Compare the observed condition and alarm with the operating procedure. Perform only the checks permitted for volume, clots, air, reagent status and controls.
- Hold affected work and refer persistent alarms or failed required QC. Preserve the original results and error details. Use the authorized release and clinical interpretation pathways.
See CSMLS analytical knowledge requirements, ICSH reference and alternative ESR methods, APL ESR substitution bulletin, Roche analyser measurement principles, Health Canada biotin interference, Health Canada analytical validation guidance, Lipemia and sodium measurement study, Hemolysis measurement interference study, Cold agglutinin reflex study, Sysmex optical and impedance analyser, Sysmex DC sheath flow detection, APL Calgary February 2026 critical values, FDA primary hCG assay interference review.
An identified, markedly lipemic specimen gives a substantially lower sodium value on indirect ISE than on a comparable direct ISE. Which method effect can contribute?
Lipid causes the ion-selective electrode to measure emitted light
Biotin makes sodium bind to the reagent bottle
Cold agglutination changes the speed of the ion-selective electrode
A reduced plasma-water fraction affects the fixed-volume dilution in the indirect method
In a valid Westergren reference-method demonstration, the plasma column above the settled red cells measures 18 mm at one hour. What is the corresponding ESR reading?
18 mmol/L
18 mm/h
18 mOsm/kg
1,080 mm/h
A chemistry analyser repeatedly flags the sample probe and required QC fails. What should the MLA do?
Suppress the flags and resume testing
Report the control value as a patient result
Hold affected work, record the alarm and QC, and alert the designated professional
Change the calibration value to make the QC pass