9.1 Reagent, Solution, Stain & Media Preparation, Dilutions & Labelling
Key Takeaways
Verify the preparation instructions, water or supplied diluent, and the lot and expiry before mixing.
Use suitable measuring equipment and follow the reagent’s specific mixing and dissolution instructions.
Use consistent units and distinguish stock volume from final volume in a dilution.
Verify a prepared buffer’s pH at the specified stage with the approved, calibrated instrument.
Record the preparation, expiry and storage instructions and apply the workplace label where required.
A working reagent should have a reproducible composition and a traceable record. Start by identifying the approved procedure and matching it to the product and lot. Confirm the intended final strength and volume, the required water or diluent, equipment, storage and verification. Check who may prepare the material and who authorizes its use.
Select Water and Measure Carefully
CLSI GP40, fifth edition, distinguishes clinical laboratory reagent water (CLRW), special reagent water (including molecular biology water), instrument feed water, manufacturer-supplied water, wash or autoclave water, and bottled purified water. Check the method specification. Some methods require a supplied diluent or a specified grade of distilled or deionized water. Verify the applicable water-system status or lot and use the validated source. The laboratory’s monitoring program may track ionic, organic and microbial contamination.
Reconstitution adds the instructed volume and type of liquid to a prepared dry product. Some controls are freeze-dried. The manufacturer’s coagulation control plasma instructions, for example, call for distilled water, gentle swirling and a standing period, and specifically direct the operator to avoid inversion for that product. Other products can have different mixing instructions. Inspect the vial and the finished preparation against its acceptance criteria, then document the actual method and storage time.
A buffer helps maintain pH over a defined range of use. Verify the calibration and suitable standards for the pH meter before checking the prepared buffer. Use the specified temperature and wait for a stable reading. If the value falls outside the procedure’s limit, document the reading and follow the authorized adjustment or remake instructions. For hazardous reagents, review the safety data sheet and follow the controls for the specific task. CCOHS warns that diluting sulfuric acid releases heat and can spatter. Use the authorized procedure and safe order of addition.
Calculate a Working Solution
For an ordinary dilution of one stock, the amount of solute transferred equals the amount in the final mixture. If the concentration units are compatible, use . The subscripts distinguish the stock (1) from the final mixture (2).
A classroom example calls for 500 mL of a 0.5% mass/volume dye solution from 5.0% mass/volume stock. Divide the final concentration by the stock concentration and multiply by the final volume. The stock volume is 0.5 ÷ 5.0 × 500 mL = 50 mL. Measure the stock and use the diluent to reach 500 mL final volume. Adding stock to 500 mL of diluent would overshoot the specified total.
When a method defines a fivefold dilution as one part stock in five total parts, 1,000 µL final volume requires 200 µL stock and a nominal 800 µL diluent. One part stock plus five parts diluent instead gives six total parts. State the actual parts and final volume when documenting a ratio. In a serial dilution, multiply the factor at each stage. Four successive twofold dilutions give factors of 2, 4, 8 and 16, each relative to the starting material.
| Target | Calculation |
|---|---|
| 250 mL of 0.9% mass/volume sodium chloride | 0.9 g per 100 mL × 250 mL = 2.25 g. Bring the dissolved solute to the required 250 mL final volume. |
| 500 mL of 0.20 mol/L sodium chloride | 0.500 L × 0.20 mol/L = 0.100 mol. Using 58.44 g/mol gives 5.844 g before rounding under the preparation procedure. |
Label, Verify and Record
Clinical preparation records commonly identify the working solution, concentration, stock lot, preparation date and time, preparer, assigned expiry and storage conditions. Add the preparation and verification details the laboratory requires. Use the assigned stability rule. If a hypothetical site specifies 30 days after preparation or the parent expiry, whichever comes first, a preparation made on October 1 from stock that expires October 15 has an October 15 limit.
Workplace hazard communication and the reagent record serve different purposes. A workplace label for a hazardous product generally identifies the product, describes safe handling and refers to the safety data sheet when available. Apply the rules for the jurisdiction and laboratory, including any specific conditions for decanted products. A complete inventory record cannot replace required hazard information on the container. Store the finished preparation according to the validated instructions and hold it if the required verification is incomplete.
See CSMLS reagent competencies, CLSI GP40 fifth edition, Thermo Fisher coagulation control plasma instructions, METTLER TOLEDO buffer preparation and pH checks, CCOHS sulfuric acid profile, Ontario WHMIS regulation, CCOHS laboratory WHMIS requirements, NIST sodium chloride reference, NIST volumetric calibration.
The coagulation control instructions specify distilled water, gentle swirling and a standing period. Which preparation follows the instructions?
Substitute tap water because water is only a solvent
Measure the specified distilled water, swirl gently and wait the prescribed period
Invert and vortex the vial instead of waiting
Use the control immediately while undissolved material is visible
A practice calculation calls for 400 mL of 0.5% mass/volume dye from a 5.0% mass/volume stock. What volume of stock should be used?
4 mL of stock made up to 400 mL
20 mL of stock made up to 400 mL
40 mL of stock added to 400 mL of diluent
40 mL of stock made up to 400 mL final volume
A lab procedure assigns the earlier of the working stability limit and parent expiry. A buffer is prepared October 1, its working limit is 30 days, and the parent expires October 15. Which date applies?
October 15
October 31
November 15
No expiry if the preparation date is logged