8.1 Centrifuges, Pipettes, Water Baths, Incubators & Maintenance Logs
Key Takeaways
Check the instrument, selected unit and approved configuration before use.
Use the validated rotor loading pattern and compare total assembly mass when balancing.
Use the pipette mode, tips and technique appropriate for the liquid and the instrument.
Read TC and TD markings before using calibrated volumetric ware.
Record measured values, the applicable limits and the disposition of affected work.
The MLA checks that equipment is fit for the assigned task before using it. Confirm the instrument identifier, applicable procedure, inspection status and scheduled verification. Read the acceptance limits before recording the observation. A space labelled “pass” cannot replace the actual reading when the procedure requires the numerical value.
Centrifuge Readiness
A fixed-angle rotor holds tubes at a set angle. A swinging-bucket rotor allows carriers to pivot outward under rotation. The tube, rotor and adapter must suit the method. For example, Newfoundland and Labrador Health Services’ clinical centrifuge reference specifies use of the approved setup and notes a local restriction on accepting specimens from fixed-angle centrifuges for the affected instrumentation. The allowed rotor and run settings are part of the procedure.
Check the rotor and compatible accessories for visible damage. Use the tube and cap specified and include each cap, carrier or insert when comparing the mass of assemblies. Follow the loading pattern permitted for that rotor. Two identically sized tubes filled to the same line can differ in mass if their liquids differ in density. The manufacturer of the Eppendorf 5427 R, for example, calls for compatible tube types and symmetric loading verified by weighing the assemblies.
Compare the selected unit against the procedure. Revolutions per minute (rpm) tell you how fast the rotor turns. Relative centrifugal force (RCF) is proportional to radius and the square of speed. At a fixed 3,000 rpm, the 15 cm rotor in the Chapter 6 example delivers 1.5 times the RCF of the 10 cm rotor. The number in rpm cannot replace the specified reading in × g.
If breakage is suspected, use the safe stop defined for that instrument, keep the lid closed and notify the designated responder. Procedures may set a waiting time and specify protective equipment and approved cleanup materials. Follow that procedure rather than opening early to see what broke. The unit remains out of service until reviewed and cleared as required.
Mass and Volume
| Instrument or marking | Use and pre-use check |
|---|---|
| Top-loading or analytical balance | Select a balance whose range and readability suit the measurement. Check level, zero and required reference-weight checks. Use the draft shield on a balance so equipped. |
| TC or “to contain” | A volumetric flask holds the calibrated volume at its mark. Use it to make a solution up to the intended final volume. |
| TD or “to deliver” | A transfer pipette delivers its calibrated volume under the stated drainage conditions. Check whether the individual design calls for a blowout. |
Tare a clean weighing vessel so the displayed net mass excludes the vessel. If a loaded vessel is 4.82 g and the empty vessel is 2.17 g, the net mass is 2.65 g. Verify that the balance is suitable for that precision. A check with a reference weight shows whether performance is within the allowed limits. It is not permission to adjust a calibration.
In an air-displacement pipette, the piston moves an air cushion above the sample. The Eppendorf Reference 2 manual describes forward pipetting from the first stop, followed by delivery and a blowout. Reverse mode takes up an extra amount and delivers to the first stop, leaving the surplus behind. This can help with a viscous liquid. A positive-displacement device uses a piston in contact with the liquid and can be better suited to challenging liquids. Use the assigned method and volume range. During gravimetric verification, the measured mass of water is converted to volume under the specified conditions. Use the applicable correction and tolerance.
Temperature, Cycle and Maintenance Records
For a refrigerator, freezer, water bath or incubator, verify the required operating conditions for the actual test or material. Use the specified measurement point and checking frequency. If a buffer labelled 2–8 °C is found in a unit displaying 9 °C, record the reading and follow the temperature-excursion procedure. Review the duration, monitoring record and affected lots with the responsible staff before deciding whether the material can still be used.
A steam sterilizer requires a validated cycle for its load. Physical readings and chemical and biological monitors answer different questions. A mark on process tape can show exposure to a condition. Review the required cycle record and indicators before authorizing release. Hold the load if the applicable monitor fails or a required result is pending.
Record equipment ID, date and time, the actual observation and limit, and any notification, interim action or return-to-service approval. Keep records linked to the affected samples and materials so a later reviewer can see what was protected, repeated or referred.
See CSMLS equipment competencies, NL Health Services centrifuge verification, Eppendorf 5427 R instructions, Eppendorf Reference 2 instructions, NIST volume calibration procedures, PHAC spill response handbook, Canadian Biosafety Standard, third edition.
Two methods use 3,000 rpm. One approved rotor configuration has a radius of 10 cm and the other 15 cm. Why are their RCF readings different?
The timer lowers the speed when the carrier is heavy
At the same rpm, relative centrifugal force increases with rotor radius
A different rotor changes rpm into an elapsed-time setting
Tube mass determines the RCF displayed at a fixed radius and speed
An approved method calls for reverse pipetting with the Eppendorf Reference 2 to handle a viscous control. What should the operator do?
Aspirate from the first stop and blow out all liquid
Set a higher nominal volume to compensate without documenting it
Lay the pipette flat with liquid in the tip before delivery
Take up the extra blowout volume, dispense to the first stop and leave the excess in the tip
An operator suspects a tube broke inside a running centrifuge. What is the appropriate immediate response?
Use the instrument’s safe stop, keep the lid closed and notify the designated responder
Lift the lid to identify the broken tube
Reach into the rotor once the sound stops
Tag the run as acceptable and continue using the instrument