12.1 MLA Analytical Scope & Point-of-Care Screening Tests
Key Takeaways
Confirm your assignment, the approved method and the responsible reviewer.
Use the device’s specimen, collection, quality-control and cleaning instructions.
Treat timing, temperature and strip preparation as method-specific.
Verify the internal control at the specified reading time for the applicable cassette.
Communicate invalid, unexpected and urgent findings through the site pathway.
The February 2027 CSMLS profile includes principles of delegated analytical testing. An MLA performs the assigned task under the applicable supervision and procedures. A test run near the patient requires the same attention to identification, method suitability and documentation as work done elsewhere. Confirm the device, approved task, training, responsibility for verification and who receives results or exceptions.
Recognize Three Different Out-of-Process Findings
| Finding | Appropriate response |
|---|---|
| A required external control fails | Hold affected testing and results, preserve the controls and instrument messages, and notify the designated reviewer. |
| A patient test is invalid | Record the invalid event and follow the approved repeat or referral procedure with the correct specimen and supplies. |
| A valid result is flagged as urgent | Follow the lab’s immediate communication and result procedure. Any repeat or confirmation is governed by the applicable policy. |
The record can include identifiers, time, device, consumable lot and control status, result or error message, and whom the operator notified. Check the units the Canadian site’s device and record require. Where blood glucose is reported in mmol/L, enter or transmit the reading with those units. Apply the site’s critical and out-of-range rules. A meter message can mark the limit of an instrument’s reporting range and may require a distinct pathway from a numeric critical result.
Use a Professional Glucose Meter
Alberta Health Services describes an approved professional glucose meter for inpatient testing. A professional meter uses the specified test strips and quality control materials. Check the approved patient population, specimen and consumables before collection. Follow the device and collection procedure for skin preparation, drop selection and volume. Avoid forced squeezing or applying blood before the preparation is complete. If a specimen type or patient condition may invalidate a reading, seek the approved alternative from the clinical or laboratory team. An AHS adult inpatient order set notes that critically ill patients can have inaccurate capillary readings and gives a confirmation pathway.
Between patients, reprocess the meter as required by the facility and device instructions. Cleaning removes soil, and disinfection needs the approved contact time and product. Protect the test strip port. Use a single-use lancing device as required and dispose of it safely. If a required control fails immediately before a requested glucose measurement, report the equipment issue promptly and arrange the authorized backup pathway. A plausible number on the display cannot clear a failed control.
Read a Urinalysis Strip as Directed
Check the product, expiry, specimen stability and applicable temperature. Siemens’ Multistix instructions direct users to mix the specimen before testing and allow an appropriately refrigerated specimen to return to room temperature. Follow the local validated specimen procedure. Some workflows use well-mixed, uncentrifuged urine for the strip and reserve a separate preparation for sediment assessment.
Remove excess urine as the product directs. The Siemens instructions distinguish visual reading from readers such as the CLINITEK Status, where the strip edge is blotted. For visual work, compare each pad with its own colour field at the time given for that pad. An automated reader uses its programmed timing. A spoiled or expired strip, incorrect timing or sample contamination can affect the result. Use the insert to check a specific interference.
If a site changes from visual interpretation to an instrument, compare the two instructions for strip preparation and result entry. Record the method actually used. A questionable colour change in a contaminated specimen should prompt a documented check with the bench, rather than copying the closest colour or treating the result as a diagnosis.
Verify the Cassette Control
The ICON 25 urine hCG insert posted by Shared Health Manitoba uses a control region and a test region. Read according to the actual sample type and validated instructions. In that example, the urine read time is three minutes. A test-region band without a control-region band at the specified time is an invalid test. Record the invalid result and follow the method’s direction to repeat with a new device. If the problem persists, hold the testing supplies as required and escalate.
A procedural control checks aspects of that device’s run. Check any required external controls and acceptance criteria separately. Read and record the actual format and result permitted by the procedure, then pass the record to the authorized team.
Apply the Same Validity Checks to Rapid Antigen Screening
Use the product and workflow authorized for the setting. For example, the Flowflex professional COVID-19 antigen test instructions identify a missing procedural control line as invalid. Record that outcome and follow the approved repeat or referral procedure. A valid negative antigen result with ongoing symptoms may need reassessment under the site’s clinical process. The authorized reviewer decides what follow-up is required.
See CSMLS analytical competency 4.1, AHS adult glycemic management, AHS DKA order set example, Roche Inform II operator resources, Shared Health published Siemens test strip insert, Siemens Multistix insert, Shared Health ICON 25 procedure, Health Canada posted professional antigen test instructions, and PHAC antigen test accuracy.
A site uses a Siemens Multistix method and has verified that a refrigerated urine specimen remains within its storage limit. What should the operator do before testing?
Read every pad at the same elapsed time
Let the urine reach the temperature specified in the insert and mix it as directed
Use sediment alone for the strip
Run the strip immediately after removing the cold specimen from storage
At the specified three-minute urine reading for ICON 25, a test band appears but the control band does not. What is the result?
Valid weak positive
Negative because the control is missing
Reportable if the band is darker later
Invalid. Document and repeat using a new device according to the procedure.
A professional glucose meter’s required liquid control fails just before an urgent patient test. What should the MLA do?
Hold use of that device and urgently arrange the authorized backup or escalation
Run the patient test and accept a plausible reading
Change the control range until the reading passes
Wait to tell the team until the end of the shift
Sections you finish are checked off in the contents.