9.3 Carbon Adsorption Systems: Granular Activated Carbon (GAC), Empty Bed Contact Time (EBCT), and Worker-Polisher Tanks
Key Takeaways
Granular activated carbon (GAC) reduces oxidizing disinfectants that can cause hemolysis if they reach dialysate.
For GAC under the CMS-incorporated RD52:2004 standard, the minimum iodine number is 900. Acid-washed carbon is recommended.
Each series GAC bed needs at least five minutes of empty bed contact time (EBCT) at maximum flow through the bed.
Divide the volume of media by the flow rate in compatible units. More flow means less contact time.
A total chlorine result at or above 0.10 mg/L after the first bed requires testing after the second bed.
9.3 Carbon Adsorption Systems: Granular Activated Carbon (GAC), Empty Bed Contact Time (EBCT), and Worker-Polisher Tanks
Why Dialysis Water Needs Carbon Pretreatment
Water utilities can use free chlorine or chloramines to control microorganisms in drinking water. Water that meets drinking-water requirements can still injure patients if used to prepare dialysate. When oxidants reach the dialyzer, exposure can damage red blood cells and cause hemolysis. Free chlorine can also damage susceptible reverse osmosis (RO) membranes. Carbon treatment protects the downstream purification system as well as the patient.
Granular activated carbon (GAC) has a large reactive surface. It removes chlorine compounds through surface reactions and also adsorbs organic substances. Chloramine removal is slower than free chlorine removal, so both the type of media and its contact time matter. Higher pH, colder feed water, higher disinfectant loading and competing substances in the feed can reduce performance. This is why a carbon tank's age alone cannot determine whether it is still effective.
Media Selection and Replacement
For the GAC configuration described in CMS Appendix H, the minimum iodine number is 900. It is a media specification. Safe operation still depends on monitoring the installed beds with the actual water supply and flow. Acid-washed carbon is recommended to minimize aluminum leaching. Other carbon configurations require performance data for chloramine removal at maximum anticipated flow through the full interval between scheduled tests. Exhausted media is discarded and replaced with new media. Regenerated carbon is excluded.
Empty Bed Contact Time (EBCT): Principles and Calculations
Chloramine destruction depends directly on Empty Bed Contact Time (EBCT)—the nominal time it would take to pass a volume equal to the empty bed volume at the given flow. It is a design calculation based on the full media bed volume.
- GAC design under RD52:2004: At maximum flow through the beds, plan for at least 5 minutes per bed in series, for at least 10 minutes total.
- Dual-Tank Configuration: Two tanks piped in series:
- Primary (Worker) Tank: Minimum 5 minutes EBCT.
- Secondary (Polisher) Tank: Minimum 5 minutes EBCT.
With exactly five minutes per bed, the total is ten minutes.
[Softener] ──► [Worker Tank (5 min EBCT)] ──► [Port 1] ──► [Polisher Tank (5 min EBCT)] ──► [Port 2] ──► [Pre-RO]
Why the Two Beds Have Separate Sample Ports
The first bed removes most of the disinfectant load. The second provides a backup. The port between them is the routine test location because a detectable rise there gives time to test the backup before relying on it. If a carbon bank uses parallel trains, each independently fed series must be monitored at its own sample ports. A passing test in one train gives no information about another.
| Test location | What the result establishes |
|---|---|
| Port 1, between beds | At a total chlorine reading of 0.10 mg/L or more, test the outlet of the backup bed promptly. Port 1 alone cannot establish whether water passed the last bed within limits. |
| Port 2, after the backup | A reading clearly below the total-chlorine trigger can support short-term operation under facility policy while the first bed is replaced. An out-of-limit final reading requires immediate action to protect patients. |
Test before the day's first treatment and before each set patient shift. A facility without set shifts tests approximately every four hours. The next chapter explains test methods, documentation, and the facility response when either port is out of limits.
Backwashing and Monitoring
Backwashing in a system designed for it can redistribute the media and reduce channeling, in which water follows a path through the bed without using its full surface. Backwashing does not regenerate spent carbon. Follow the equipment and facility schedule and document the date of bed changes or backwashing checks. Review changing water demand with the responsible team because higher flow shortens contact time even when the tank size stays the same.
Worked Example: Sizing the Worker and Backup Beds
A facility is considering adding stations. The existing conventional water system has two series GAC beds, each with 10 cubic feet of media. The proposed maximum flow through the carbon beds is 18 gallons per minute. With approximately 7.48 U.S. gallons per cubic foot, each bed has about 74.8 gallons of media volume. Divide by 18 gallons per minute to get about 4.16 minutes per bed, or 8.31 minutes total before rounding. Either compare each bed with the 5-minute minimum or compare the two-bed total with 10 minutes.
To meet 5 minutes per bed at 18 gallons per minute, each bed needs at least 90 gallons of media, about 12.03 cubic feet before choosing an installed size. The current configuration could provide 5 minutes per bed at no more than about 14.96 gallons per minute. Bring this result to the medical director and water-treatment team for review before adding the demand. A favorable chlorine sample at today's lower demand cannot establish adequate contact time at the proposed higher demand.
Sources: 42 CFR § 494.40 and CMS Appendix H, revision 246, V192–V197.
Which specification applies when replacing granular activated carbon in a conventional U.S. dialysis water system governed by CMS Conditions for Coverage?
Use regenerated media once a backwash has been completed.
Choose any carbon by particle size alone and rely on reverse osmosis to remove the remaining chloramine.
Use new granular activated carbon with an iodine number of at least 900. Acid-washed carbon is recommended.
Replace the exhausted carbon with sodium-exchange softener resin.
A conventional two-bed granular activated carbon system has a maximum flow through the beds of 15 gallons per minute. To meet an empty bed contact time of at least five minutes per bed, what is the minimum volume of carbon media required in each bed?
15 gallons of carbon per tank
30 gallons of carbon per tank
50 gallons of carbon per tank
75 gallons of carbon per tank
A water treatment team proposes increasing flow through two unchanged carbon beds from 12 to 15 gallons per minute. Each bed contains 75 gallons of media. How does the change affect the empty bed contact time per bed?
It rises from 5.0 to 6.25 minutes.
It stays at 6.25 minutes because the beds are in series.
It drops from 12.5 to 10 minutes per bed.
It drops from 6.25 to 5.0 minutes per bed.