10.1 Total Chlorine and Chloramine Monitoring: Testing Schedule, Sample Ports, AAMI Limits, and Breakthrough Actions
Key Takeaways
Total chlorine includes both free and combined chlorine. A separate-testing strategy has to account for chloramine as well as free chlorine.
For routine manual checks, sample before the first treatment of the day and before each set shift, or approximately every four hours when shifts are not set.
When total chlorine at the first port is 0.10 mg/L or higher, promptly test the port after the backup bed and log both readings.
A compliant final sample can permit temporary operation while the first bed is replaced, up to 72 hours under RD52:2004 if the final port remains compliant.
An above-limit result after the last component requires immediate medical director notification and verified immediate correction or termination of affected treatments.
10.1 Total Chlorine and Chloramine Monitoring: Testing Schedule, Sample Ports, AAMI Limits, and Breakthrough Actions
Free Chlorine, Chloramines, and Total Chlorine
These procedures describe conventional two-stage water systems under § 494.40(a) and (b). For preconfigured systems, follow the FDA-approved labeling and requirements in § 494.40(e).
Water utilities may use chlorine or monochloramine to control microorganisms. Free chlorine refers to the uncombined chlorine species available to disinfect. Combined chlorine includes chloramines, which form when chlorine reacts with ammonia. Total chlorine measures the two fractions together. A free-chlorine test alone cannot show whether chloramine remains in the water.
Dialysis water has different safety requirements from drinking water because the water is used to make dialysate. Chloramine exposure during dialysis can cause oxidative injury to red blood cells and hemolysis. Oxidation of iron in hemoglobin also produces methemoglobin, which cannot carry oxygen at the affected sites. Suspect a shared water or dialysate problem when unexplained symptoms occur in several patients using the same supply. Alert the nurse and activate the facility's response. A water test establishes exposure risk. Clinical assessment determines whether an individual patient's symptoms are due to hemolysis or another cause.
Choosing a Test and Collecting a Sample
The federal rule gives different first-port triggers for different tests. If measuring the fractions separately, test the second bed when free chlorine exceeds 0.5 mg/L or chloramine exceeds 0.1 mg/L. With a total-chlorine test, test the second bed at 0.10 mg/L or higher. A total-chlorine test is a common way to screen both fractions with one measurement. For example, a validated free-chlorine result of 0.02 mg/L and a total-chlorine result of 0.18 mg/L from the same sample imply 0.16 mg/L combined chlorine. The total-chlorine result alone already triggers the second-port test.
CMS Appendix H describes N,N-diethyl-p-phenylenediamine (DPD) test kits and dip-and-read strips, as well as suitable online monitoring. Choose the facility-approved method with the sensitivity needed for the applicable threshold. A strip whose first readable step above zero is 0.5 mg/L cannot resolve a 0.10 mg/L threshold. Follow the method instructions for sample volume, calibration or controls, and reading time.
- When: For routine manual monitoring, test before the first treatment of the day and before each set patient shift. When there are no set shifts, test approximately every four hours. Apply the facility's verification and calibration policy when an online monitor with automated alarm is in use.
- Where: The routine sample comes from the port after the first bed and before the backup. If it reaches the applicable trigger, take a separate sample from the port after the backup.
- How: Draw after the system has operated for at least 15 minutes. Flush the sample port according to the facility's procedure and analyze on-site promptly. Record the time, port, method, result and any notifications or corrective action. Distinguish the actual time of a follow-up test from a generic shift entry.
Decision Path After the First Port Reaches the Total-Chlorine Trigger
First bed breakthrough, compliant backup. Log the actual test times and call the designated clinical and technical staff. Under RD52:2004, a compliant backup bed may support a short period of continued operation, up to 72 hours, while the spent bed is replaced, provided the second port remains compliant. Follow the facility plan for testing after the backup. CMS explains that an increase in testing frequency depends on past performance and changes in the feed water. When possible, move the existing second bed to the first position and place the new bed second. If the positions cannot be exchanged, replace both beds.
Final reading above limit. Immediately notify the medical director. Take immediate corrective action and verify by testing that it was effective, or terminate affected treatments to protect patients from exposure. Follow the facility emergency procedure to keep suspect water out of use and report possible exposure to the clinical team. Water in a holding tank can be considered only if appropriate for that system and tested compliant. Prevent incoming unsafe water from reaching that tank. Before resuming, confirm acceptable levels and document steps to address the cause. Decisions about patient evaluation, returning circuit blood or rescheduling are made by the clinical team using the applicable emergency protocol.
Worked Scenario: Following Both Readings to a Decision
Before the next scheduled shift, a technician tests after the first bed and obtains 0.13 mg/L total chlorine. The water system is running and the test method is appropriate. The technician promptly tests after the backup, which reads 0.02 mg/L. The results show a first-stage problem with a compliant final sample. The technician records both readings and the time, contacts the designated staff, and follows the replacement and monitoring plan. The next backup test reads 0.14 mg/L before the first bed has been replaced. That is an above-limit final reading. The facility must now protect patients, notify the medical director immediately, and confirm effective correction by testing or terminate affected treatments. The earlier 0.02 mg/L reading cannot establish safety after the new result.
Sources: 42 CFR § 494.40(a)–(b), CMS Appendix H, revision 246, V196–V197 and V270–V273, CDC investigation of dialysis-associated hemolysis, and case series on dialysis-associated methemoglobinemia.
Why does a free-chlorine result of 0.02 mg/L by itself fail to clear a water sample from the first carbon port?
The test does not measure chloramines, which can injure red blood cells during dialysis.
A free-chlorine test measures the chlorine adsorbed inside the bed instead of the water sample.
The test cannot be conducted in a dialysis facility.
The reported unit cannot be used for a chlorine result.
Several patients dialyzing on a shared water system develop unexplained symptoms. The clinical team identifies hemolysis and methemoglobinemia, and the water investigation finds chloramine breakthrough. Which mechanism fits these findings?
Chloramine raises hemoglobin's oxygen-carrying capacity.
Chloramine causes oxidative damage to red cells and oxidizes ferrous iron in hemoglobin to ferric iron.
Chloramine causes hemolysis only by making dialysate hypotonic.
Chloramine triggers the blood pump to draw air into the venous line.
A conventional dialysis facility records 0.14 mg/L total chlorine after the first carbon bed, then 0.02 mg/L after the backup. What is the appropriate interpretation and action?
The first reading proves all patients were exposed and every circuit has to be discarded.
The low second reading proves the first reading was inaccurate. Follow the normal schedule and make no record.
The backup is compliant at this check. Log both values, notify designated staff, and follow the temporary-operation, follow-up testing and replacement plan.
Exchange the tank connections with no further testing and use the original first bed as backup.