12.2 Blood Leak Detectors, Air Bubble Detectors, and Temperature Alarms: Mechanism and Automated Bypass

Key Takeaways

  • The dialysate blood-leak detector checks for blood on the spent dialysate side. A detector alarm may have other causes, so use the applicable verification procedure.

  • A positive blood-leak result calls for the facility's blood-leak response and clinical assessment. Use the machine and dialyzer instructions when deciding how to handle circuit blood.

  • On the 2008T the ultrasonic level detector monitors the venous chamber. The separate optical detector identifies blood versus clear fluid in the return tubing.

  • When the level alarm detects air at a low chamber level, the 2008T stops blood pumping and occludes the venous line.

  • During dialysate bypass, fluid is routed away from the dialyzer. Assess the patient and confirm blood pump and UF status separately.

Last updated: November 2026

Identify what triggered the monitor

MonitorWhat it observesWhy it matters
Dialysate blood leakAn optical change in fluid leaving the dialyzer on the dialysate sideMay indicate a leak across the membrane. Optical interference can also produce an alarm.
Venous chamber level or air monitorAir or a drop in fluid level at the detector locationThe model can stop the pump and close the venous safety clamp to limit passage of air.
Optical blood detector in the venous tubingOn the 2008T, whether opaque blood or clear fluid is present below the clampUsed for the device's alarm logic. It does not replace the ultrasonic chamber monitor.
Temperature and conductivity monitorsThe measured conditions of the prepared dialysateThe machine can keep fluid out of the dialyzer when it goes into bypass. The criteria depend on settings and the model.

Verify pre-use checks

Before connection, verify that the 2008T Rev. AF alarm and hydraulic pressure tests have passed. The test sequence depends on the startup and prime mode. Confirm successful checks for the blood-leak, level and pressure monitoring functions. During the specified final-setup level check, confirm that the pump stops and the venous clamp occludes. A test that does not pass calls for holding the machine from use and following the model’s service procedure. Do not use the absence of an alarm during setup as proof that the alarms have been tested.

Investigate a possible blood leak

Assess the patient and identify the exact blood-leak message. The 2008T Rev. AF describes optical alarms caused by blood or interference such as air. It specifies testing dialysate with a compatible blood-leak strip and repeating a negative result as directed for the applicable alarm. Read the strip according to its instructions. Clear-looking fluid alone cannot exclude a leak. One negative result cannot resolve a recurring alarm. Do not resume clinical use while verification or troubleshooting remains incomplete.

A positive result supports the presence of blood on the dialysate side. A breach can cause blood loss or expose the blood side to dialysate. Notify the RN and follow the established blood-leak procedure. The F180NR Rev. E labeling refers staff to facility protocol, and the 2008T Rev. AF directs them to the unit's blood-leak policy on a positive test, including whether return is allowed in that situation. The clinical team also checks for other hazards that can make reinfusion unsafe. Do not independently override an alarm to rinse back. Stop use of the affected dialyzer and record the event, verification and disposition.

Respond to air, temperature and conductivity alarms

Check that the drip chamber and bloodline are seated in the correct holders, the chamber is filled to the specified level and the circuit is free of air. On the 2008T, when low chamber level and air are detected, the blood pump stops and the venous line clamp closes. Check the patient, trace for air or a loose connection and follow the precise alarm procedure. If air may have been infused, keep air-contaminated blood from being returned and activate the emergency response. A corrected equipment alarm in a patient with no suspected air exposure is a different scenario.

Dialysate temperature and conductivity are checked against the specified values and limits. Hypotonic or excessively hot fluid may injure blood cells, but an alarm value does not diagnose an injury. In 2008T bypass, fresh dialysate no longer goes through the dialyzer. A separate blood-side alarm can stop the pump. Verify the displayed status of blood flow and UF rather than assuming they continue or that the alarm establishes patient safety. Assess the patient and follow the instructions for the specific message. If there is possible unsafe exposure or an unresolved fault, notify the RN and qualified technical staff.

Worked example — clear fluid with a positive strip

During treatment the 2008T displays a blood-leak alarm, and the effluent looks clear. An appropriate blood-leak strip is positive under its reading instructions. Treat the finding as a potential membrane breach. Alert the RN, maintain the protective alarm response while the team applies the facility's positive-test protocol, and avoid restarting treatment on the affected dialyzer. Document the actual result, actions and clinical decision about circuit blood. Use the patient's observed blood loss and circuit disposition, rather than an assumed volume, when recording the event.

Sources 2008T operator's manual, Rev. AF, pp. 43–44, 48, 74–80, 224–225 and 251–252, Optiflux F180NR instructions, Rev. E and CDC investigation of dialysis-associated hemolysis.

Test Your Knowledge

A 2008T blood-leak alarm persists. The effluent is visually clear, but an appropriate blood-leak strip is positive. What is the appropriate next step?

A

Resume dialysis because visible red effluent is required.

B

Report the positive finding to the RN and follow the unit's blood-leak policy and both device instructions before deciding what to do with circuit blood.

C

Switch to bypass and independently rinse back.

D

Change the blood-leak threshold and continue on the same dialyzer.

Test Your Knowledge

On the 2008T, what does the venous level detector module do when a low chamber level and air are detected?

A

It uses the conductivity cell to raise dialysate sodium.

B

It reverses the blood pump to clear the chamber.

C

It alarms, stops the blood pump and closes the venous safety clamp.

D

It relies on the optical blood detector to measure the pH of the return path.

Test Your Knowledge

A 2008T enters dialysate bypass after an out-of-range conductivity alarm. What does bypass establish?

A

Dialysate is routed away from the dialyzer. Check the patient and separately verify blood pump and UF status.

B

The blood circuit is automatically safe to continue unattended.

C

The machine has determined the patient's serum sodium concentration.

D

The previous treatment time lost in bypass has already been made up.