11.1 Extracorporeal Circuit Components: Arterial Line, Blood Pump, Drip Chambers, and Dialyzers
Key Takeaways
Arterial and venous name the blood removal and return limbs of the circuit. Identify pressure sensor locations before interpreting a reading.
A roller pump compresses a specified segment of tubing. Match the pump setting to the approved segment and avoid kinks.
Blood flows inside hollow fibers and dialysate flows outside them. The fiber walls and seals keep the streams separated while allowing transfer across the membrane.
Prime to remove air and prepare the dialyzer and lines using the actual device instructions. Confirm the correct orientation, required volume, fill level and absence of air.
Verify the prescribed dialyzer and inspect its label and integrity. Reprocessing programs must use eligible devices on the same patient under the applicable reuse checks.
Follow the blood path
| Component | Function and what to check |
|---|---|
| Arterial line and its pressure monitor | The arterial line brings blood from the access to the pump. A sensor upstream of the pump usually reads suction when the pump is drawing blood. A configured post-pump arterial sensor measures a different site. |
| Roller pump and pump segment | Rollers move blood through compatible tubing. The machine's flow display depends on setup and operating conditions. Check the approved tubing, pump segment size, loading and routing. |
| Dialyzer blood compartment | Blood is distributed through a header into hollow fibers and collected at the opposite header. Dialysate passes on the outside of the fibers. |
| Venous drip chamber and level detector | The chamber separates air from the returning blood. In the 2008T example, the ultrasonic detector monitors chamber level, and the venous pressure port measures pressure at the chamber. |
| Venous return, clamp and optical detector | Check the path to the patient and keep it visible. On the 2008T, the optical detector distinguishes blood from clear fluid in the return tubing. The clamp and pump are used by the machine's protective alarm logic. |
A clamp or kink can compromise flow, and a loose connection can cause air entry or blood loss. Identify the actual location instead of using fixed arterial and venous readings for every setup. Trace the whole blood path during preparation, and again if the machine alarms.
Understand the pump and dialyzer together
At the same roller speed, blood moved per revolution depends on the pump segment and its filling. Incorrect loading or tubing diameter can make actual delivery differ from the displayed setting. Excessively restricted blood flow can damage red cells. An unexpected decline in delivered dialysis dose calls for a review of treatment time, actual flow, access and sampling as well as equipment. A pressure reading within the patient’s usual pattern does not exclude access dysfunction. Remove a suspected faulty machine from clinical use and refer pump verification or calibration to qualified personnel.
A hollow-fiber dialyzer uses sealed fiber bundles to keep blood inside the lumens while dialysate passes around them. The F180NR Rev. E labeling is one example with a polysulfone-based membrane, 185 µm fiber lumen and 35 µm membrane wall, with polyurethane potting. Device specifications vary. Hydraulic permeability, expressed by the ultrafiltration coefficient, describes water transfer under specified conditions. The device's solute clearance and sieving data describe different properties. A more permeable membrane can help with removal of solutes such as beta-2 microglobulin, depending on the membrane and treatment conditions. Use the prescribed dialyzer and compatible UF control.
Prepare and verify before connection
Examine the package, dialyzer label, ports and caps. Set up compatible tubing with aseptic technique, verify the labels and secure connections. Prime and flush the blood side to remove air and reduce residues, and fill the dialysate side as specified. For example, the 2008T setup shows arterial-end down for blood-side priming. The F180NR Rev. E instructions then rotate the dialyzer arterial-end up to fill the dialysate compartment, and separately direct staff to the compatible bloodline instructions for treatment orientation. Check the drip chamber fill and confirm the circuit is free of visible air and is ready before connection. Do not infer a saline volume from another dialyzer's label or infuse recirculated prime.
Quality control when a facility reprocesses dialyzers
Use only eligible devices labeled for multiple reuse, only for the same patient and with the required exclusions. The F180NR label used here is single-use. A reuse program follows the incorporated ANSI/AAMI RD47:2002/A1:2003 requirements and the applicable device and germicide instructions. Water at each reprocessing system is sampled under the approved water monitoring plan. Validate automated volume measurements and verify function after applicable repairs or changes before releasing equipment to service. Service and reprocessing records identify the equipment and the checks completed.
For hollow-fiber dialyzers using a total cell volume (TCV) performance test, the baseline is the original measured volume before first use. Between uses, require at least 80% of original TCV and a separate blood-path integrity test. At an original 100 mL and a current 78 mL, the result is 78%, so discard under this criterion even if the integrity test passed. Meeting the volume threshold does not substitute for a passed integrity test.
At release, check identity, labeling, caps, damage, leaks and storage conditions, and verify germicide presence as required before rinsing. After the approved rinse, use the required low-level residual test before treatment. Product-specific presence testing and residual testing serve different purposes. Record results and who performed the checks. CMS Appendix H describes a two-person identity check before use in a facility, with patient participation if possible.
Worked example — a flow discrepancy
A patient’s reported Kt/V fell from 1.45 to 1.12 with a prescribed blood flow of 400 mL/min. The displayed flow and pressure trends looked usual. The RN checks the delivered treatment time, access findings and sample collection, and the technical team verifies the removed machine on a test circuit. At a displayed 400 mL/min, the test setup measures 312 mL/min. The difference is 88 mL/min, or 22% of the setting. Verify the compatible pump segment and measurement setup, and report the bench result to qualified staff. This finding supports investigating delivery. It cannot establish how much blood flow the patient received or account for the entire Kt/V change by itself.
Sources 2008T operator’s manual, Rev. AF, Optiflux F180NR instructions, Rev. E, FDA Optiflux technical filing K203062, 21 CFR §876.5860, and CMS Appendix H, V302–V303, V314–V317, V330, V336–V338 and V347–V353.
During setup, the pump-segment diameter displayed on the 2008T differs from the diameter specified for the approved tubing in use. What should the operator do before connecting the patient?
Verify the tubing and correct the setup under the device instructions so the selected diameter matches the pump segment.
Proceed because the venous pressure monitor independently measures blood flow.
Adjust the mechanical occlusion by trial and error until both values match.
Use the dialysate conductivity display to verify the pump segment size.
What do the headers, fibers and potting in a hollow-fiber dialyzer accomplish?
They mix blood directly with dialysate and then re-separate both liquids.
They direct blood through the fiber lumens and keep the flow paths separated while allowing transfer across the fiber walls.
They set the exact same ultrafiltration coefficient for all high-flux models.
They produce fluid removal by pumping dialysate through the blood tubing.
An eligible dialyzer in a reuse program has original TCV of 110 mL, current TCV of 90 mL and a failed blood-path integrity test. What should staff do?
Release it because more than 80% of the original TCV remains.
Replace the identity label and release it without further review.
Keep it out of use because passing the volume test cannot compensate for a failed integrity test.
Raise machine venous pressure to overcome the defect.