5.4 Severe Technical Emergencies: Air Embolism, Hemolysis, and Exsanguination

Key Takeaways

  • Loose prepump connections, residual air after inadequate priming or an open catheter path can admit room air. Inspect the circuit and maintain its air safeguards.

  • If air may have reached the patient, activate emergency help, stop the pump, isolate the circuit and keep air-contaminated blood from being returned.

  • For life-threatening gas embolism, the AHA recommends 100% oxygen and supports level positioning. Start CPR and use an AED as indicated.

  • In suspected acute dialysis-associated hemolysis, stop treatment, clamp the circuit and do not return its blood. Hemolysis releases potassium and free hemoglobin.

  • Venous needle dislodgement can escape machine pressure detection. Keep access sites and connections visible and secure. For major bleeding, stop pump-driven loss, apply direct pressure to the site and summon emergency help.

Last updated: November 2026

Suspected air embolism

Air can enter through a loose connection on the suction side of the blood pump, through an open central venous catheter, or when residual air is left in tubing during setup. In the circulation, gas can obstruct flow and cause acute shortness of breath, chest discomfort, hypotension, neurologic changes or collapse. Visible air in the return line or an observed connection failure can be critical evidence. Recognition does not depend on hearing a murmur.

If air may have reached the patient, call for emergency help and the RN while stopping the pump and clamping and isolating the affected bloodlines or access under the emergency procedure. Do not return blood from the air-contaminated circuit. This situation differs from a detected alarm addressed before any air reaches a stable patient. Assess the patient even if the alarm clears.

For life-threatening gas embolism, the 2025 AHA guideline recommends 100% oxygen. It supports a level surface with the patient supine or in a recovery position as appropriate to breathing and airway status. If CPR is indicated, position supine and begin resuscitation without delay. Older teaching on routine head-down positioning has been superseded in this guidance. Emergency clinicians may arrange hyperbaric oxygen. The technologist helps with the immediate response within training and local scope.

Suspected acute hemolysis

Red-cell rupture releases free hemoglobin and potassium. Potential causes during dialysis include severe flow restriction from kinked or defective tubing, hypotonic or overheated dialysate, and contaminants such as chloramine or residual disinfectant. Loss of the expected pressure pattern can help locate a problem, but a single pressure, conductivity or temperature value cannot diagnose hemolysis. Chest or back pain, dyspnea, nausea, unusual translucent red or burgundy circuit blood, or pink plasma can raise concern. Neither normal color nor the absence of an alarm rules it out.

Stop dialysis immediately, stop the pump, clamp and isolate the circuit, and do not reinfuse potentially hemolyzed blood. Call the RN and emergency team. Reinfusion may add potassium released by ruptured cells. The team assesses and monitors for hyperkalemia, rhythm changes and anemia. Remove the affected equipment from use under the site's procedure. Preserve relevant tubing, supplies and lot information for investigation and let the clinical team direct treatment and testing. If more than one patient is affected, promptly report the possibility of a shared supply, water or dialysate problem.

Major bleeding from disconnection or needle dislodgement

An open venous return path can discharge pump-driven blood. A venous pressure alarm can miss dislodgement because the pressure changes by less than the alarm window. Measured pressure still includes resistance from the tubing and needle, as well as differences related to height and pump flow. Keeping the patient's access and connections visible allows staff to notice blood loss before a monitor does. Secure lines and needles according to the applicable device and facility procedure and check them when the patient moves.

If disconnection or dislodgement is causing serious bleeding, summon help, stop the blood pump, clamp open bloodlines and apply firm direct pressure to the bleeding access site using protective equipment as available. Act in parallel when additional staff are present. Assess for shock, support emergency care and record the observed time and equipment condition.

FindingPriority circuit actionCare-team concern
Air may have reached the patientStop the pump and isolate the air-contaminated circuit. Withhold reinfusion.Support oxygenation, circulation and urgent evaluation for gas embolism.
Possible acute hemolysisStop dialysis, clamp the circuit and withhold reinfusion.Assess potassium, rhythm, anemia and the possible source.
Dislodged needle or separated line with major blood lossStop the pump, clamp and apply pressure to the bleeding site.Control blood loss and evaluate shock.

Worked example — sudden pain and an abnormal bloodline

Ten minutes into treatment at a prescribed blood flow of 420 mL/min, Donald reports severe back and chest pain. The technologist observes an unusual translucent red color in the venous tubing and a new kink between the blood pump and dialyzer. That combination raises concern for mechanical hemolysis. The technologist stops the pump, clamps the lines, withholds reinfusion and calls the RN and emergency team. A patient blood sample later shows elevated potassium. The team assesses and treats the patient and retains the circuit and machine information for investigation. A kink supports a mechanical mechanism, yet its presence alone does not prove it was the only cause.

Sources 2025 AHA resuscitation guidance, gas embolism, Optiflux F180NR instructions, Rev. E, 2008T operator's manual, Rev. AF, CDC hemodialysis hemolysis investigation, BC Renal complication protocol, clinical context, needle dislodgement simulation and CMS Appendix H, access visibility.

Test Your Knowledge

A patient develops abrupt chest and back pain during hemodialysis. The venous line looks unusually translucent red and a bloodline is kinked. What is the immediate response to suspected hemolysis?

A

Straighten the tubing and return the circuit blood to avoid anemia.

B

Continue treatment if there is no blood leak alarm.

C

Stop dialysis and the blood pump, clamp the lines, withhold reinfusion and call for emergency clinical help.

D

Increase blood flow to determine whether the color clears.

Test Your Knowledge

Air is seen moving toward the patient through the return line. The patient becomes acutely dyspneic but remains responsive and breathing. Staff stop the pump, isolate the circuit and call for emergency help. Which additional measure follows the 2025 AHA gas embolism guidance?

A

Stand the patient up to redistribute the gas.

B

Provide 100% oxygen and support the responsive patient supine on a level surface under the emergency protocol.

C

Place the patient routinely head-down before supporting oxygenation.

D

Restart the pump at lower speed while waiting for the team.

Test Your Knowledge

A venous needle has dislodged. The dressing is visibly bloody, yet venous pressure has not crossed the low alarm limit. Which explanation and preventive measure fit?

A

An intact air detector prevents venous blood loss, so observation can be deferred.

B

A quiet alarm establishes that the visible blood is only from the skin.

C

Increasing blood flow is the usual way to verify that the needle remains in place.

D

The pressure change can stay within the alarm window. Secure the needle and keep the access and line connections visible during dialysis.