11.2 Dialysate Delivery System: Proportioning, Acid & Bicarbonate Concentrates, Conductivity, and pH Monitoring
Key Takeaways
Match concentrate composition and proportioning ratio to the prescription and machine configuration. Verify the label instead of relying on connector color.
The acid component helps limit carbonate precipitation. The bicarbonate component provides buffer.
Observe the labeled preparation, storage and discard time for bicarbonate concentrate. Follow the facility's distribution and sanitation controls.
Conductivity measures aggregate ionic conductance. Check it against the applicable expected value with the appropriate temperature basis; approximate pH is a separate check.
Unexpected readings need verification and correction before connection. A dialysate bypass alarm requires attention to the patient's condition and the model's instructions.
Mix the prescribed fluid
A conventional bicarbonate system combines treated water, an acid component and a bicarbonate component. Metering establishes the mixture. On the 2008T, balancing chambers control the hydraulic balance of dialysate and the UF system handles net fluid removal, while concentrate pumps provide the concentrate for proportioning. Confirm all components and the prescribed settings. A 1:44 acid labeling example means one volume of acid concentrate for 44 volumes of the other streams combined. Thus 45 L of final mixture contains 1 L of that acid concentrate under the nominal ratio. Bicarbonate and water have their own specified contributions. Verify the exact configuration and concentrate label rather than applying one recipe across machines and products.
| Component | What to verify |
|---|---|
| Acid concentrate | Confirm species and electrolyte content, formula, ratio, product identity and whether additives are prescribed. The acid helps limit calcium and magnesium carbonate precipitation after mixing. |
| Bicarbonate concentrate | Verify the compatible product, dated preparation and expiration or use time. Bicarbonate provides buffer and liquid concentrate can support microbial growth. Follow the product's allowed holding time and the facility's cleaning and handling procedures. |
| Treated water | Confirm supply from the facility's approved system. Acceptable final conductivity does not replace chemical or microbial water controls. |
| Final dialysate | Match the prescription and selected concentrate. Independently check the required conductivity and approximate pH by the approved method. |
Potassium and calcium values on the prescription are individual concentrations. Reconfirm an unfamiliar or substituted product with the nurse. A mixture with the wrong potassium or another ingredient can have a plausible total conductivity. The nurse and prescriber establish the bath. Do not choose an electrolyte formula on the basis of a single laboratory value.
Verify conductivity and approximate pH
Conductivity describes how readily the dissolved ions conduct current. Temperature affects the reading. Use the device and meter instructions for temperature compensation and compare values on the same basis. The expected conductivity depends on the selected composition and settings. It is distinct from the limits of the alarm window and from the actual machine reading. Approximate pH tests a different property. Check it independently even when the conductivity is plausible. Also verify the prescribed dialysate temperature. On the 2008T Rev. AF, compare the displayed actual temperature with the ordered setting after the temperature stabilizes. If the device has a temperature alarm or remains in bypass, hold connection and escalate the failed check. Follow the applicable device and facility instructions for troubleshooting or an independent temperature check.
CMS Appendix H V250 calls for independent approximate pH before each patient's treatment and when a different acid composition is used. Follow the machine manufacturer's conductivity instructions and the facility's established limits for comparison with external instruments. For the 2008T Rev. AF setup, independent means can include the machine's independent conductivity test, distinct from the ordinary displayed value. Where a meter or strips are used, the instructions specify pH 6.9–7.6 and conductivity within 0.4 mS/cm of the theoretical conductivity (TCD). Follow the exact device and facility procedure for pH, including any provisions for the bibag setup and concentrate changes. Record the verification.
The machine's bypass function can route dialysate away from the dialyzer in an out-of-range condition. Assess the patient and the actual alarm, and do not use bypass clearance as evidence that the correct bath has been mixed. Follow the manufacturer and facility steps if the unit cannot be verified or if unsafe fluid might have reached the patient.
Worked example — two out-of-range checks
A 2008T has a TCD of 13.9 mS/cm for the ordered bath. The independently verified sample reads 13.1 mS/cm and pH 6.7. Conductivity is 0.8 mS/cm below TCD, exceeding the model's 0.4 mS/cm allowance, and pH is below the device's lower bound. Do not connect the patient. Report both results and verify the meter, sampling, concentrate labels and machine configuration. The team finds a mismatched concentrate. After correction under the device and facility procedure, a fresh sample reads 13.8 mS/cm and pH 7.2. The conductivity difference is 0.1 mS/cm. Both results meet the device criteria in this example. Document the correction and finish the remaining pretreatment and prescription checks before use.
Sources 2008T operator’s manual, Rev. AF, pp. 17, 72, 80–81, 94–95 and 317, 2008T technician’s manual and CMS Appendix H, V233, V249–V250.
Why does a conventional bicarbonate dialysate system use a separate acid concentrate?
To determine the patient's ultrafiltration target from the supply pressure.
To sterilize the dialysate and eliminate the need for water quality controls.
To replace prescribed anticoagulation in the blood circuit.
To support the final dialysate pH and reduce carbonate precipitation when calcium and magnesium are present.
On a 2008T with TCD 14.0 mS/cm, the independent sample is 13.7 mS/cm and pH 6.7. What should staff do before connecting?
Withhold treatment and investigate. The conductivity is within 0.4 of TCD, but the pH fails the device check.
Start dialysis because a conductivity within 0.4 mS/cm proves pH is acceptable.
Start treatment because pH cannot be checked independently before connection.
Widen the conductivity alarm window and omit the pH result from the record.
Two acid concentrates have the same dilution ratio but list different potassium concentrations. The delivered conductivity is within the expected range. How should staff verify the bath?
Use either product because conductivity identifies each dissolved electrolyte.
Match the connector color to the machine and omit the label check.
Match the actual product composition and configured recipe to the prescription and complete the applicable fluid checks.
Use the product with the lowest potassium whenever the patient's dialysis time is shortened.