6.3 Blood Sampling Technique: Post-BUN Collection Protocols and Avoiding Recirculation Rebound
Key Takeaways
Confirm the order, the patient's identity, the specimen type and the collection time. Obtain the pre- and postdialysis BUN from the same treatment.
Collect the predialysis specimen before treatment and protect it from saline, heparin or catheter lock contamination. Use the catheter and facility instructions for any discard.
For KDOQI's slow-blood-flow method, use 100 mL/min and wait long enough for less-recirculated blood to reach the inflow sampling port, typically 15 seconds. The stop-dialysate-flow method is a separate alternative.
A low post-BUN caused by recirculation or saline dilution can inflate calculated adequacy. A delay with tissue rebound can bias in the opposite direction.
Quick Answer: For a paired BUN calculation, obtain the predialysis specimen before dialysis begins and collect the postdialysis specimen from the dialyzer inflow side at the approved sampling time before rinseback. KDOQI describes a slow-blood-flow method and a separate stop-dialysate-flow method.
Confirm the Order Before Collecting
Confirm the patient's identifiers, specimen type, ordered collection time and any requirements for the container, labeling, storage or prompt transport. Record the site and timing when requested. Predialysis, intradialytic and postdialysis results can answer different questions. For example, an ordered blood culture after a new fever requires the facility's culture and aseptic collection procedure. A timed urine specimen, if ordered to assess residual kidney function, uses the ordered collection interval and laboratory handling instructions.
Collect the predialysis BUN before starting dialysis in the same session as the postdialysis sample. Protect the blood specimen from saline and heparin contamination. When sampling from a central venous catheter, perform hand hygiene, clamp before removing the cap, use an approved antiseptic with friction on the hub and let it dry. Clear the lock and any required discard using the catheter-specific and facility procedures for the ordered specimen. The catheter model, lumen and test can affect this procedure.
Obtain the End-of-Treatment BUN Before Rinseback
- Identify the approved method and sampling site. KDOQI describes collecting from the dialyzer inflow side with either slow blood flow or stopped dialysate flow. Verify the correct inflow port on the tubing set.
- Use the method's timing. With the slow-blood-flow approach, reduce pump flow to 100 mL/min and wait long enough for blood from the access to pass the sampling port, usually 15 seconds. At that flow, about 25 mL passes in 15 seconds. The volume upstream of the port depends on the tubing. This change minimizes access recirculation for the draw. The separate stop-dialysate-flow method allows about 3 minutes with dialysate flow stopped before sampling, so the BUN concentrations of the circuit inflow and outflow become closer.
- Collect and document promptly before rinseback. Use the facility protocol and machine and tubing instructions for ultrafiltration, pump and clamp handling, aseptic port access and specimen collection. Label the sample as postdialysis and document a deviation for the team and laboratory.
Recognize the Direction of Sampling Error
| Collection problem | Likely bias and mechanism |
|---|---|
| Full-speed sample with known access recirculation | Recently dialyzed blood recirculates into the withdrawal line. It can lower measured post-BUN and overestimate URR and spKt/V. |
| Saline enters the postdialysis sample | Dilution can lower measured post-BUN and overestimate calculated adequacy. |
| Sample from the outlet while the dialyzer is still clearing urea | Urea concentration in the dialyzer outflow can be lower than at the inflow. The outlet sample can overestimate the patient's urea reduction. |
| Delayed collection after clearance stops, without dilution | Blood urea can increase as solute redistributes. Using the later value in an end-of-treatment calculation can underestimate spKt/V or URR. |
| Predialysis sample diluted by saline or lock solution while post-BUN is valid | A falsely low baseline BUN makes the apparent reduction smaller and can underestimate calculated adequacy. |
Access recirculation brings already dialyzed blood back to the withdrawal line. Cardiopulmonary recirculation and later tissue-to-blood redistribution can also change the concentration measured near or after treatment completion. Those changes explain why the intended sampling time matters. The extent and duration of rebound vary among patients and sessions.
Worked Case: Two Errors in One Sample
The predialysis BUN is 68 mg/dL and the postdialysis result is 33 mg/dL, an apparent URR of . The postdialysis sample was drawn late from the circuit after saline rinseback had begun. Delayed redistribution can raise BUN, and saline contamination can lower it. The recorded sequence leaves the direction of the net error undetermined.
The technologist reports the collection problem. The care team arranges a repeat valid pair at a subsequent session. The new pre-BUN is 70 mg/dL and the post-BUN is 20 mg/dL, giving URR ≈ 71.4%. The repeat pair estimates the BUN reduction for that session. The earlier biased pair cannot establish the earlier dose.
Sources KDOQI 2006 sampling guideline § 3.1–3.4, NKF materials explaining the 2015 adequacy update, HEMO study of rebound, CDC catheter hub protocol, CDC blood culture collection and 42 CFR § 493.1242 specimen handling.
The unit uses the KDOQI slow-blood-flow method for a postdialysis BUN. Which approach reflects that method?
Begin saline rinseback and collect from the dialyzer outlet.
Keep the blood pump at the full treatment rate and sample the circuit immediately.
Disconnect a bloodline and wait for tissue urea to equilibrate before taking the sample.
Reduce blood flow to 100 mL/min and allow the blood to advance past the inflow sampling port, usually about 15 seconds, then sample there before rinseback using the approved procedure.
In a session with confirmed access recirculation, a sample drawn from the dialyzer inflow at full blood pump speed contains recently dialyzed blood. With a valid predialysis BUN, how does this affect the calculated dose?
A falsely low post-BUN can make both URR and spKt/V appear higher.
A falsely high post-BUN can make both URR and spKt/V appear lower.
Only the predialysis BUN changes.
URR is unaffected because the calculation uses the dose prescribed by the clinician.
Which error can dilute a predialysis BUN specimen drawn from a central venous catheter?
Confirming the patient's identity before collecting the sample.
Using the correct tube for the requested BUN assay.
Collecting while lock solution remains in the sampled lumen.
Letting the approved catheter hub antiseptic dry before sampling.