12.3 Extracorporeal Clotting, Power Failure Protocols, and Manual Blood Return Safeguards

Key Takeaways

  • Fibrin, clots, unexplained pressure changes or a loss of clearance can suggest clotting. Inspect the circuit and assess alternative causes.

  • Stop a visibly occluded circuit and do not force blood through it. Notify the nurse and follow the disposition and documentation procedure.

  • Facility outage plans cover equipment limits, communication and patient disposition. An audible loss-of-power alarm does not establish that dialysis safeguards still work.

  • Use only the applicable manufacturer procedure and assigned clinical direction for manual return. The 2008T gives different instructions for manual circulation and manual return.

  • When manual return is permitted, directly observe the venous chamber and return line for air. Keep air-contaminated, hemolyzed or fully clotted blood from being returned.

Last updated: November 2026

Recognize clotting and protect the return path

Blood can clot on contact with the extracorporeal circuit, particularly when flow is interrupted or anticoagulation is limited. Look for new fibrin or clot in the venous chamber or fibers. A significant pressure change at the same blood flow or a loss of clearance can support concern. Interpret the direction with the clot location and pressure sensors. Color and pressure alone cannot grade the extent of clotting. Check flow interruptions, access and circuit resistance, and report the pattern and anticoagulation delivered to the nurse.

Follow the prescribed plan for a circuit that is still flowing and may be developing clot. Anticoagulation and any prescribed flushes depend on the patient and facility procedure. If the line or chamber is fully obstructed by clot, stop the pump and isolate the circuit. Do not overcome resistance by forcing saline or blood through the obstruction. Notify the RN, follow the procedure for circuit loss and record any blood not returned. The care team determines whether and how dialysis continues.

Inspect the used dialyzer and document what changed

After treatment, handle the used dialyzer as potentially contaminated. Look through the housing for visible clots or fibers with retained blood, and inspect the housing, ports and connections for damage or leakage. Review any blood-leak alarms and records of interrupted flow. Report what you can observe and any unexplained drop in delivery. Appearance helps identify a possible circuit problem but cannot calculate clearance or replace device quality checks. Follow the applicable disposal or reprocessing policy and preserve equipment for investigation if the facility directs.

Respond to a power interruption

Report the interruption, check the patient, inspect the access and circuit and determine which functions are available. Follow the facility emergency plan and the machine's power-failure instructions. In the 2008T Rev. AF example, a mains power failure stops the blood pump and closes the venous clamp. The dialysate pump, heater, blood-leak detector and venous level detector become nonfunctional. The device can sound a battery-powered loss-of-power alarm for seven minutes. That alarm does not provide normal treatment monitoring. Other devices and installed backup sources may behave differently. The unit's response includes communication, patient disposition and alternatives if treatment cannot be resumed. During the handoff, record the outage time, treatment delivered, whether blood was returned and which machine functions were confirmed unavailable. This lets the nurse and technical team distinguish a patient issue from a unit-wide failure and decide when the system is ready for use.

Distinguish manual return from manual circulation

Only use manual return when authorized and when the circuit is suitable for return. Stop and get clinical direction if there is air in the return path, suspected hemolysis, extensive clot or any condition that makes reinfusion unsafe. The 2008T instructions separate hand-cranked saline reinfusion from temporarily circulating blood manually. For reinfusion, the manual describes the supplied hand crank and clockwise rotor direction as shown on the pump housing, with strict visual control and the unit's saline-return procedure. The manual describes 6–10 rotations per minute in its manual circulation section. This device number does not establish a rate for reinfusion or for another pump.

Watch the venous chamber and bloodline continuously for air during authorized manual return. A power outage can make automated air detection unavailable. Use the device instructions for the clamp and circuit setup and the stopping point, with adequate light and assigned staff. Verify manual-return readiness and staffing as part of drills. If you cannot safely observe, the path is obstructed, or air is detected, stop and follow the emergency procedure.

Worked example — outage and two different circuits

A power outage stops the powered pumps on two 2008T stations. The charge nurse activates the facility response.

  • Intact circuit. The first patient's circuit is eligible for manual return after assessment. If the nurse directs return, trained staff use the 2008T procedure, watch the chamber and return line throughout, and document the time and disposition.
  • Obstructed circuit. The other patient's venous chamber contains a clot that blocks return. Staff isolate the circuit, notify the nurse and document its loss without forcing reinfusion.

Sources 2008T operator's manual, Rev. AF, pp. 171–173, FDA hemodialysis device quality guidance, ch. 8, BC Renal complication protocol, clinical context CMS Appendix H r246 V331 and V352 and 42 CFR §494.62 emergency preparedness.

Test Your Knowledge

The venous chamber has a visible obstructing clot, and flow cannot be maintained. What is the appropriate response?

A

Force saline through the chamber to restore blood return.

B

Stop and isolate the circuit, report the obstruction and withhold forced reinfusion.

C

Increase pump speed until the clot passes.

D

Use manual return to bypass the obstruction alarm.

Test Your Knowledge

On a 2008T, trained staff are directed to return blood manually after an outage. The circuit is eligible for return. What instruction applies?

A

Use a fixed 20–30 RPM rate for every model.

B

Treat the loss-of-power alarm as proof that the blood-leak detector is still active.

C

Use the manual circulation rate as the mandatory rate for reinfusion.

D

Follow the 2008T manual-return instructions with the specified crank direction, saline procedure and strict visual control.

Test Your Knowledge

Why must staff directly observe the venous chamber and return line during authorized hand-cranked return on a 2008T after mains power loss?

A

The battery alarm verifies the level detector is functioning.

B

Automated UF measures the balance of manual saline during the outage.

C

Normal venous level detection is unavailable. Direct observation helps prevent air from being infused.

D

Hand-cranking uses the dialysate blood-leak detector to show when to stop.