2.3 Fluid Removal Dynamics: Ultrafiltration Rate (UFR) and Transmembrane Pressure (TMP)
Key Takeaways
The programmed ultrafiltration rate in mL/hr is distinct from the weight-normalized rate in mL/kg/hr. Use the weight specified by the question or the reporting method.
For a planning estimate using prescribed target weight, divide the planned machine ultrafiltration volume by planned hours and then by target kg.
The former CMS ESRD QIP reporting formula used measured pre- and postdialysis weights, delivered treatment time and measured postdialysis weight. CMS removed that reporting measure starting in payment year 2026.
An ultrafiltration rate above the patient's ability to refill and compensate can cause symptomatic hypotension. Risk depends on the person and session.
The dialyzer ultrafiltration coefficient describes water permeability. Interpret the machine's transmembrane pressure display and alarm limits using its instructions for use.
Ultrafiltration, Convection, and Plasma Refilling
Ultrafiltration (UF) is the movement of water across a semipermeable membrane. A pressure difference across the dialyzer drives net water removal. Some dissolved solutes travel with that water, a process called convection or solvent drag. Plasma proteins create an opposing oncotic pressure. Blood and dialysate pressures can both vary along the hollow fibers as the two fluids travel in opposite directions.
The machine removes fluid from the circulating plasma. Water moves from surrounding tissue into the circulation through plasma refilling. The rate varies with fluid distribution, the patient's condition, and how far the session has progressed. If removal outpaces refilling, circulating volume falls. Cardiac output and vascular tone determine how well the patient compensates. A patient can be fluid overloaded overall and still develop low blood pressure during rapid removal.
Isolated ultrafiltration uses a prescribed mode focused on fluid removal without routine dialysate flow. Sequential dialysis can include an isolated UF period followed by hemodialysis with dialysate flow for diffusive solute removal. Check the ordered phase, duration and fluid goal for each part of treatment. Time used for isolated UF may expose the patient to a higher rate during that phase and reduces the time available for diffusive clearance unless the plan accounts for it.
Calculating Ultrafiltration Rate
The machine's average planned UF rate in mL/hr is the prescribed machine UF goal in mL divided by the planned UF time in hours. The machine goal may include fluid expected to enter the patient during treatment, according to unit practice and the current prescription. Distinguish it from the change expected on the scale. The problem must say which weight to use when it asks for a weight-normalized rate. For a planning question that specifies target weight:
The historical Centers for Medicare & Medicaid Services (CMS) End-Stage Renal Disease Quality Incentive Program (ESRD QIP) calculation used the measured pre- to postdialysis weight decrease times 1,000, divided by delivered dialysis hours and then by measured postdialysis weight. If time is recorded in minutes, divide by 60. The historical reporting calculation and a treatment estimate can yield different answers because measured weight change, actual time and post weight can differ from planned values. CMS removed the UFR reporting measure as of payment year (PY) 2026.
Higher removal rates have been associated with worse outcomes in observational research, including studies comparing rates above 13 mL/kg/hr with rates at or below that value. Kidney Disease: Improving Global Outcomes (KDIGO) advises considering clinical factors and the competing risk of leaving fluid overload. Symptoms and trends matter at any rate. Confirm a high planned rate or one the patient has not tolerated with the registered nurse (RN). The clinical team may reconsider treatment time, the amount to remove, the target weight, or whether an additional treatment is appropriate.
Worked Examples: Planning and Reviewing a Treatment
Planning case. Elena weighs 63.4 kg before treatment. The prescribed target is 60.0 kg. Assume the team expects a net 200 mL of fluid to enter the patient during treatment, and that the 3.4 kg gap represents 3,400 mL. Planned machine UF is 3,600 mL. Over 4.0 hours, 3,600 ÷ 4.0 is 900 mL/hr. Divide 900 by the stated 60.0 kg target to get 15.0 mL/kg/hr. Holding the goal and weight constant, 4.75 hours would give about 12.63 mL/kg/hr. Report the rate with the patient's symptoms and vital signs so the team can confirm the plan. The prescriber determines the actual treatment time.
Delivered case. Marcus weighs 88.4 kg before treatment and 85.4 kg after 225 delivered minutes. His prescribed target was 85.0 kg, and his recorded machine UF was 3,300 mL, including 300 mL estimated net input. Under the historical CMS method, use the 3.0 kg measured difference and 3.75 delivered hours. The calculation is 3,000 ÷ 3.75 ÷ 85.4, or 9.37 mL/kg/hr. Substituting 3,300 mL or the 85.0 kg target would answer a different question. Record the weight above target and report why the session ended early if known.
Dialyzer Ultrafiltration Coefficient and Transmembrane Pressure
The dialyzer ultrafiltration coefficient, , describes water movement for a given pressure under specified test conditions. Units are mL/hr/mmHg. For the same UF rate and other comparable conditions, a higher coefficient requires a smaller driving pressure. Compare the model's instructions rather than using a single cutoff to identify membrane performance.
In a physical model, mean blood-side hydraulic pressure minus mean dialysate-side hydraulic pressure describes the hydraulic gradient toward dialysate. On-screen transmembrane pressure (TMP) values depend on the device's measurement locations and sign convention. For example, the Fresenius 2008T calculates the displayed value as measured dialysate pressure minus measured venous pressure. With a dialysate reading of +75 mmHg and measured venous pressure of +95 mmHg, the displayed TMP is −20 mmHg. The sign on that display does not reverse the direction of prescribed fluid removal. On the 2008T a balancing system maintains the dialysate flow balance, while the UF pump removes additional fluid from the dialysate circuit at the programmed rate. Check the exact model's controls and monitor the UF removed over time.
| Observation | What to check within role |
|---|---|
| Pressure trend changes or TMP alarm occurs | Check the patient, circuit and access first. Note the displayed message, baseline and direction of change. |
| A dialyzer or line problem is suspected | Look for visible clotting, kinks, loose connections or leakage. Check the prescribed settings and follow the model's alarm instructions. |
| Pressure is inconsistent with the UF setting or a leak or fault persists | Report promptly to the RN and involve qualified technical staff. Follow policy for stopping treatment or removing equipment from service. An acceptable pressure reading alone cannot exclude all UF delivery errors. |
Sources: KDIGO blood pressure and volume management report, UFR cohort study, historical CMS measure specifications, CMS final rule, and Fresenius operator manuals.
A 3.5-hour treatment is planned with a machine UF goal of 3,150 mL. Use the prescribed target weight of 75.0 kg for normalization. What is the planned weight-normalized ultrafiltration rate?
9.0 mL/kg/hr
12.0 mL/kg/hr
14.2 mL/kg/hr
10.5 mL/kg/hr
A patient develops dizziness and hypotension during fluid removal, although the planned rate is 11 mL/kg/hr. Why can this happen?
Plasma refilling increases the amount of fluid withdrawn by the machine.
Every rate below 13 mL/kg/hr maintains circulating volume.
Removal can outpace that patient's plasma refilling, and cardiovascular compensation may be inadequate.
A rate normalized by weight directly measures plasma refilling.
On a Fresenius 2008T, measured dialysate pressure is +80 mmHg and measured venous pressure is +120 mmHg. What TMP does the model display using these readings?
+40 mmHg
+200 mmHg
−200 mmHg
−40 mmHg