10.2 Reverse Osmosis (RO) Technology: Membranes, Percent Rejection, Conductivity, and Operating Pressures
Key Takeaways
Feed pressure drives water across a semipermeable membrane. Crossflow carries retained solutes toward the concentrate outlet.
Calculate rejection as (feed conductivity − product conductivity) ÷ feed conductivity × 100 using comparable readings.
For a system without reject recycle, calculate recovery as product flow ÷ (product flow + reject flow) × 100.
Compare conductivity, rejection, pressure and flow trends with the manufacturer and facility limits. Under RD52:2004, rejection below 90% prompts chemical analysis.
If RO is the final chemical purification step, staff must be able to prevent unsafe product water from reaching points of use when a water quality alarm sounds.
How the membrane separates two streams
In osmosis, water tends to move across a selectively permeable membrane toward the side with higher solute concentration. An RO pump raises hydraulic pressure on the feed side above the opposing osmotic pressure difference. Some water permeates the membrane. The rest flows across the feed surface and carries concentrated solutes away. This crossflow limits buildup at the membrane surface. Some units recycle a portion of concentrate, so distinguish fresh feed and net drain from internal membrane flows.
A common design wraps membrane sheets and flow spacers around a tube that collects permeate. A thin-film composite element uses a selective polyamide layer on supporting material. A suitable intact system reduces dissolved inorganic solutes, bacteria and endotoxin. Product quality depends on the feed, element and operating conditions. A second pass can further reduce solutes by taking first-pass permeate into a second RO stage. Upstream carbon reduces municipal disinfectant exposure. Free chlorine can damage susceptible polyamide elements, and carbon performance is itself a patient safety control.
Read three different measurements
| Measurement | Calculation or observation | Interpretation |
|---|---|---|
| Conductivity-based rejection | Fractional reduction in a proxy for dissolved ions. Both readings need the same units and a comparable temperature basis. | |
| Water recovery | Share recovered as permeate. Without reject recycle, feed equals product plus reject. With recycle, product divided by product plus net drain flow describes overall utilization, while single-pass recovery uses the total membrane feed including recycle. | |
| Pressure drop across an RO stage | Membrane-stage inlet pressure minus concentrate outlet pressure | A rise at comparable conditions suggests increased feed-channel resistance. Verify the taps and trends before diagnosing a cause. This reading is distinct from pressure across the membrane. |
Product water conductivity reflects the aggregate presence of mobile ions. It does not identify individual chemicals or measure endotoxin or bacteria. The same product conductivity gives a different rejection percentage if the feed changes. For example, at a product reading of 12 µS/cm, rejection is 97% with a feed of 400 µS/cm, but 88% with a feed of 100 µS/cm. Log the actual numbers and trends, including pressures and flow readings available on the unit. A higher recovery concentrates salts in a smaller reject stream, which can increase scale risk. Keep flow and pressure settings within the unit's approved range.
Respond to quality alarms and changing performance
Under the U.S. Medicare requirements, the installation must demonstrate that RO product water meets applicable chemical limits if RO is the last purification stage. Online instruments must allow staff to determine rejection and product-water conductivity. A suitable resistivity or TDS instrument may be used in place of the conductivity monitor. The system uses a preset product quality alarm, with an alarm audible in the treatment area if RO is the final chemical purification step. In that situation, the facility needs a means to stop unsafe product water from reaching points of use. Automatic diversion is one design. Staff must know how to stop flow where the system has no automatic diverter.
Record operating parameters each treatment day and compare them with established limits. Obtain chemical analysis at installation, after RO membrane replacement and at least annually. Falling rejection below 90% or a seasonal change suggesting poorer source water also calls for chemical analysis under RD52:2004. The 90% value is a test trigger. Follow facility alarm settings and response procedures even if the calculated rejection exceeds 90%. For a water quality alarm, notify the charge nurse and responsible water staff, verify that affected equipment cannot receive unsafe product, and follow the contingency plan.
Compare stage pressure drop at comparable conditions or normalize performance according to the equipment instructions. Fouling can increase pressure drop. Colder feed can reduce permeate output even without fouling. Select a permitted cleaning or disinfection cycle based on the fault and equipment instructions. With chemical disinfection, complete the specified rinsing and residual checks before return to service.
Worked example — stable readings followed by an alarm
An RO with no reject recycle has feed conductivity of 420 µS/cm and permeate conductivity of 12 µS/cm, both at the same reference temperature. Product flow is 7.5 gpm and reject flow is 5.0 gpm. Membrane-stage inlet and concentrate outlet pressures are 190 and 174 psi, measured at the same taps used for a clean baseline of 15 psi differential.
- Rejection is .
- Feed flow is gpm. Recovery is .
- Stage pressure drop is psi, up 1 psi from baseline. Compare all three results with this unit's approved limits. A reading near baseline says little about the concentration of an individual contaminant.
- Later, a high conductivity alarm sounds at 48 µS/cm while feed remains 420 µS/cm. Rejection is now . This unit's diverter activates. Verify isolation, notify the clinical team and water-system staff, and follow contingency and testing procedures. The fall warrants chemical analysis even though the percentage cannot determine whether oxidant damage, a leak, or another fault caused the change.
Sources 42 CFR §494.40(a), CMS Appendix H, V177 and V199–V201, and DuPont FilmTec cleaning criteria, Rev. 13.
Feed conductivity is 400 µS/cm and product conductivity is 16 µS/cm, reported on the same temperature basis. What is the RO rejection, and what does it establish?
84%; replace the membrane based on this reading
92.5%; apply an acid cleaning cycle
94%; the number proves there is no endotoxin
96%; compare the trend with operating limits and verify individual contaminants separately
Why is carbon adsorption placed upstream of a chlorine-sensitive polyamide RO membrane in a municipal water system?
To reduce disinfectant exposure that can damage the membrane, while protecting patients from oxidant exposure
To add bicarbonate to the permeate
To refrigerate the feed before it enters the membrane housing
To increase the suspended sediment in the membrane feed
At comparable flow, the RO membrane-stage inlet and concentrate outlet pressures are 180 and 155 psi. The clean differential was 14 psi. What does the change suggest?
The membrane pressure drop is 14 psi, so no resistance has changed
The stage pressure drop rose to 25 psi, suggesting increased feed-channel resistance that needs investigation
The water recovery rate is now 25%
The pressure readings confirm sodium has precipitated out of the product