20.3 Hemodialysis Access Interventions & Central Venous Lines
Key Takeaways
Select native fistula, graft or catheter according to the patient’s dialysis and vein-preservation plan.
Flow, caliber and depth help assess maturation. The older Rule of 6s is a heuristic, and clinical usability requires a broader assessment.
Treat clinically important lesions. Verify the specific angioplasty pressure limit and thrombectomy indication, and check the entire circuit after salvage.
Use device-specific tip targets for infusion and dialysis catheters. Verify the implanted port and access set before power injection.
Preserve central venous access options in advanced CKD. A line that flushes without blood return needs evaluation for several causes, including a fibrin sheath.
20.3 Hemodialysis Access Interventions & Central Venous Lines
Clinical Core: Follow the dialysis circuit from arterial inflow through the access to central venous outflow. Match clinical dysfunction to its cause before treating a narrowing. Distinguish local circuit repair from central venous catheter placement and malfunction.
Access design and maturation
The 2019 U.S. KDOQI update bases access selection on the patient's kidney-replacement and access plan. Consider anatomy, urgency, expected dialysis duration, prior access and patient preferences. A functioning native fistula has no implanted prosthetic conduit. Grafts may be useful when appropriate vein or maturation time is limited. A central venous catheter can meet immediate needs and in selected patients may be the longer-term choice. Catheters bring risks including infection, thrombosis and central vein stenosis.
| Access | Anatomy or feature | Assessment |
|---|---|---|
| Arteriovenous fistula (AVF) | Native artery connected to a native vein. Radiocephalic joins radial artery and cephalic vein. A brachiocephalic AVF connects brachial artery and cephalic vein. A brachiobasilic AVF can use vein transposition for superficial access. | Examine the inflow, anastomosis and draining vein. Juxta-anastomotic narrowing can compromise maturation. |
| Arteriovenous graft (AVG) | A prosthetic conduit bridges an artery and vein. A forearm loop and upper-arm configuration are possible. | Check the graft body and both anastomoses. The graft-vein anastomosis is a common site of narrowing. Time to use depends on the graft and the clinical plan. |
| Dialysis catheter | Venous access with blood withdrawal and return lumens. A nontunneled line can meet immediate needs. A tunneled cuffed line can be used over a longer interval. | Review position, mechanical function, infection risk and a vein-preservation plan. |
The older 2006 KDOQI Rule of 6s used flow greater than 600 mL/min, diameter at least 6 mm, depth no more than 6 mm and discernible margins to describe fistulas more likely to be usable. It called for evaluation if a fistula had not matured at six weeks. Current decision-making includes examination and clinical usability, supported by imaging when appropriate. In the prospective Hemodialysis Fistula Maturation study, flow, diameter and depth helped predict maturation. No single ultrasound cutoff confirms that a fistula can be cannulated successfully.
Worked comparison. An access was easy to cannulate last week. It now has prolonged postdialysis bleeding and a stronger pulse with a diminished thrill. Those findings point toward impaired venous outflow, so the team can examine the draining vein and central outflow, among other circuit segments. A culprit narrowing must explain the clinical change.
Fistulography, angioplasty and declot
Clues to access dysfunction include new cannulation difficulty, prolonged postdialysis bleeding, altered thrill or bruit, high venous pressures or ipsilateral arm swelling. Use imaging to examine the arterial inflow, anastomosis, conduit, draining veins and central outflow as needed. A puncture or catheter aimed toward venous outflow can show the downstream circuit. Reflux views with transient outflow compression or a separately directed approach can help evaluate arterial inflow. Dual puncture is one option.
KDOQI supports balloon angioplasty as a primary treatment for a clinically and angiographically significant AVF or AVG stenosis, using higher pressure as needed. Select balloon size and design for the anatomy. A resistant waist may prompt reassessment of fit or an alternative device. Rated burst pressure is a labeling limit. It does not prescribe the pressure needed to treat a lesion. For example, BD lists an 18-atm rated burst pressure for a 12 mm × 6 cm Atlas Gold model on an 80 cm shaft. The Conquest 40 CQF5078 is listed at 40 atm. Confirm the exact item and stay within its labeling. A stent-graft may be appropriate in selected cases, including rupture or certain graft-vein anastomosis lesions. Review whether it would block future access options.
A thrombosed access requires timely evaluation of clot burden, suspected cause, patient risk and dialysis needs. Depending on anatomy and available expertise, options include surgical or percutaneous salvage, aspiration, a mechanical technique or a pharmacomechanical approach. For example, the U.S.-labeled AngioJet AVX set is intended for qualifying AV access conduits at least 3.0 mm in diameter with its console. It uses directed fluid jets and active removal of thrombus. Match the specific set and check the instructions, including rhythm, hemolysis and operating precautions. Doses, device choices and opposing sheath directions vary. A balloon catheter can help retrieve an arterial plug in a selected case. Check the reopened circuit for a clinically important culprit lesion and confirm inflow and outflow. Anatomic stenosis is common, but causes and salvage plans vary.
Central lines and appropriate tip position
- A PICC enters peripherally, often through an arm vein, for central infusion. Many upper-body infusion devices target the lower SVC near the superior cavoatrial junction (CAJ), where the SVC joins the right atrium. The BD PowerPICC example specifies the lower third of the SVC. Use the verification method appropriate to the model and setting, including imaging or a validated tip confirmation system where applicable.
- A tunneled cuffed dialysis catheter serves high-flow extracorporeal treatment. KDOQI describes a mid-right-atrial tip for dialysis CVCs in the relevant upper-body setting. That target differs from the infusion PICC example. A tunneled device's cuff provides tissue anchoring. Its external lumens require the appropriate exit-site and hub care.
- The carina offers an approximate radiographic landmark for the CAJ. Its relationship varies with the patient and projection. Confirm the tip against anatomy, catheter type and local verification method. Examples of concerning malposition include unintended upper venous branches or extension across the tricuspid valve.
Implanted reservoirs
A totally implanted port consists of a subcutaneous reservoir with a septum connected to a venous catheter. It provides repeated access without an external hub between uses. This can suit intermittent long-term therapy such as chemotherapy. Depending on anatomy and the plan, the operator selects a central venous entry site and creates a pocket for the reservoir. Confirm the reservoir, catheter path, continuity and tip, and follow the specific device's access and asepsis instructions. Use the indicated sterile noncoring needle to enter the septum.
When contrast power injection is planned, verify the specific implant and the complete access set. For example, the U.S. PowerPort K251253 clearance permits power injection with the PowerLoc Safety Infusion Set, up to a device maximum of 5 mL/s. The specific set and other components can impose lower limits. Identify and confirm the assembly and tip position and verify patency before use. A routine noncoring needle, a generic port appearance or an infusion from a different visit cannot establish power-injection compatibility.
Preserve veins in advanced CKD when future dialysis access is likely. Coordinate before inserting a PICC or choosing a route that could compromise future access. For upper-body dialysis access, the right internal jugular is often preferred when suitable. The subclavian route can jeopardize future access through central stenosis. Symptomatic central outflow disease can produce ipsilateral edema, chest-wall collaterals or access dysfunction. KDOQI advises against treating an incidental central narrowing without clinical indicators.
A line that flushes but cannot withdraw blood may have a fibrin sheath, which can act like a one-way valve. Tip-wall contact, malposition, kink and intraluminal thrombus can also cause dysfunction. Evaluate the likely cause before attempting salvage. A catheter-directed lytic lock can address suspected thrombotic occlusion under the applicable protocol. Exchange and selected sheath disruption are other options for persistent dysfunction. KDOQI leaves the decision to disrupt a fibrin sheath during exchange to clinical judgment and does not establish a preferred disruption method.
At follow-up a radiocephalic fistula has an estimated flow of 650 mL/min and a 6.1-mm lumen but lies 9 mm below the skin. How should these results be interpreted?
Flow above 600 mL/min by itself establishes clinical readiness.
Depth establishes that the fistula is thrombosed.
Flow and diameter are encouraging, but depth and access course still affect cannulation. Assess clinical usability and use imaging as needed.
Begin cannulation automatically at six weeks regardless of findings.
Angioplasty for a clinically significant AVG stenosis leaves a persistent balloon waist. The selected labeled balloon has an 18-atm rated burst pressure. Which interpretation is appropriate?
Continue past 18 atm as long as the balloon retains a waist.
Inflate to 18 atm in every lesion because the rated burst pressure is the nominal diameter pressure.
Place a stent-graft in any residual narrowing without reassessing the result.
Reassess sizing and the lesion. If a different balloon or strategy is chosen, verify its labeled limits without exceeding the current balloon’s rating.
Which comparison fits typical upper-body placement of an adult infusion PICC and a tunneled hemodialysis catheter under the stated U.S. guidance?
Both tips should end in the mid-subclavian vein.
An infusion PICC such as PowerPICC targets the lower SVC. A tunneled upper-body dialysis catheter commonly targets the mid-right atrium under KDOQI.
Both tips should extend through the tricuspid valve to the ventricular apex.
The dialysis line should always stop at the carina and the PICC should lie in the upper SVC.
An adult has an implanted venous port for repeated infusions and is scheduled for power-injected contrast. The port is reported as power rated, but the access set cannot be identified. What should the team do before connecting the injector?
Proceed at the full port rate because the reservoir can withstand it.
Proceed with any noncoring needle that provides blood return.
Confirm the port model, power-rated access set, placement and applicable limits before proceeding.
Use the prescribed IV drug rate as the contrast injector setting.