11.3 Catheter Types, French Sizing & Selective Curves
Key Takeaways
A catheter's nominal outer diameter is one-third of its French size in millimeters. Check the actual sheath, catheter lumen and wire compatibility separately.
Liner, braid and jacket influence catheter control. Confirm torque response and tip position under imaging.
A multi-sidehole pigtail distributes contrast along the distal catheter. Confirm the indicated use, expected pressure, delivered flow and position.
Embolization uses an agent-compatible selective delivery device. Multihole infusion catheters serve specific catheter-directed thrombolysis applications.
Reverse-curve motion depends on the segment, curve formation and approach. Confirm the observed relationship of tip, wire and target branch.
11.3 Catheter Types, French Sizing & Selective Curves
Clinical Core: A catheter's outer size, inner lumen, usable length, tip shape and exit ports answer different questions. Read each specification for the intended device and procedure. A flush catheter opacifies a larger region. A selective catheter positions its tip in a branch. A coaxial microcatheter can reach a smaller treatment target.
French size and device compatibility
For the nominal outer diameter of a catheter shaft,
| Catheter size | Approximate outer diameter |
|---|---|
| 3 Fr | 1.00 mm |
| 4 Fr | 1.33 mm |
| 5 Fr | 1.67 mm |
| 6 Fr | 2.00 mm |
| 7 Fr | 2.33 mm |
| 8 Fr | 2.67 mm |
An introducer sheath's nominal French marking commonly refers to the catheter profile it accepts. Its wall adds to the outer diameter at the access site, and actual dimensions can differ across sheath models. Terumo's Glidesheath Slender body and tip specifications show why the French designation and the separately specified dimensions matter. Catheters with the same outer diameter can have different internal lumens. Some microcatheters have a tapered profile and list different proximal and distal sizes.
A composite catheter can use a low-friction liner, a reinforcing braid and polymer jacket. The reinforcement helps transfer torque and resist deformation. Liner and wall construction affect device passage, the inner lumen and the way the shaft responds through bends. Check the labeled guidewire, lumen and pressure or injection limits. Torque at the hub may not produce the same instantaneous tip motion in tortuous anatomy.
Worked compatibility check. A catheter shaft marked 5 Fr has a nominal outer diameter of mm. A device passing through it is specified at 0.035 in, or 0.889 mm. Those values alone say nothing about clearance at the hub or tip. Check the catheter's labeled compatible wire or device diameter, any narrower component, usable length and the sheath needed for the catheter itself.
Flush, selective and infusion designs
A multi-sidehole pigtail distributes contrast through several exits along the distal portion. The curved tip helps avoid pointing a single open end directly at the wall. Hole geometry, catheter position and delivery conditions still affect the jets and tip stability. Use the recommended pressure and flow for the exact catheter and tubing. Published flow and pressure pairs can be test-condition values. Merit labeling, for example, identifies the fluid viscosity used to characterize its angiographic catheters.
| Task | Typical catheter function | Check before use |
|---|---|---|
| Aortography or cavography | A suitable flush catheter opacifies the trunk and branch origins or renal veins. | Anatomical coverage, catheter and port position, intended injection and ratings. |
| Pulmonary angiography during acute PE intervention | A small hand injection may confirm local anatomy. A main pulmonary artery pigtail is a nonselective option when that run is indicated. | Review the plan and condition with the operator. Match the catheter and bolus to the hemodynamics and clinical question. |
| Selective diagnostic study | A straight or curved selective catheter helps place the outlet in a branch. | Assess ostial fit, outlet position, flow and potential reflux before using the prescribed injection. |
| Superselective embolization | A dedicated compatible catheter or microcatheter delivers the chosen agent to the planned target. | Review its instructions for inner lumen, exit design, agent compatibility and technique. Confirm selective position and monitor for nontarget branches or reflux. |
| Catheter-directed thrombolysis | Some specifically labeled infusion catheters have multiple ports for drug distribution along a clot. | Confirm the infusion segment and system labeling. A diagnostic flush catheter and a treatment infusion catheter serve different purposes. |
In bronchial artery embolization, the CIRSE standard describes end-hole diagnostic catheters and a coaxial microcatheter for superselective access. Distribution depends on position, flow, collaterals and the prescribed agent. A sidehole flush injection and an end-hole embolic delivery need different position checks.
Selective curves and target geometry
| Shape | Typical use of the geometry |
|---|---|
| Straight or simple angled, such as Berenstein | A simple profile can direct the wire into an accessible branch or serve in selective diagnostic work. The approach and direction of the ostium guide the choice. |
| Headhunter | A shaped catheter that can be used to select supra-aortic branches in suitable anatomy. |
| Cobra | A forward-facing compound curve can align with selected visceral or bronchial origins. |
| Simmons or Sidewinder | A reverse curve can direct the tip toward a branch that is difficult to align with a simpler curve. This shape is used in selected arch anatomy and during some bronchial studies. |
| Judkins Right | A preformed coronary catheter developed for right coronary engagement. Use the curve and access route discussed in the cardiac anatomy lesson. |
A reverse curve may need to be formed in the vessel under fluoroscopy. Its distal tip can change direction as the secondary curve is advanced, rotated or withdrawn. The observed motion depends on the approach, curve formation and ostial orientation. Watch the operator's view of the tip and wire. If the curve displaces, wire movement meets unexpected resistance or the outlet abuts the wall, reassess before contrast or therapy.
Worked curve check. A forward-facing catheter fails to stay aligned with an arch branch. The operator evaluates a reverse curve after reviewing the arch and approach. The relevant observations are whether the secondary curve has formed and whether the tip and wire point toward the target without force. The technologist confirms the requested shape and wire compatibility and prepares for imaging of the landing position.
An interventional technologist is preparing a vascular catheter for an adult diagnostic arteriogram. The catheter package indicates that the device has an outer diameter of 1.67 mm. What is the corresponding French scale size of this catheter?
3 French
5 French
6 French
4 French
Why can a suitable multi-sidehole pigtail be used for a nonselective diagnostic contrast run in a large vessel?
Its sideholes concentrate all the contrast in a jet directed forward from the tip.
Its sideholes guarantee that any embolic agent reaches one selected branch.
The nominal French size guarantees the same power-injection limit across pigtail models.
Several exits distribute contrast around the distal catheter. The model's limits and the positions of its tip and ports still matter.
A simple forward-facing catheter repeatedly loses alignment with an arch branch. What feature may make a Simmons-type catheter useful?
Its tip always advances farther into a branch when the hub is pushed straight forward.
Its French size automatically increases when the catheter meets resistance.
A formed reverse curve can reorient the tip relative to the branch takeoff. Alignment and motion must be observed under imaging.
A pigtail bumper automatically centers its distal end in the target branch.