23.4 Intracardiac Hemodynamics

Key Takeaways

  • Level the pressure transducer to the cardiac reference height and zero it to atmosphere. Recheck height and the waveform before comparing measurements.

  • Direct Fick output uses measured oxygen consumption and the arterial-to-mixed-venous oxygen content difference. Convert oxygen content from mL/dL to mL/L when reporting output in L/min.

  • Calculate Qp/Qs from the systemic divided by the pulmonary oxygen content difference. Saturations approximate the ratio when hemoglobin is comparable and dissolved oxygen is negligible.

  • Use the mean simultaneous left ventricular to aortic gradient in the Gorlin aortic valve area calculation. Low flow can produce severe stenosis with a mean gradient below 40 mmHg.

  • Divide the systemic pressure drop by systemic flow for systemic vascular resistance and the pulmonary pressure drop by pulmonary flow for pulmonary vascular resistance. One Wood unit equals about 80 dyn·s/cm⁵.

Last updated: November 2026

23.4 Intracardiac Hemodynamics

Transducer Setup and Waveform Discipline

Leveling sets the height of the pressure reference. Zeroing sets the instrument to zero while it is open to atmosphere at that reference. For supine right heart catheterization, the 2022 European Society of Cardiology and European Respiratory Society guidelines use the mid-thoracic level. A transducer above the reference reads low and one below reads high. Recheck the level if the patient or table moves.

An arterial pressure system's fast-flush test assesses its dynamic response. A brief square pulse should be followed by a rapid return to baseline with minimal oscillation. Overdamping blunts the systolic peak and can raise the apparent diastolic pressure. Excess ringing can exaggerate the systolic peak. Examine the circuit and catheter position when a waveform changes and assess the patient at the same time.

Intrathoracic pressure shifts the recorded pressures as the patient breathes. In routine right heart catheterization, record pressures, including the pulmonary artery wedge pressure (PAWP), at quiet end expiration without a breath hold. With large respiratory swings, the operator may average measurements over three to four cycles. Document the measurement convention so tracings can be compared.

Reading Chamber and Vessel Pressures

The table gives illustrative recordings for one resting, supine patient. Read the waveform and report the measurement type with each number.

SiteExample pressureMeasurement
Right atrium (RA)4 mmHgMean atrial pressure
Right ventricle (RV)25 / 5 mmHgSystolic / end-diastolic pressure
Pulmonary artery (PA)25 / 10 mmHgSystolic / diastolic pressure
PA wedge9 mmHgMean wedge pressure, an estimate of mean left atrial pressure in an adequate wedge recording
Left ventricle (LV)120 / 9 mmHgSystolic / end-diastolic pressure
Aorta120 / 75 mmHgSystolic / diastolic pressure

The right ventricular and PA systolic readings are similar in this example. As the catheter moves into the PA, the higher diastolic pressure and the pulmonary valve closure notch help identify the new position. The 2022 pulmonary hypertension guideline's resting right atrial reference range is 2–6 mmHg and its cardiac index reference range is 2.5–4.0 L/min/m². Interpret clinical measurements with the other pressure and flow data.

The Fick Principle

At steady state, systemic blood flow multiplied by systemic oxygen extraction equals oxygen consumption. With no shunt, systemic and pulmonary flows are equal, and pulmonary artery blood provides the mixed venous sample.

CO=V˙O2CaO2−CvO2CO=\frac{\dot{V}O_2}{C_aO_2-C_vO_2}

Here CaO2C_aO_2 and CvO2C_vO_2 are the arterial and mixed venous oxygen contents. Direct Fick uses measured oxygen consumption. An estimated Fick method substitutes predicted consumption. One published adult estimate uses 125 mL/min/m² times body surface area, but actual consumption varies.

Worked example. Measured consumption is 250 mL/min, hemoglobin is 14 g/dL, arterial saturation is 96%, and mixed venous saturation is 75%. For this exercise assume dissolved oxygen is negligible and use the supplied binding coefficient of 1.36 mL of oxygen per gram of hemoglobin.

  • Arterial content is 1.36×14×0.96=18.2784 mL/dL1.36\times14\times0.96=18.2784\,\text{mL/dL}.
  • Mixed venous content is 1.36×14×0.75=14.28 mL/dL1.36\times14\times0.75=14.28\,\text{mL/dL}.
  • The difference is 3.9984 mL/dL, or 39.984 mL/L.
  • Cardiac output is 250/39.984≈6.25 L/min250/39.984\approx6.25\,\text{L/min}.

For height 170 cm and weight 70 kg, the Mosteller formula gives BSA=170×70/3600≈1.82 m2BSA=\sqrt{170\times70/3600}\approx1.82\,\text{m}^2. The cardiac index is cardiac output divided by body surface area, about 3.44 L/min/m².

Shunt Quantification

A step-up in oxygen content during an oximetry run can locate where oxygenated blood enters the right heart. For the pulmonary and systemic circuits, the steady-state Fick relationships give:

QpQs=CaO2−CmvO2CpvO2−CpaO2\frac{Q_p}{Q_s}=\frac{C_aO_2-C_{mv}O_2}{C_{pv}O_2-C_{pa}O_2}

The terms are systemic arterial, mixed systemic venous, pulmonary venous, and pulmonary arterial oxygen contents. For a left-to-right shunt, estimate the mixed systemic venous content from samples upstream of shunt entry. The pulmonary arterial sample includes blood added by the shunt.

Worked example. Assume a steady-state ventricular septal defect study, comparable hemoglobin in all samples and negligible dissolved oxygen. Saturations are systemic artery 96%, mixed systemic venous 70%, pulmonary vein 96% and pulmonary artery 84%. The shared hemoglobin factor cancels and Qp/Qs=(96−70)/(96−84)=26/12≈2.17Q_p/Q_s=(96-70)/(96-84)=26/12\approx2.17. In this example pulmonary flow is more than twice systemic flow and the net shunt is left to right. A ratio below 1 means pulmonary flow is below systemic flow. Interpret the ratio in the context of anatomy and sampling. Equal net flows can coexist with bidirectional shunting. Use full oxygen content if dissolved oxygen contributes materially, such as during high inspired oxygen testing.

Valve Gradients and Valve Areas

The pressure drop across the aortic valve is estimated from simultaneous LV and aortic tracings during ejection. For the mitral valve, measure the diastolic gradient from left atrial and LV pressures. If the operator uses a wedge tracing as a surrogate for left atrial pressure, account for waveform delay. A wedge-based gradient can still overestimate the direct gradient after correction.

The invasive Gorlin equation estimates aortic valve area (AVA) in cm² from the time-averaged pressure gradient during systolic ejection.

AVA=COSEP×HR×44.3×ΔP‾AVA=\frac{CO}{SEP\times HR\times44.3\times\sqrt{\overline{\Delta P}}}

Enter cardiac output in mL/min when it represents forward flow across the aortic valve, systolic ejection period (SEP) in seconds per beat, heart rate (HR) in beats per minute, and the mean gradient in mmHg. With a forward flow of 5,000 mL/min, an ejection period of 0.30 seconds, a heart rate of 75 and a mean gradient of 40 mmHg, the result is about 0.79 cm². The Hakki formula provides a simplified estimate with different inputs. On echocardiography the continuity equation uses the left ventricular outflow area multiplied by the ratio of outflow-tract to aortic valve velocity-time integrals.

The ACC/AHA valve guideline identifies high-gradient severe aortic stenosis with a peak velocity of at least 4 m/s or a mean gradient of at least 40 mmHg. Aortic valve area is typically at most 1.0 cm² in that group. Severe disease can also present with lower gradients in low-flow states. For rheumatic mitral stenosis, a mitral area of 1.5 cm² or less is the usual severe threshold in the ACC/AHA staging framework.

Vascular Resistance

Use the pressure drop across the circuit and the flow through that circuit. With pressure in mmHg and flow in L/min, the results are in Wood units.

SVR=MAP−P‾RAQsSVR=\frac{MAP-\overline{P}_{RA}}{Q_s}

PVR=P‾PA−P‾AWPQpPVR=\frac{\overline{P}_{PA}-\overline{P}_{AWP}}{Q_p}

Here MAP is mean arterial pressure, and the subscripts RA, PA and AWP are right atrial, pulmonary arterial and pulmonary artery wedge pressures. In the absence of a shunt, QsQ_s and QpQ_p both equal cardiac output. Multiply Wood units by approximately 80 to convert to dyn·s/cm⁵.

Worked examples with no shunt. A mean systemic arterial pressure of 93 mmHg, mean RA pressure of 5 mmHg and output of 5 L/min give an SVR of 17.6 Wood units, or 1,408 dyn·s/cm⁵. A mean PA pressure of 20 mmHg, wedge pressure of 10 mmHg and output of 5 L/min give a PVR of 2.0 Wood units, or 160 dyn·s/cm⁵.

References

Test Your Knowledge

In a steady-state study with no shunt, measured oxygen consumption is 250 mL/min, hemoglobin is 14 g/dL, arterial saturation is 96%, and pulmonary artery saturation is 75%. Use 1.36 mL of oxygen per gram of hemoglobin and assume dissolved oxygen is negligible. What is the closest direct Fick cardiac output?

A

2.5 L/min

B

6.3 L/min

C

12.5 L/min

D

8.4 L/min

Test Your Knowledge

The patient's table is raised before another intracardiac pressure measurement. What should the technologist verify before comparing the new and previous pressures?

A

The pressure transducer is leveled to the cardiac reference height and properly zeroed to atmosphere.

B

Zeroing once per case wherever the transducer happens to sit

C

Reading mean pressures during mid-inspiration

D

Calibrating the system with contrast injected through the catheter