18.1 Mechanical Embolics: Coils & Vascular Plugs
Key Takeaways
Choose an occlusion site after identifying inflow, outflow and routes of collateral filling.
Check coil or plug diameter, length, delivery lumen and release requirements against the exact instructions.
The tether can permit controlled repositioning in designated detachable devices. Confirm limits before attempting resheathing.
Coil fibers and hydrogel change packing and clot formation. Durable intent still requires a documented flow endpoint and follow-up when indicated.
18.1 Mechanical Embolics: Coils & Vascular Plugs
Clinical Core: Mechanical embolization aims to interrupt flow at a planned location. Identify inflow, downstream runoff, branches to preserve and potential alternate routes before checking hardware fit. Recheck the actual flow endpoint after placement.
Choose an occlusion site
Coils and vascular plugs provide a physical obstacle that can slow flow and promote thrombosis. They are generally intended for durable occlusion. The result depends on device design and packing, available landing length, local flow and the patient's capacity to form clot. Durable intent does not guarantee immediate stasis or permanent exclusion. An absorbable agent may be selected for temporary occlusion in a different clinical situation. Particles and liquid agents can reach other levels of the vascular bed, with different risks and delivery requirements (§18.2).
| Clinical goal | Questions to answer on imaging |
|---|---|
| Focal arterial hemorrhage | Which artery feeds the injury, and is there a second route to the bleeding site? |
| Aneurysm or pseudoaneurysm exclusion | Can the sac be isolated while maintaining perfusion in necessary branches? Which vessels enter or leave it? |
| Shunt or fistula flow reduction | How fast does flow reach the draining vein? Could the device migrate into another circulation? |
| Endograft-related branch exclusion | Does the problem arise at an attachment seal or from branch backflow? Different mechanisms require different plans. |
A gastroduodenal artery pseudoaneurysm may fill through the hepatic arterial side and from the superior mesenteric side via pancreaticoduodenal collaterals. If both contribute, closing only the proximal route may leave retrograde filling. An isolation approach can address both when appropriate. A covered stent can be an alternative when maintaining the parent artery matters and there is a suitable landing zone. Define the operator's endpoint and the views needed to verify it.
Coil structure and release
The coil's wound wire forms a delivery profile, and the coil assumes its intended secondary shape as it leaves the delivery catheter. Radiopaque materials allow fluoroscopic tracking. Fibered coils provide a matrix for clot formation. Some hydrogel-coated platinum coils expand after hydration, increasing the material volume available to fill the target. For example, the manufacturer describes limited early expansion and full expansion within about 20 minutes for the AZUR Peripheral HydroCoil system. Expansion, placement time and effect depend on the specific model. Hydrogel technology does not guarantee a durable clinical outcome.
| Delivery type | Implication |
|---|---|
| Pushable coil | The approved pusher moves the coil through the catheter without a controllable final detachment from a tether. A released coil cannot be pulled back with its pusher. Separate retrieval may be possible if a complication occurs. |
| Detachable coil | The coil remains coupled to a delivery system until an intentional release. Many systems allow specified positioning or resheathing before release. Read the device instructions for the marker, time and resheathing constraints. |
Match the compatible catheter lumen, coil, pusher or release system. French outer size alone does not establish delivery lumen compatibility. Assess usable length along the actual catheter path and avoid forcing any component that meets resistance. Coil diameter is selected for the measured landing zone and expected geometry. CIRSE describes 10% to 20% oversizing for vessel occlusion in its visceral and renal aneurysm standard. This example does not establish a universal rule for every coil, sac and vessel. An undersized coil may migrate. Excessive size may prevent the intended shape or move the catheter.
In selected sac embolizations, a framing coil establishes a supporting boundary for packing. Subsequent softer coils can fill the remaining space. Packing density expresses nominal coil volume divided by target volume. The result depends on assumptions about both volumes, particularly with a coating that expands. The operator assesses the residual flow and parent vessel directly. A specific ratio does not substitute for angiography.
Select a plug by its chart
Many plugs use a self-expanding nitinol mesh and remain attached to a cable during positioning. The actual device specifies where a plug can be deployed, how it can be repositioned, the delivery path and compatible lumen. Abbott's current U.S. chart lists the 12 mm AVP II for an 8.0–9.0 mm vessel. In an 8.0 mm vessel, its nominal oversize is . This implant needs the model's listed delivery lumen, at least 0.070 inch for that size. AVP 4 has a different profile and chart. Its 8 mm model is listed for 5.5–6.0 mm vessels, so an 8 mm vessel is outside that model's chart. Check the relevant instructions and actual landing length before opening the device.
Worked endpoint. A plug appears seated, but a delayed run still shows flow into the intended exclusion zone from a separate branch. Report where and when flow enters. Inspect the supply again before marking the case complete. Flow around an implant and collateral flow into the target indicate different mechanisms.
A tethered detachable coil projects into a branch that must remain open. It has not been released. What does the connection allow the team to consider?
Repositioning or resheathing when allowed by the device instructions
Assuming the coil will move back by itself after release
Imaging only the main vessel and ignoring the branch
Delivering it with a different pusher selected by French outer size alone
An 8.0 mm vessel is in the given 8.0–9.0 mm range for a 12 mm AVP II. If all other suitability checks are pending, what nominal oversize does this chart entry represent?
12%
25%
50%
150%
A repair excludes proximal flow into a gastroduodenal pseudoaneurysm. A selective superior mesenteric run still opacifies the sac through pancreaticoduodenal vessels. Which interpretation fits?
The sac is excluded because its original proximal access was closed.
Collateral backfilling persists and must be reported as residual perfusion.
The finding establishes a complete seal at both ends.
The finding proves that all distal vessels are occluded.