23.6 Permanent Pacemakers and Implantable Defibrillators
Key Takeaways
A pacemaker supports a slow rhythm. An ICD treats dangerous ventricular tachyarrhythmias and may also provide pacing.
CRT-P has pacing functions and CRT-D adds defibrillation. A coronary venous lead can provide left ventricular pacing.
Device appearance helps identify hardware. Confirm the model and therapy from the implant record.
Check lead position and testing independently. Electrical capture requires depolarization after the stimulus.
New instability after implantation warrants immediate communication and assessment for complications.
23.6 Permanent Pacemakers and Implantable Defibrillators
Permanent rhythm devices can look similar in a chest image, so begin with the therapy and then trace the hardware. A conventional generator includes a battery and electronics inside a pocket. One or more connected leads sense cardiac activity or carry electrical therapy. The operator selects an implant for the rhythm problem, expected pacing requirement, anatomy and patient goals. The imaging team helps map the route, document placement and recognize possible complications. The device team verifies electrical performance.
Match the implant to the clinical problem
A permanent pacemaker uses timed, low-energy impulses to sustain an adequate rhythm when indicated. A pacing stimulus that depolarizes the intended chamber has electrical capture. In sinus node dysfunction, the relationship between symptoms and bradycardia guides decisions. Acquired Mobitz II, high-grade or complete atrioventricular block can warrant permanent pacing even in an asymptomatic patient when reversible or physiologic causes have been excluded. This is a different decision from temporary pacing to support unstable bradycardia during an emergency or another procedure.
An implantable cardioverter-defibrillator (ICD) detects programmed ventricular tachyarrhythmias and can deliver a shock. Many transvenous systems also provide pacing for slow rhythms and antitachycardia pacing to terminate selected ventricular tachycardias. An ICD may be appropriate after ventricular fibrillation or unstable sustained ventricular tachycardia in selected patients after assessment of the cause and prognosis. Other patients qualify for primary prevention based on disease and risk with no previous arrest. An ejection fraction by itself cannot determine the full indication.
Cardiac resynchronization therapy (CRT) coordinates ventricular activation in selected patients with heart failure and electrical dyssynchrony. CRT-P provides resynchronization with pacing; CRT-D adds defibrillation. One conventional biventricular approach uses a right ventricular lead and a left ventricular pacing lead carried through the coronary sinus to a coronary venous branch. That lead can pace the left ventricle from the epicardial side. The implant may also have an atrial lead. Venous anatomy and electrical findings affect the final lead positions.
Trace the system on two views
| System | Feature to verify |
|---|---|
| Conventional transvenous pacemaker | Generator in the pocket. Trace the leads along the central veins to their intended chambers. In a dual-chamber system, one lead typically ends in the right atrium and another in the right ventricle. |
| Transvenous ICD | A right ventricular defibrillator lead has one or more radiopaque shock coils. Additional leads depend on pacing or resynchronization needs. |
| Totally subcutaneous ICD | The EMBLEM system uses a generator and a subcutaneous electrode. It provides neither long-term bradycardia pacing nor antitachycardia pacing or CRT. |
| Leadless pacemaker | The original Micra is a self-contained example implanted inside the right ventricle via a transcatheter femoral venous route. It has no separate transvenous lead or chest pocket. Other leadless devices have their own configurations. |
For conventional transvenous devices, the operator may enter by cephalic venous cutdown or image-guided axillary access. Recognize the lead course through the central veins and check the tip in the designated projection. A left ventricular coronary venous lead runs away from the right ventricular lead and follows the venous branch on the heart surface. Review frontal and additional views to resolve overlap. Inspect continuity and lead slack, especially after repositioning or connecting the generator. Check the device identification and the implant record before assigning its features from imaging alone.
Pair anatomic and electrical confirmation
Sensing describes detection of the patient's intrinsic signal. Capture threshold describes the minimum stimulus that reliably depolarizes the chamber at a specified pulse duration. Impedance measures the electrical opposition in the lead circuit. Use the device and implant protocol when interpreting values. These measurements answer different questions. For example, a ventricular lead can project over the intended chamber yet fail to capture during testing. Report the discrepancy promptly. The device team checks the measurements and lead connections while imaging evaluates position and motion.
Inspect the final images and hand off concerns including a changed lead course or suspected pneumothorax. After transvenous lead placement, implant consensus recommends chest radiography to document position and evaluate for pneumothorax, plus electrocardiography and device checks before discharge. An acute blood pressure drop with concerning signs warrants rapid assessment for complications such as perforation and pericardial effusion. Document the actual device, lead locations, access, final checks and notable events. Subsequent procedures may need the exact model and device team input for pacing dependence and electromagnetic interference planning.
References
A conventional biventricular device has one lead running into a lateral cardiac vein through the coronary sinus. What is that lead intended to do?
Deliver a shock from a subcutaneous electrode.
Sense pressure in the left atrium.
Pace the left ventricle through the coronary venous route.
Enter the left ventricular cavity through the mitral valve.
The operator confirms that the new right ventricular lead projects at the intended location. During the pacing test, the stimulus repeatedly fails to produce ventricular depolarization. Which finding needs to be addressed?
Failure of electrical capture despite the apparent lead position.
Successful capture because a pacing stimulus was delivered.
Confirmation of an adequate shock threshold.
Proof that the lead is a subcutaneous electrode.
A patient assessed for an ICD also needs continuous long-term bradycardia support. What should the team recognize about the EMBLEM S-ICD by itself?
Its subcutaneous electrode provides chronic dual-chamber pacing.
It automatically adds a coronary sinus lead if pacing is needed.
Its shock function makes the pacing requirement irrelevant.
It cannot provide long-term bradycardia pacing, so the device plan must address both needs.