2. Bundle Branch Blocks and Paced Rhythms
Key Takeaways
Adult complete bundle branch block typically has a QRS of at least 0.12 seconds and a characteristic morphology. Width alone has several possible explanations.
An rSR' (rabbit-ears) pattern in V1 suggests right bundle branch block; a broad notched R wave in V5–V6 and lead I suggests left bundle branch block — the provider confirms the diagnosis.
LBBB can change the baseline ST-T pattern. Escalate ischemic symptoms with a suggestive tracing promptly and let the clinician evaluate the findings in context.
A visible pacing spike marks a stimulus on the tracing. Look for the response in the intended chamber. Ventricular QRS shape depends on the pacing site and method.
A correctly timed ventricular stimulus without a resulting QRS suggests noncapture. An absent spike alone cannot establish device failure. Assess the patient and report suspected malfunction promptly.
A wide QRS can result from altered ventricular conduction, a ventricular rhythm or ventricular pacing. Use lead-specific morphology and pacing information to describe what you see.
Bundle Branch Block: One Ventricle Gets the Message Late
Below the AV junction, the His bundle divides into the right bundle branch and the left bundle branch, each feeding its ventricle's fast-conduction fibers. When one branch is blocked, the unblocked ventricle depolarizes normally and quickly, while activation of the affected ventricle is delayed. This changes the morphology and prolongs the QRS. Adult complete RBBB or LBBB generally requires a QRS of at least 0.12 seconds with the corresponding lead findings.
Right Bundle Branch Block (RBBB)
In right bundle branch block, the right ventricle is the late one. In leads V1 and V2, look for the classic rSR' pattern — two positive peaks, nicknamed rabbit ears or an M-shaped complex — where the second peak (the R-prime) is the delayed right-ventricular wave. In the lateral leads (I, V5, V6) the same late force arrives as a wide, slurred S wave. RBBB can accompany right-heart strain, lung disease, or congenital conditions, but it also appears in otherwise normal hearts — one more reason confirmation belongs to the provider, not the technician.
Left Bundle Branch Block (LBBB)
In left bundle branch block, the larger left ventricle is the late one. In leads V5, V6, and I, look for a tall, broad notched R wave (another M shape) with the normal septal Q wave absent; in V1, look for a deep, wide negative complex (QS or rS). LBBB can be associated with structural heart disease; report the pattern for clinical evaluation.
Keep the skill at recognition level: you flag a wide QRS with the suggestive lead pattern, and you say what you see — the provider confirms which branch is blocked, because other wide-QRS situations (a ventricular rhythm, pacing, some drug effects) share the width.
Why LBBB Gets Special Respect
LBBB changes ventricular activation and can lead to secondary ST-T changes, often opposite the main QRS deflection. In a patient with ischemic symptoms, report the pattern and the symptoms immediately. The clinician evaluates the tracing for concordant or disproportionate changes using the applicable criteria. Newness of LBBB alone is insufficient to classify it as a STEMI equivalent.
Isolated left fascicular block can shift the frontal QRS axis without a wide QRS. The clinician evaluates the axis and morphology, including whether another conduction disturbance is present.
Artificial Pacemakers: When the Device Drives the Beat
A pacemaker can support patients with selected sinus node and AV conduction disorders. A conventional implanted pacemaker has a generator and one or more leads, but system types and programmed modes vary. Record the known device, rhythm and patient condition.
Pacing Spikes and What Follows Them
A pacing spike, when visible on the surface ECG, identifies a pacing stimulus. Atrial electrical capture produces a P wave and ventricular electrical capture produces a QRS. In AV sequential pacing, a visible atrial stimulus can precede a P wave and a visible ventricular stimulus can precede the QRS. A dual-chamber system can also sense a native beat and withhold a stimulus, so two spikes need not appear every cycle.
Conventional right ventricular myocardial pacing often makes the QRS wide and LBBB-like. Conduction system pacing can produce a narrower or different pattern. A surface ECG can miss small pacing artifacts. Electrical capture alone does not establish a mechanical pulse or adequate perfusion.
Failure Modes the Technician Recognizes and Reports
| Observed finding | What to consider | Technician action |
|---|---|---|
| Ventricular pacing stimulus without the expected QRS | Possible loss of electrical capture; compare rhythm and timing | Assess the patient and report promptly |
| Unexplained pauses without expected pacing artifacts | Possible output failure or inappropriate inhibition from oversensing; also consider recording settings and the programmed pacing mode | Record the pause and symptoms; escalate promptly for clinical review |
| Apparent stimuli inappropriately timed to native beats | Possible undersensing; device interpretation depends on timing, programming and signal quality | Save the tracing and report the timing |
Oversensing is detection of a signal the device should ignore, which may inappropriately inhibit a stimulus. Undersensing is failure to detect a true intrinsic depolarization; a stimulus may then appear at the wrong time. The clinical team checks the rhythm, programmed settings and device function.
Bedside Safety Around Pacemakers
- Use the standard chest landmarks where feasible. If an incision, dressing or clinical restriction prevents standard placement, use the approved alternate site and document it.
- Magnet application is provider-directed. A magnet changes how some pacemakers behave; never place or remove one on your own initiative.
- Document the paced rate, spike presence, and capture consistency, and compare with previous tracings when available, so changes in device behavior stand out.
Before a stress test, a patient with conventional RV pacing has wide QRS complexes following ventricular stimuli at 75 bpm. Two expected ventricular stimuli then have no following QRS, and the patient becomes lightheaded. Pause the test, assess the patient and activate urgent help for possible loss of capture. Save the relevant segment and report the symptoms and timing. Follow BLS if arrest is suspected.
A wide QRS of 0.14 seconds shows an rSR' pattern in lead V1 and wide slurred S waves in leads I and V6. Which finding does this pattern suggest?
Left bundle branch block
A ventricular escape rhythm from complete heart block
Right bundle branch block
Normal septal depolarization
A ventricular pacing stimulus arrives at the expected time, outside an intrinsic QRS, and there is no ensuing ventricular depolarization on a clean tracing. Which problem does this suggest?
Normal dual-chamber pacing
Failure to capture
Failure to pace
A junctional escape beat
Why is a wide QRS common with conventional right ventricular myocardial pacing?
The battery slows conduction through the AV node
Atrial depolarization is always absent during ventricular pacing
The pacing artifact adds width to the QRS measurement
The stimulus activates ventricular muscle locally first and changes the usual ventricular activation sequence
Sections you finish are checked off in the contents.