2. Waveform Morphology and Intervals

Key Takeaways

  • At 25 mm/s and 10 mm/mV, each 1 mm square represents 0.04 seconds horizontally and 0.10 mV vertically. A large square spans 0.20 seconds and 0.50 mV.

  • The PR interval is measured from the beginning of the P wave to the beginning of the QRS complex; normal is 0.12 to 0.20 seconds.

  • Q comes before any positive QRS deflection, R is the first positive deflection, and S follows R. An adult QRS of at least 0.12 seconds is wide.

  • A T wave is generally upright and asymmetric in lead II. New narrow peaked T waves can occur with hyperkalemia, while broader hyperacute T waves can be an ischemic finding. Report the shape, leads and symptoms.

  • U waves are often easier to see in V2–V3 at slower rates. Prominent U waves can support suspicion of hypokalemia, especially with T-wave flattening.

Last updated: November 2026

The Grid: Your Measuring Stick

At the standard settings of 25 mm/s paper speed and 10 mm/mV gain, one small square is 1 mm on the grid and represents 0.04 seconds horizontally and 0.10 mV vertically. One large square spans five small squares along each axis, or 0.20 seconds by 0.50 mV. Five large squares therefore span one second, and 30 large squares span the 6-second strip used for rate counting. Every amplitude, interval, and ST measurement in this section is read against that grid. If the speed or gain was changed — for example, a 50 mm/s strip taken to spread out a very fast rhythm — the printed calibration marks change meaning and the change must be documented; the guide's Equipment chapter covers calibration standards and machine operation.

The Baseline

The isoelectric line (baseline) is the zero-potential reference on the recording between waveform deflections. The true resting baseline is the TP segment, the flat stretch from the end of one T wave to the start of the next P wave. Use a stable baseline for amplitude and ST measurements. The TP segment may be difficult to distinguish at faster rates; document obscuring artifact and follow the measurement protocol.

The Waveforms

P Wave: Atrial Depolarization

The P wave is the first small deflection of each cycle and represents atrial depolarization. In sinus rhythm, the impulse starts near the SA node high in the right atrium. Normal values: smooth, round, and upright in lead II, less than 2.5 mm tall, and less than 0.12 seconds wide (under three small squares). A tall, peaked P wave can suggest right atrial abnormality; a wide, notched P wave suggests left atrial enlargement. Those are descriptive findings the technician records factually for the provider's interpretation.

QRS Complex: Ventricular Depolarization

The QRS complex represents ventricular depolarization and can show one or more deflections, with size and direction varying by lead. Its letters are assigned purely by deflection direction:

  • Q wave — the first negative (downward) deflection, before any positive deflection.
  • R wave — the first positive (upright) deflection.
  • S wave — any negative deflection that follows an R wave.

A mostly upright complex with a small terminal dip may be written Rs. A QRS of 0.08 seconds is within the expected adult range. A duration greater than 0.11 seconds is prolonged under adult ECG standards. At 0.12 seconds or more, consider wide-QRS patterns such as bundle branch block, ventricular ectopy, or ventricular pacing and review the morphology.

T Wave: Ventricular Repolarization

The T wave is the broad hump after the QRS representing the ventricles electrically resetting (repolarizing). Normal: upright in lead II, rounded, and asymmetric — a gentler upstroke followed by a steeper downstroke. That normal asymmetry is a subtle but testable point. New narrow, symmetric, tented T waves can occur with hyperkalemia. Broader, unusually prominent T waves in neighboring leads can be an early ischemic change. Check lead distribution and symptoms, compare an available prior tracing and report concern promptly.

U Wave: The Small Afterthought

The U wave is a small, rounded deflection immediately after the T wave, best seen in leads V2 and V3 at slower heart rates. It is easy to overlook — which is exactly why exams test it. A prominent U wave, large enough to clearly alter the contour after the T wave, is a classic clue to hypokalemia, and U waves are often easier to see at slower rates. The morphology alone cannot establish the potassium level. Report it as a descriptive finding; do not dismiss it as noise or artifact.

ComponentRepresentsNormal durationNormal appearance
P waveAtrial depolarizationUnder 0.12 sRound, upright in lead II, under 2.5 mm
QRS complexVentricular depolarization>0.11 s is prolonged in adults; ≥0.12 s is wideDirection and amplitude vary by lead
T waveVentricular repolarizationVaries with rateRounded, asymmetric, upright in lead II
U waveDeflection with an uncertain mechanismVariableUsually small and round, follows T, often clearest in V2–V3

Intervals and Segments

PR interval. Measured from the start of the P wave to the start of the QRS complex; normal is 0.12 to 0.20 seconds (three to five small squares). A PR over 0.20 seconds with all P waves conducted fits first-degree AV delay. A PR under 0.12 seconds can occur with pre-excitation or a low atrial or junctional origin. Compare consecutive intervals and account for any P waves without a following QRS before describing a conduction pattern.

QRS duration. Measured from the first deflection of the QRS to the J point — the junction where the QRS ends and the ST segment begins. At 25 mm/s, three small squares span 0.12 seconds.

ST segment. Runs from the J point to the start of the T wave. Describe the direction, amount and distribution of any deviation from the baseline. The criteria for an urgent ischemic pattern depend on the leads, morphology and clinical context; see Section 4.

QT interval. Measured from the beginning of the QRS to the end of the T wave; it captures the total time of ventricular electrical activity. The QT naturally shortens as the rate rises and lengthens as the rate slows, so machines print a rate-corrected value, the corrected QT (QTc). QTc interpretation depends on age, sex, QRS width and correction method. In adult hospital monitoring, values above roughly 450 ms for men or 460 ms for women are prolonged. A QTc above 500 ms warrants prompt clinical review per protocol because torsades risk rises.

MeasurementStarts atEnds atNormal value
PR intervalBeginning of the P waveBeginning of the QRS0.12–0.20 s
QRS durationFirst deflection of the QRSJ point, where the QRS ends≥0.12 s is wide
ST segmentJ pointStart of the T waveAssess deviation, lead distribution and morphology
QT intervalBeginning of the QRSEnd of the T wave, excluding a distinct U waveRate-dependent; document the method and QTc

Waveform Quality Descriptors

When describing waveforms — in documentation or on the exam — use precise, standard adjectives. These descriptors carry specific meaning:

DescriptorWhat you are describingWhat it can suggest
Symmetric / asymmetricWhether the two halves of the waveform mirror each otherNarrow tented T waves may suggest hyperkalemia; regional broad hyperacute T waves may suggest ischemia
DirectionPositive (upright), negative (inverted), or biphasic (both directions)An inverted or biphasic P wave in lead II suggests the impulse did not start in the SA node
AmplitudeHeight or depth in millimeters against the baselineA measured change can be compared with a prior tracing in the same lead
DurationWidth in seconds (small squares)A wide QRS or a wide notched P wave points to conduction or chamber findings
NotchedA split or double-peaked contourA notched P wave suggests atrial enlargement; a notched QRS can accompany bundle branch block
Fragmented / slurredA broken or smoothly blurred upstroke or downstrokeConduction abnormality within the ventricles

Notice the word biphasic — it means the waveform swings both upright and inverted within the same deflection, and it is a favorite exam descriptor because it is easy to confuse with simply inverted.

How the Technician Handles Interval Findings

Many machines print estimates of the PR, QRS and QT when they identify the relevant landmarks. Treat those printouts as a starting point, not an authority: scan them for obvious discrepancies against the tracing — a printed PR of 0.30 seconds on a strip whose P waves sit nearly fused to the QRS deserves a recapture or an immediate flag. Never write over or obscure the tracing's printed values; record your verification note separately, and include flagged interval values when you report findings. The final interpretation of any interval belongs to the provider.

Warning

Flag a markedly prolonged printed QTc or new concerning T- or U-wave changes and report them with any symptoms according to clinical urgency. Record that an automated number is machine-generated; rate, artifact, QRS width or an indistinct T-wave end can affect it.

Test Your Knowledge

Where does the PR interval begin and end?

A

End of the P wave to the end of the QRS complex

B

Beginning of the P wave to the peak of the R wave

C

Beginning of the P wave to the beginning of the QRS complex

D

Beginning of the Q wave to the end of the T wave

Test Your Knowledge

In QRS nomenclature, what is the first negative deflection after an R wave called?

A

An S wave

B

A Q wave

C

An R-prime wave

D

A U wave

Test Your Knowledge

A slow rhythm shows unusually prominent U waves, most clearly in V2 and V3, after flattened T waves. Which electrolyte change can produce this pattern?

A

Hyperkalemia

B

Pericarditis

C

Right atrial enlargement

D

Hypokalemia

Test Your Knowledge

Which QRS duration is within normal limits?

A

0.14 seconds

B

0.08 seconds

C

0.12 seconds

D

0.16 seconds

Sections you finish are checked off in the contents.