2. Maintenance and Inspection of EKG Equipment

Key Takeaways

  • Inspect wires for damage and loose connections before use. Tagged equipment with recurring signal dropout should be checked off the patient under an approved procedure.

  • At the standard scale of 25 mm/s and 10 mm/mV, a 1 mV reference pulse is 10 mm high, a small box is 0.04 seconds, and a large box is 0.20 seconds.

  • Measure against the recorded speed and gain. An intentional alternate scale can produce a usable tracing if its settings are clearly identified.

  • Clean and disinfect reusable patient-contact and high-touch surfaces between patients using a compatible method and the product’s labeled contact time.

  • Never tape or splice a damaged cable: remove it from service, tag it, and submit a work order to biomedical engineering.

Last updated: November 2026

Scheduled Versus As-Needed Maintenance

Equipment care comes in two layers. Scheduled (preventive) maintenance follows a calendar: the quick checks you perform every shift, plus the deeper periodic inspections and internal calibration performed by the facility's biomedical engineering (clinical engineering) department under the manufacturer's preventive-maintenance schedule. As-needed (corrective) maintenance responds to a fault: a frayed lead wire, a jammed printer, a machine that fails its self-test. The technician owns the daily layer and the reporting; opening the machine is never the technician's job.

Daily and Weekly Checks

FrequencyCheckWhat to look for
Every shiftPower and batteryAdequate charge or a secure connection to a grounded outlet; cord intact
Every shiftPatient cable and lead wiresCracked insulation, frays, kinks, corroded or bent snaps
Every shiftThermal paperAdequate supply; paper loaded as the device instructions show
Every shiftElectrode stockSealed pouches, in date, gel moist
Every patient useConsole, display, keyboardClean surfaces; controls respond correctly
Weekly (per policy)Full cable set with an approved check or simulatorEvery lead records cleanly; clips cleaned of dried gel; cart wheels and drawers functional

Inspecting the Electrode Tray and Lead Wires

Lead wires and clips take more physical abuse than any other part of the system, and can develop visible or intermittent faults. Inspect for:

  • Cracked or stiffened insulation along the wire, especially near the snap boots and the cable yoke where bending concentrates.
  • Frayed cable jackets and kinked sections that hint at broken conductors inside.
  • Corroded or bent snaps and clips, and dried gel crusted onto contact surfaces.
  • Bent or missing pins in the patient cable connector.

If gentle movement reproduces dropout, the cause may be the connector, contact or cable. Remove suspect hardware and test it off the patient only with approved equipment. Substitute a known-good compatible cable if permitted. Tag the faulty part and report it rather than trying to repair the insulation or connector.

Battery and Charging Discipline

Return the machine to its charger or connect it to its approved power supply after use according to facility policy. For a scheduled test, use an adequately charged or connected device because a power failure can interrupt capture. Batteries age: a unit that no longer holds a charge through a shift should be reported to biomedical engineering for battery service.

Test Your Knowledge

During your daily inspection you find a lead wire with cracked insulation and a corroded snap. What is the correct action?

A

Wrap the cracked section with tape and keep the wire in service

B

Scrub the corrosion off with an abrasive pad and continue using it

C

Remove the lead wire from service, replace it, and report it per facility policy

D

Move the wire to another department's cart so it becomes their problem

Paper and Supply Storage

Thermal paper is heat-sensitive: store spare rolls in a cool, dry place away from direct sunlight, radiators, and heat-producing equipment, because heat and light darken the paper before it ever reaches the printer. Keep the correct roll model for your machine stocked in the drawer. Electrodes are sealed in foil pouches so the conductive gel cannot evaporate; store them in a cool, dry cabinet, follow the label for storage and expiry after opening, and never use an electrode whose gel has dried stiff — dried gel raises the electrical resistance of the skin contact and produces noisy, wandering tracings. Rotate stock by earliest expiry first. Among stock with the same expiry, use the older supply first.

Calibration and Standard Settings

Two settings define a standard EKG, and you should understand what each one actually does:

  • Paper speed — 25 mm/second. This is the horizontal measurement. At 25 mm/second, one small square equals 0.04 seconds and one large square (five small squares) equals 0.20 seconds. At 50 mm/s a small box is 0.02 seconds and a large box is 0.10 seconds. The same 0.20-second interval occupies 5 mm at 25 mm/s and 10 mm at 50 mm/s. Applying the 25 mm/s grid rule to the faster printout would misread it as 0.40 seconds.
  • Gain — 10 mm/mV. This is the vertical measurement. One millivolt of signal moves the tracing 10 mm (two large squares) upward. Machines print a calibration pulse — a square-wave mark — at the start or end of the strip; a 1 mV mark at standard gain should be 10 mm tall. A 5 mm mark indicates the half-standard (5 mm/mV) setting used when complexes are unusually tall and would overlap the neighboring channels; a 20 mm mark indicates the double-standard (20 mm/mV) setting used to magnify unusually small complexes.

A setting changes how the same signal appears on paper or screen. Read durations using the printed speed and voltages using the gain. If a 10 mm deflection prints at 20 mm/mV, its amplitude is 0.5 mV. Verifying that the standard settings and the calibration mark are correct is technician work; internal recalibration belongs to biomedical engineering.

Cleaning Between Patients and Terminal Cleaning

  • Between every patient, clean and disinfect the reusable patient-contact and high-touch parts with products and methods compatible with the manufacturer’s instructions. Remove visible soil, observe the product label’s contact time, and protect connectors from liquid ingress.
  • Additional cleaning after isolation use or at the end of a session follows the facility’s infection-control procedure, the required pathogen coverage, and the device instructions.
  • Dry your hands before touching the console, avoid harsh solvents on screens, and follow the manufacturer's cleaning instructions.

Storage After Use

Hang lead wires loosely rather than wrapping them tightly around the machine, coil the patient cable in loose loops, park the cart in its designated secure location, plug it in to charge, and keep electrode pouches sealed.

Documenting Maintenance Issues and Work Orders

Any fault you cannot correct at the bedside gets documented, not ignored: remove the equipment from service, attach a do-not-use tag, record the symptom and any error code, and submit a work order to biomedical engineering per facility policy. Keep the equipment log entry complete — machine identifier, date, what you observed, and whom you notified. A repaired machine returns to service only after biomedical engineering clears it.

Test Your Knowledge

At standard settings, how tall should the calibration pulse printed on an EKG strip be?

A

5 mm

B

2 mm

C

20 mm

D

10 mm

Test Your Knowledge

Where should unopened electrode pouches and spare thermal paper rolls be stored?

A

In a cool, dry cabinet away from heat, sunlight, and radiators

B

On a windowsill so the stock is easy to see

C

On top of the printer so the rolls stay warm and print darker

D

In the warmest utility room in the suite