8. Artifact Recognition and Correction During Acquisition
Key Takeaways
Shivering or muscle tension can add irregular, jagged artifact. Offer warmth, support and relaxation, then repeat the recording if appropriate.
In North America, a uniform 60 Hz pattern can suggest mains interference. Separate signal and power cables, check approved connections, and ask the care team before turning off other equipment.
When the baseline drifts, check movement, breathing, and electrode contact. Use the approved skin-prep and hookup procedure before repeating.
RA–LA reversal inverts lead I. RA–LL reversal inverts lead II and gives an inverted exchange of I and III. LA–LL reversal inverts III and exchanges I and II.
Assess the patient first. A stable pulse does not rule out VT, while independent native beats visible through the disturbance can support artifact. Escalate concern promptly and follow the emergency protocol if indicated.
An artifact is any deflection on the tracing that does not come from the heart. The EKG machine amplifies tiny electrical potentials, so it records skeletal muscle, wall outlets, and cable motion just as faithfully as myocardium - and artifact does not merely blur a tracing, it can mimic lethal rhythms. Patients have been shocked and treated for ventricular fibrillation that was never there. Task 4.1.4 is the skill this section builds: identify the artifact, correct it at the bedside during acquisition, and know when a frightening rhythm is real.
The Three Baseline Artifacts
Somatic tremor
Signature: jagged, irregular, spiky baseline of varying height - muscle electricity recorded by electrodes near moving skeletal muscle. Causes: shivering from cold, anxiety, parkinsonian tremor, tense handgrip, discomfort.
Fixes, in order: offer a blanket, reassure the patient, support the limbs in a relaxed position, and re-record if appropriate. A tremor that cannot be quieted gets protocol-specified handling; the machine's filter trade-offs belong to facility policy, and the physical fixes come first.
60-cycle AC interference
Signature: a perfectly uniform, fine, fuzzy oscillation across the baseline - the rhythm of alternating-current electricity in North American wiring. Causes: electrical coupling from nearby sources or problems with signal contacts or equipment. Cables routed close to power cords may contribute.
Fixes: separate the patient cable from power cords, verify the manufacturer-approved connections and coordinate with the clinical team about nearby devices. Keep essential patient equipment operating. If the fuzzy pattern is perfectly regular, think electricity before you think muscle.
Wandering baseline
Signature: a slow, rolling drift of the whole baseline up and down the grid, dragging every waveform with it. Causes: respiration (deep or labored breathing moves the chest), patient movement, and electrode gel drying out or oily skin breaking the contact.
Fixes: prepare the skin according to the device instructions, reseat or replace the electrode, check that the intended sites have not shifted, and coach the patient to breathe quietly and stay still. Then re-record.
Electrode problems
Dried gel, expired pads, and poor adhesion generate noisy, jump-prone leads and eventually a flat or pegging line when contact is lost entirely. Check electrode expiration dates, keep pads sealed until use, and replace - do not reuse - electrodes that have dried. A sudden flat line in some leads while others record normally, or a stylus slamming to the grid edge, can reflect signal loss or a recorder issue. Check the patient, inspect the electrode contacts and connections, and use a compatible spare when indicated.
Lead Reversals and Their Signatures
Swapped wires do not look like noise - they look like disease, which is why their signatures are exam material. Learn the pattern table:
| Reversal | Tracing signature |
|---|---|
| RA and LA swapped | Lead I fully inverted (upside-down P, QRS, T); leads II and III trade places; aVR and aVL trade; chest leads stay normal |
| RA and LL swapped | II is inverted. I becomes the negative of the original III, and III becomes the negative of the original I. aVR and aVF exchange. |
| LA and LL swapped | III is inverted. I and II exchange; aVL and aVF exchange. |
| RA and RL swapped | II can become very small because it samples the difference between the two leg sites. Verify the labels and hookup. |
The first pattern deserves one more line, because it has a famous impostor: RA-LA reversal and dextrocardia both invert lead I and turn aVR upright. The tell is the chest leads - RA–LA reversal leaves the original precordial waveforms unchanged, while mirror-image dextrocardia often shows poor or reversed R-wave progression with left-sided chest placement (see the special-situation section for the mirrored-placement protocol). The fix for a reversal is mechanical: find the two swapped wires, reconnect them correctly, and re-record.
Artifact or True Arrhythmia?
Check responsiveness, symptoms and the pulse as appropriate to the patient’s condition. In suspected arrest, activate emergency response and start the indicated BLS actions without waiting to troubleshoot. If the patient has a pulse, wide-complex tachycardia still warrants prompt clinical assessment.
Compare simultaneous leads when available. For example, if a monitor looks very fast but another simultaneous channel shows clear native QRS complexes continuing independently through the noise at the rate of the patient’s pulse, report the discrepancy and check for movement or a loose electrode. The waveform may be clearer in one lead than another, so verify the signal and report persistent concerns even when the patient appears stable.
The Discipline of Acquisition
Artifact correction is a during-acquisition duty, not a post-subject apology:
- Watch the tracing as it prints or displays - do not discover the problem after handing it off.
- Correct the cause at the bedside (warm, reposition, correct contact or route cables safely) and re-record. A distorted tracing is not fixed by hoping the provider tolerates it.
- Never submit a distorted tracing without documenting the attempts: note the artifact identified, the corrective actions taken, and the re-recording, per protocol.
- Handle the erroneous tracing according to record policy. Identify any repeat and why it was needed. For a tracing already transmitted or filed, notify the responsible staff and use the approved amendment procedure.
The deep patient factors that create artifact - tremor and shivering, hairy chest needing clipped patches and firm prep, obesity and diaphoresis weakening adhesion, movement and restless positioning - all have their fixes taught here because acquisition and artifact control are the same skill. You caught the wandering baseline before you pressed print, or you caught it after and re-recorded: either way the interpreter receives a tracing that reflects the heart, not the room.
Warning
Two screens are always in front of you: the tracing and the patient. When they disagree, the patient wins. A patient can have a real tachyarrhythmia with a pulse. Check the patient, report promptly, and follow the relevant response pathway.
A 12-lead tracing shows a perfectly uniform, fine, fuzzy baseline oscillation across all leads. What is the artifact, and what corrects it?
Somatic tremor, corrected only by changing the paper speed
Possible 60 Hz interference; check the approved connections, separate signal cables from power sources, and confirm the cause safely
Wandering baseline, corrected by increasing gain
Confirmed limb-lead reversal
A tracing shows lead I completely inverted with an upright aVR, but normal R-wave progression across V1 through V6. What does this mean?
Suspect a right-arm/left-arm swap. Verify the physical hookup and repeat if a swap is found.
Diagnose dextrocardia from lead I alone
Assume the tracing is a normal variant without checking
Turn on filtering to correct the connection
One monitor channel shows apparent wide-complex tachycardia while the patient is alert. A separate simultaneous channel shows distinct normal QRS complexes continuing at the rate of the palpable pulse despite interference. What should you do?
Treat the first channel as confirmed ventricular tachycardia
Label the strip as VT and wait for routine review
Check the patient and hookup, promptly report the discrepancy, and correct and repeat the recording as directed
Increase gain until both channels look alike