9.1 Rate Calculation, Rhythm Analysis, Mean QRS Axis & PR/QRS/QTc Intervals

Key Takeaways

  • Check the printed speed and gain first. At 25 mm/s use 300 divided by large boxes for a regular R–R interval. At 50 mm/s use 600.

  • Estimate the mean rate for irregular rhythms over a known duration. When atrial and ventricular activity differ, describe both rates.

  • With positive I and negative aVF, lead II helps tell a normal leftward axis from left axis deviation.

  • QT measurement includes the QRS and ends at the end of the T wave. Report the correction method and assess difficult measurements or broad QRS separately.

Last updated: November 2026

Read in a consistent order

Start by checking identifiers, acquisition settings and lead quality. A standard report often uses 25 mm/s and 10 mm/mV. At those settings, a small box is 40 ms across and 0.1 mV tall. If the report uses a different scale, convert before comparing amplitudes or intervals. Consider the patient's symptoms and capture a previous tracing when available. AHA/ACCF/HRS technology statement

StepWhat to record
RateVentricular rate and, when different, atrial rate
RhythmR–R and P–P regularity, atrial morphology and timing of P waves relative to QRS
AxisFrontal direction of the mean QRS and any uncertainty near a boundary
IntervalsPR, QRS, and measured QT with rate correction
Chambers and conductionVoltage and morphology clues, bundle branch or fascicular patterns
ST, T and Q wavesDistribution, change from baseline, symptoms and the need for prompt review

Rate and rhythm

For a regular rhythm, divide 60 by the R–R interval in seconds. At 25 mm/s, the equivalent constants are 300 divided by large boxes or 1,500 divided by small boxes. Four large boxes take 0.80 s and correspond to 75/min. The same horizontal distance at 50 mm/s takes 0.40 s and corresponds to 150/min. For an irregular rhythm use a measured window. Nine QRS complexes over six seconds give an approximate mean ventricular rate of 90/min. Six seconds spans 30 large boxes at 25 mm/s and 60 large boxes at 50 mm/s. Prefer a longer available recording for irregular timing. Atrial rate can be measured separately from P–P or flutter-wave intervals when atrial activity is visible.

Inspect P waves in multiple leads, their regularity and their relationship to each QRS. Sinus P waves are typically upright in I and II and inverted in aVR. Similar P shapes with alternating acceleration and slowing with respiration support sinus arrhythmia. Sinus atrial activation can still be present when AV conduction is blocked. A negative P and QRS in I with upright aVR can suggest arm-lead reversal; compare with the baseline and verify the electrode labels. With that swap, precordial signals are unchanged, including any pre-existing abnormal progression. AHA ECG technology statement 2019 ESC supraventricular tachycardia guideline

Frontal QRS axis

Estimate net QRS polarity in I and aVF using the dominant positive and negative components. A clearly positive lead generally lies within 90° of the mean QRS direction, and a near-equiphasic lead is approximately perpendicular to it. The AHA adult reference for normal axis is –30° to +90°. Positive I and aVF place the axis between 0° and +90°. With positive I and negative aVF, lead II at +60° divides the quadrant at –30°. If II is positive, the axis is on the normal side. If II is negative, it is more leftward than –30° and meets the left axis deviation classification. Negative I with positive aVF suggests right axis deviation. Two negative leads put the axis in the extreme quadrant. If I or aVF is nearly equiphasic, examine adjacent leads instead of forcing a positive or negative decision. Axis deviation has several possible causes, including conduction changes and technical error. AHA/ACCF/HRS intraventricular conduction statement

Worked axis example. If aVL at –30° is nearly equiphasic, the perpendicular possibilities are +60° and –120°. A positive lead II at +60° selects the +60° direction.

Intervals and QT correction

Measure PR from the P wave onset to the first QRS deflection. A typical adult PR is 120–200 ms, with first-degree AV block defined by PR above 200 ms and 1:1 conduction. Measure the duration of ventricular activation from QRS onset to offset. A QRS of at least 120 ms can occur with complete bundle branch block, ventricular rhythm, pre-excitation or ventricular pacing. The rest of the tracing determines the cause. Measure QT from the QRS onset to the end of T, excluding a discrete U wave. Use a clearly defined T-wave end and review other leads or the baseline when T and U overlap. The QT includes ventricular depolarisation, so a broad QRS can lengthen measured QT even without the same increase in repolarisation time. AHA/ACCF/HRS ST, T, U and QT statement

Bazett divides QT by the square root of R–R in seconds. Fridericia divides by the cube root of R–R. Both can retain some heart-rate dependence. Bazett often gives higher corrected values at fast rates and lower ones at slow rates. Report the method used and confirm a value that may change clinical interpretation. With a QT of 360 ms and R–R of 0.64 s, Bazett gives 360/0.64=450360/\sqrt{0.64}=450 ms. Part IV of the AHA/ACCF/HRS statement uses practical adult prolonged-QT boundaries of over 450 ms in men and at least 460 ms in women with appropriate rate adjustment. A value around 500 ms or more, measured appropriately, warrants attention to relevant medications and electrolyte findings. QT correction comparison study 2025 AHA adult advanced life support

Loading diagram...
Estimate frontal QRS axis with I, aVF and II
Test Your Knowledge

On a 25 mm/s strip, 8 QRS complexes appear in 30 large boxes during an irregular rhythm. On a second strip at 50 mm/s, a regular R–R interval spans 4 large boxes. What are the ventricular rates?

A

40/min and 75/min

B

About 80/min and 150/min

C

80/min and 37.5/min

D

160/min and 150/min

Test Your Knowledge

QRS is positive in I and negative in both aVF and II. Which frontal-axis classification fits?

A

Normal leftward axis between 0° and –30°

B

Right axis deviation

C

Extreme axis

D

Left axis deviation, more negative than –30°

Test Your Knowledge

A woman has a measured QT of 450 ms, R–R of 0.81 s and a narrow QRS. What does the Bazett calculation give?

A

500 ms, a prolonged corrected interval

B

450 ms, the same as the measured QT

C

405 ms after multiplying by the square root

D

556 ms after dividing by R–R directly

Test Your Knowledge

I shows inverted P and QRS complexes and aVR is upright. Precordial progression is the same as the patient’s prior ECG. Which acquisition error should first be checked?

A

V1 and V2 exchanged

B

V5 placed below V6

C

Right and left arm wires exchanged

D

Incorrect gain on a single chest lead

Sections you finish are checked off in the contents.