11.2 Cardiac Catheterization, Angiography, Electrophysiology Studies & Telemetry Support

Key Takeaways

  • Radial and femoral arterial access are both used for coronary angiography. Routine Allen or Barbeau screening cannot reliably predict access outcomes.

  • Right-heart catheterisation records chamber and pulmonary pressures, including wedge pressure. Coronary angiography images coronary vessels with contrast.

  • Electrophysiology studies combine surface ECG and intracardiac signals. AH reflects conduction to the His signal and HV reflects conduction from His to ventricular activation.

  • Check pacing threshold, intrinsic signal amplitude and impedance as separate measurements. Magnet response and telemetry settings depend on the actual device and approved protocol.

Last updated: November 2026

Coronary angiography and haemodynamic measurements

Coronary angiography uses an arterial catheter to deliver contrast near the coronary openings. A radial or femoral access site can be used. Radial access often reduces access-site bleeding in PCI, and the operator selects the route for the patient and procedure. Allen and Barbeau manoeuvres can describe collateral hand perfusion in a given setup, but SCAI does not recommend routine preprocedural collateral testing to select access. For a right-heart study, a venous catheter can record right atrial, right ventricular and pulmonary arterial pressures. A valid pulmonary artery wedge pressure can estimate left atrial pressure. Crossing the aortic valve to measure LV pressure is a separate manoeuvre. An angiogram alone does not supply the LV end-diastolic pressure. SCAI radial best practices 2025 ACS guideline ESC/ERS pulmonary hypertension guideline

Before the procedure, help verify identity, the planned monitoring, baseline observations and any missing safety information. Relay previous contrast reactions, renal function and medication information. The treating team decides how to manage antithrombotics and metformin according to the procedure, kidney function and applicable protocol. Contrast route and acute kidney injury affect the decision. During angiography keep an interpretable ECG and relay important changes as they occur. A damped pressure waveform can have several causes and can occur when the catheter impedes flow at an ostium. The operator evaluates waveforms and catheter position. A new pause, ST change or run of ventricular beats with a pressure change needs immediate attention. Assess the patient and communicate the change while the operator decides what to do. SCAI cath laboratory consensus ACR Manual on Contrast Media Original report on damped and ventricularised waveforms

After access, note the new bleeding, swelling, pain, altered sensation or poorer distal circulation that should be reported. Check the actual orders for compression, positioning and mobilisation. These depend on the route, the closure method and clinical findings. Do not use a universal bed-rest interval. SCAI cath laboratory consensus

Worked comparison. During a right-heart study a report lists a right atrial pressure and a wedge pressure. These are measured in different positions and describe different sides of the circulation. Compare the displayed waveform and the site documented by the operator before recording the value. In a coronary study, record the moment of an ECG change and whether it was accompanied by symptoms or a pressure drop. This helps the team separate a transient recording problem from a change that needs action. ESC/ERS pulmonary hypertension guideline SCAI cath laboratory consensus

Electrophysiology and implant measurements

In an electrophysiology study, intracardiac electrodes record local atrial, His bundle and ventricular electrical signals alongside the surface ECG. AH measures from the recorded atrial signal to His activation and largely reflects AV nodal conduction. HV measures from His to the earliest ventricular activation. For example, in unexplained syncope with bifascicular block, a baseline HV of at least 70 ms is a specific finding used in the ESC pacing guidance. Programmed stimulation can test conduction or provoke a rhythm for mapping. Ablation can target tissue supporting the arrhythmia. The team decides whether to use radiofrequency, cryoenergy or another approach, and which medicines to continue. ESC pacing guideline ESC supraventricular tachycardia guideline

MeasurementWhat to document
Capture thresholdLowest output that consistently evokes depolarisation at the specified pulse width
Sensing amplitudeMagnitude of the intrinsic atrial or ventricular signal under the measured conditions
Lead impedanceElectrical behaviour of the lead circuit, interpreted with the lead design and prior or repeat readings

Worked example. An implant measurement gives a 0.6 V threshold at 0.5 ms and an 11 mV intrinsic ventricular signal. After the connection is handled, impedance changes abruptly from 650 to 2,400 ohms. Report the change and the other readings immediately. The team can check for a connection or other high-resistance circuit problem. That one change does not identify exactly where the problem lies. Document device and lead details so the test team can apply the specified acceptable range. EHRA conventional implant consensus

Telemetry and device precautions

Inpatient telemetry uses an approved configuration to monitor assigned rhythms or intervals. It uses fewer recording positions than a diagnostic 12-lead ECG in many installations. Choose the displayed leads and settings according to the reason for monitoring and the device instructions. Check the patient promptly after a critical alarm. Verify signal quality and use the approved limits and escalation process rather than suppressing a crisis alarm to reduce noise. If a separate ECG or device interrogation is needed, preserve the relevant telemetry strip and time. Magnetic response differs among pacemakers, defibrillators and models. Follow the device team's procedure and confirm the exact type and programming before any magnet intervention. AHA hospital ECG monitoring statement AHA implanted-device statement, Table 2

Test Your Knowledge

How should an Allen or Barbeau test be described in relation to access selection for routine transradial angiography?

A

Either test guarantees that hand ischaemia cannot occur

B

Every patient must pass both tests before radial access

C

They can assess collateral perfusion, but routine use to decide access is not recommended by SCAI

D

They measure left ventricular end-diastolic pressure

Test Your Knowledge

During contrast injection, a new four-second pause appears with a pressure drop. The patient is awake. What should the ECG technician do immediately?

A

Alert the operator, assess the patient and continue monitoring so the team can respond

B

Silence the pause alarm before notifying anyone

C

Deliver an unsynchronised shock for the pause

D

Give sublingual nitroglycerin independently

Test Your Knowledge

During a device implant the lead impedance abruptly rises from 650 to 2,400 ohms after the connector is handled. What is the best interpretation?

A

Normal capture is proved by the impedance change

B

The value specifically proves a low-resistance insulation short

C

The finding by itself identifies phrenic stimulation

D

Report a possible high-resistance circuit or connection problem and verify it with the device team

Test Your Knowledge

A monitor intermittently alarms for a dangerous rhythm while electrodes are coming loose. What is the appropriate action?

A

Widen all crisis limits until the end of the shift

B

Assess the patient, correct the signal and follow approved crisis alarm and escalation procedures

C

Switch off lethal alarms and rely on the next ward round

D

Confirm a diagnosis from lead II alone without examining the patient