7.4 WPW Syndrome, Accessory Pathways & Pacemaker / ICD Sensing Basics

Key Takeaways

  • Ventricular pre-excitation can produce a short PR, a slurred initial QRS and a wide QRS. Orthodromic AVRT conducts to the ventricle via the AV node.

  • Pre-excited AF can produce a very rapid irregular rhythm with changing QRS width and morphology. The 2023 AF guideline lists AV nodal blockers, including IV amiodarone in this setting, as contraindicated.

  • The NBG code describes pacing and sensing. For a paced beat, check timing of the spike and whether a corresponding depolarisation follows.

  • Magnet response depends on the actual device and programming. Many pacemakers pace asynchronously in magnet mode. In a transvenous ICD, a magnet can suspend tachyarrhythmia therapies without changing bradycardia pacing.

Last updated: November 2026

Ventricular pre-excitation and AV re-entry

In a heart with an atrioventricular accessory pathway, atrial impulses may reach ventricular muscle by that route as well as through the AV node. Early local ventricular activation causes a slurred initial QRS deflection called a delta wave. The early onset can shorten the PR and widen the QRS. The degree of pre-excitation may vary because the contribution of the two routes changes. A resting ECG with manifest pre-excitation describes a Wolff–Parkinson–White (WPW) pattern. WPW syndrome involves an accessory pathway with associated symptomatic arrhythmia. A pathway that conducts only retrogradely can participate in tachycardia without a delta wave in sinus rhythm. A short PR alone does not establish an accessory pathway. 2019 ESC supraventricular tachycardia guideline

A dominant R and positive initial deflection in V1 is the historical type A appearance and often suggests a left-sided pathway. A predominantly negative V1 is the historical type B appearance and does not reliably identify a right-sided insertion. Clinical electrophysiology uses patterns across multiple leads and mapping to define the pathway more precisely. Comparison of localisation criteria Ablation-site validation study

In orthodromic AV re-entry tachycardia (AVRT), the impulse travels down the AV node and back up an accessory pathway. Ventricular activation usually produces a narrow QRS. A baseline bundle branch block or rate-related aberrancy can widen it. In antidromic AVRT, the impulse reaches the ventricle down an accessory pathway and returns through the AV node or a second pathway. The QRS is wide and the tracing can resemble VT. Atrial fibrillation with an antegrade-conducting pathway can produce an irregular, very rapid rhythm with varying QRS widths and shapes. The 2023 ACC/AHA AF guideline contraindicates AV nodal-blocking drugs in pre-excited AF, including adenosine, verapamil, diltiazem, beta blockers, digoxin and amiodarone. Rapid ventricular activation via the pathway can precipitate VF. Urgently report the rhythm and patient status so the team can apply the correct clinical pathway. 2019 ESC supraventricular tachycardia guideline 2023 ACC/AHA atrial fibrillation guideline, Section 10.6

Pacing code and the surface ECG

NBG positionWhat it describesExamples
IChamber pacedA atrium, V ventricle, D dual, O none
IIChamber sensedA atrium, V ventricle, D dual, O none
IIIResponse to sensingI inhibited, T triggered, D both, O none
IVRate modulationR rate-responsive, O none
VMultisite pacingA atrium, V ventricle, D both, O none

In VVI, the pacemaker paces the ventricle, senses the ventricle and can inhibit pacing after sensing a ventricular event. DDD provides both atrial and ventricular pacing and sensing, with both inhibited and triggered responses. The programmed and intrinsic timing determines which pacing pulses appear. An atrial pacing artefact can precede a paced P wave, and a ventricular pacing artefact can precede a paced QRS. In sequential pacing, look for both when neither chamber has already activated independently. Right ventricular apical pacing usually produces a wide LBBB-like QRS through slower spread across ventricular myocardium. Other sites and conduction-system pacing can show different morphologies. Bipolar pacing artefacts may be small and can be difficult to see at some recording settings. Original revised NBG code AHA implanted-device statement

Output, capture and sensing

Check for expected output, a corresponding depolarisation and appropriate timing. Failure to capture means a pacing artefact occurs without the expected atrial or ventricular depolarisation. With undersensing, a native beat fails to reset or inhibit the device when it should. A spike may appear shortly after a visible intrinsic beat. With oversensing, a signal such as noise is counted as cardiac activity and may inappropriately inhibit output. Missing apparent pacing can have several causes, including a programmed inhibition or a spike that is hard to see, so surface observations and device interrogation serve different purposes. Pacing reference manual, surface ECG troubleshooting AHA implanted-device statement

Worked example. A VVI pacemaker set to 70/min has a nominal lower-rate interval of about 857 ms (60,000 divided by 70). If a clearly visible intrinsic QRS occurs 320 ms before a ventricular spike that follows the prior pacing schedule, the device may have failed to register the native beat. Confirm timings on the full strip, note how the patient feels and alert the team. Interrogation can show whether the beat was sensed and explain any programmed timing exceptions.

An ICD may also deliver antitachycardia pacing or a shock. A magnet can change some pacemakers to asynchronous pacing and can suspend tachyarrhythmia therapy in many ICDs. Those responses vary by model, programming and device type. A transvenous ICD with tachyarrhythmia therapies suspended by a magnet usually retains its programmed bradycardia pacing. Some pacemakers have no magnet response or a response programmed off. Record the device details when available and follow the supervising team's device-specific protocol. AHA implanted-device statement, Table 2

Test Your Knowledge

An asymptomatic athlete has a PR of 95 ms, a slurred initial QRS upstroke and a QRS duration of 132 ms in sinus rhythm. What does this ECG show?

A

An isolated short PR with a narrow QRS

B

First-degree AV delay with RBBB

C

A ventricular escape without atrial activation

D

Manifest ventricular pre-excitation, or WPW pattern

Test Your Knowledge

An adult with known pre-excitation has a very rapid irregular tachycardia with varying wide QRS complexes. The treating team suspects pre-excited AF. Which medication is contraindicated in that setting under the 2023 ACC/AHA AF guideline?

A

Procainamide

B

Diltiazem

C

Ibutilide

D

A procedural sedative used during planned cardioversion

Test Your Knowledge

A VVI pacemaker programmed at 70/min emits a spike on its old timing schedule 320 ms after a clear intrinsic QRS. Device interrogation confirms the native ventricular event was missed. Which malfunction occurred?

A

Failure to capture despite appropriate sensing

B

Battery output failure

C

Ventricular undersensing

D

Ventricular oversensing causing inhibition

Test Your Knowledge

The device team confirms a transvenous pacemaker programmed for asynchronous magnet response and a transvenous ICD programmed to inhibit tachyarrhythmia therapy with a magnet. What response is expected?

A

The pacemaker paces asynchronously. The ICD suspends tachyarrhythmia therapies while keeping its bradycardia pacing mode.

B

Both devices stop all pacing for as long as the magnet is present.

C

The pacemaker begins antitachycardia pacing and the ICD delivers a shock.

D

The ICD becomes an asynchronous pacemaker while the pacemaker retains normal sensing.