8.1 ECG Grid, Paper Speed (25/50 mm/s), Gain Calibration (10 mm/mV) & Frequency Filters
Key Takeaways
At 25 mm/s, a 1 mm box represents 40 ms and a 5 mm box represents 200 ms.
At 50 mm/s, the time per box is halved. The large-box rate constant for a regular rhythm is 600 at 50 mm/s and 300 at 25 mm/s.
Divide vertical distance in millimetres by gain in mm/mV. Verify the gain for the actual lead, especially when a recording uses mixed scaling.
Diagnostic bandwidth preserves more waveform information than restricted filters. Verify filter settings and interpret apparent loss of pacing spikes with the patient and device information.
Horizontal and vertical scales
A surface ECG plots voltage against time. The horizontal axis tells how long an event takes. The vertical axis reports the voltage difference for the lead being measured. Read the speed and gain from the report before using a grid or comparing tracings. On a calibrated printout or screen ruler, a small square is 1 mm on each side and a large square is 5 mm on each side. If a strip has time marks, check what they represent for that recorder and speed. A screenshot or rescaled printout can change the apparent size of the boxes, so use the supplied time and voltage references. AHA/ACCF/HRS ECG technology statement
| Printed speed | Time per 1 mm | Time per 5 mm | Large boxes per minute |
|---|---|---|---|
| 25 mm/s | 40 ms | 200 ms | 300 |
| 50 mm/s | 20 ms | 100 ms | 600 |
| 12.5 mm/s, if supported | 80 ms | 400 ms | 150 |
Worked example. At 50 mm/s a regular rhythm has five large boxes between consecutive R waves, an eight-small-box PR interval and a four-small-box QRS. The R–R interval is s, giving beats/min. The PR interval is ms and QRS duration is ms. Reading the strip as 25 mm/s would produce 60 beats/min, 320 ms and 160 ms. That could make ordinary conduction appear abnormally slow. The shortcut for regular rhythms is to divide the number of large boxes in one minute by the R–R distance in large boxes. For an irregular rhythm, use a known recording duration and count complexes across that interval so one unusually long or short beat does not drive the estimate.
Voltage calibration
At 10 mm/mV the 1 mV reference spans 10 mm vertically. At half gain (5 mm/mV) a complex of the same voltage uses half as much vertical space. Half gain helps when tall complexes overlap other channels on the printout, and a larger gain can help resolve small components. Record any changed gain. The shape and layout of the calibration marks depend on the recorder and the report format. Read the individual lead or lead group calibration instead of assuming mixed gains produce a particular stepped pulse. AHA/ACCF/HRS ECG technology statement Welch Allyn electrocardiograph operating manual
| Gain | Vertical height for 1 mV | Voltage per millimetre |
|---|---|---|
| 5 mm/mV | 5 mm | 0.20 mV |
| 10 mm/mV | 10 mm | 0.10 mV |
| 20 mm/mV | 20 mm | 0.05 mV |
Worked example. On a tracing with chest leads set to 5 mm/mV, the S-wave depth in V1 is 11 mm and R in V5 is 14 mm. These are mV and mV, giving a combined voltage of 5.0 mV. At 10 mm/mV the same combined voltage would span 50 mm. Apply voltage criteria to the converted measurement. Confirm whether a threshold uses amplitude in one lead, a sum or a different combination, and remember that ECG voltage is an electrical clue to hypertrophy.
Bandwidth and visible waveforms
A high-pass filter reduces components below its lower-frequency cutoff. Slow baseline movement is one target, but part of the ST and T waveform also has low-frequency content. Raising a conventional analogue high-pass cutoff from 0.05 to 0.5 Hz can shift the ST baseline through phase distortion. A compatible zero-phase digital design can permit a cutoff up to 0.67 Hz under the AHA standardisation statement. A low-pass filter reduces higher frequencies. A diagnostic adult cutoff of at least 150 Hz preserves more QRS detail than a 40 Hz muscle filter. The same statement identifies 250 Hz as more appropriate for infants. A narrow pacing artefact can be attenuated or missed depending on bandwidth, sampling and spike-detection processing. Document an altered filter, reacquire a clean diagnostic ECG when needed and use device information when assessing pacing. AHA/ACCF/HRS ECG technology statement, bandwidth and pacing sections
A notch filter targets a narrow frequency band such as 50 Hz mains interference. India's national reference grid frequency is 50 Hz. Start with electrode contact, cable condition and interference sources before changing the device's filter. Diagnostic and monitoring modes may use different settings. An apparent ST change on a restricted monitoring strip should prompt clinical assessment and a diagnostic tracing when appropriate. Indian Electricity Grid Code AHA/ACCF/HRS ECG technology statement
The tracing is recorded at 50 mm/s. The regular R–R interval spans 5 large boxes and PR spans 7 small boxes. What are the rate and PR?
60 bpm and 0.28 s
120 bpm and 0.14 s
120 bpm and 0.28 s
60 bpm and 0.14 s
The printout labels V5 at 5 mm/mV and lead I at 10 mm/mV. The R wave in V5 is 13 mm tall. What is that amplitude?
0.65 mV
1.3 mV
1.8 mV
2.6 mV
Which filter change on a conventional analogue recorder can distort the ST segment and create an apparent elevation or depression?
Raising the high-pass cutoff from 0.05 to 0.5 Hz
Using an adult high-frequency cutoff of 150 Hz
Allowing frequencies up to 250 Hz for an infant recording
Retaining the same validated diagnostic settings on both recordings
A diagnostic ECG is printed using a restrictive 40 Hz low-pass filter after the patient shivers. Which waveform effect can result?
All time intervals double
Only the inferior ST segments can change
QRS peaks are attenuated and narrow pacing artefacts may become hard to see
Lead I inverts and aVR exchanges with aVL