7.1 Pre-Fitting Assessment: Keratometry, Topography & Vertex Conversion
Key Takeaways
Check comfort, lids, tear stability and corneal integrity before fitting. Interpret tear measurements according to the technique used.
With a keratometric index of 1.3375, conventional K equals 337.5 divided by the anterior corneal radius in millimeters. Verify the instrument setting.
Axial maps summarize broad curvature. Tangential maps emphasize local shape. Tomography adds posterior surface and thickness information.
Vertex-convert both principal powers at the measured spectacle vertex distance before reconstructing a corneal-plane spherocylinder.
Pre-fitting assessment
Ask about previous lens wear, comfort, wearing schedule, work environment, vision needs and the ability to insert, remove and care for lenses. Inspect the lid margins, tear film and cornea at the slit lamp. Upper lid eversion helps identify tarsal papillae. Interpret their appearance with symptoms, mucus, lens movement and other signs. Address active inflammation or epithelial damage before routine fitting. Suspected infection requires urgent eye-care assessment.
| Observation | How to use it |
|---|---|
| Tear breakup | A short measured time supports tear instability. TFOS DEWS III's under 10 s criterion refers to noninvasive breakup time in its diagnostic algorithm. Fluorescein volume and technique affect invasive breakup time. |
| Tear meniscus | A low meniscus can suggest an aqueous-deficient contribution. A preserved meniscus does not exclude a mixed disorder. |
| Lid margin and gland expression | Plugging and thick meibum support a meibomian gland contribution. Assess corneal staining and comfort before deciding on a lens trial. |
| Cornea and tarsal conjunctiva | Significant epitheliopathy, active keratitis or symptomatic tarsal inflammation can require deferral and further evaluation. Assess the underlying condition before choosing a replacement schedule. |
Dubai setting. ST-60 lists slit lamp and keratometry among optical-center equipment and training requirements. Its equipment section identifies tonometers, fundus cameras and OCT among instruments that should not be present in an optical center. Follow DHA HP-02 on professional privileges and facility services. Arrange corneal imaging or specialist assessment when the required service is beyond the current facility. Follow ST-60 §11 for contact items, equipment rests and diagnostic solutions.
Keratometry and reported power
A keratometer measures the reflection of known targets from the anterior corneal tear surface. The reflected image supplies an estimate of curvature within the instrument's sampling zone. A conventional keratometric reading converts anterior radius to an estimated corneal power using a calibration index rather than separately measuring the two corneal surfaces.
With :
A shorter radius gives a steeper K and a larger positive value. Some devices offer other indices, so record the setting. The approximation can be misleading when the relation between the anterior and posterior curvatures has changed, for example after LASIK. Check tear quality, alignment and repeatability before recording values.
Worked example. An horizontal radius gives at 180 and a vertical radius gives at 090. The difference is , with the vertical meridian steeper. This is with-the-rule anterior corneal toricity. It does not by itself establish the patient's refractive cylinder.
Instrument range. The Topcon KR-800, for example, specifies radii from 5.00 to 10.00 mm, corresponding to 67.50 to 33.75 D with the 1.3375 index. A 10.25 mm radius would calculate as about 32.93 D, outside that model's range. If a reliable reading cannot be obtained, follow the particular instrument's instructions or use another calibrated method. Do not extrapolate from an out-of-range display.
Topography and tomography
| Map | What it shows | Interpretation |
|---|---|---|
| Axial or sagittal curvature | Curvature referenced to the instrument's axis | Useful for broad patterns and a regular astigmatic bow tie. May smooth or shift the appearance of localized steepening. |
| Tangential or instantaneous curvature | More local curvature | Useful for locating focal steepening and transitions. Repeat poorly acquired maps and review the tear surface, fixation and scale. |
| Tomographic elevation and pachymetry | Elevation relative to a reference surface and corneal thickness distribution | Helps assess possible ectasia using anterior and posterior surface and thickness information. Read outputs in the context of the device and clinical exam. |
A Placido image is based on anterior surface reflections. It cannot directly map posterior corneal elevation. Compare repeatable topography and, when indicated, tomography before interpreting a focal finding. Lens-related warpage, tear disruption and ectasia can all affect the anterior map. A historical videokeratography study tested central power above and/or inferior–superior asymmetry above as screening features. These are study criteria. A single cutoff or map feature does not diagnose keratoconus. Progressive irregular refraction or suspicious asymmetry warrants appropriate corneal evaluation.
Vertex conversion and selection of a trial
The corneal-plane approximation for distant viewing is
where is spectacle power and is the measured positive distance in meters in front of the cornea. A spectacle minus correction usually needs less minus at the corneal plane. A spectacle plus correction usually needs more plus. For example, one manufacturer's toric fitting guide calls for vertexing when the magnitude of power in either principal meridian exceeds . Use the applicable fitting guide and vertex distance.
Worked example. Convert from 12 mm in front of the cornea. The spectacle powers are at 180 and at 090. The corneal estimates are and . Using the unrounded values puts the cylinder at about , giving approximately . Select available trial parameters and verify fit, acuity, axis and over-refraction. The optical estimate does not specify a manufactured lens.
A spherical equivalent averages the two powers and leaves some astigmatic defocus. Compare spherical and toric options based on measured acuity, visual demands and on-eye behavior. A randomized trial found better acuity with toric lenses than spherical lenses in its studied range of 0.75–1.75 D of astigmatism. The starting lens still depends on the individual fit and response.
Sources
- TFOS DEWS III diagnostic methodology, tear testing
- DHA ST-60 sections 9 and 11
- DHA Clinical Privileging Policy HP-02 v2
- Topcon KR-800 manufacturer specifications
- Study of keratometric and total corneal power after LASIK
- Comparison of corneal cone locations estimated with axial and tangential maps
- Videokeratography screening comparison
- Global Consensus on Keratoconus and Ectatic Diseases, edition 2
- Bausch + Lomb INFUSE fitting guide, principal-meridian vertex conversion
- Randomized comparison of toric and spherical soft lenses
Sources checked November 25, 2026.
A spectacle correction of −8.00 −3.50 × 180 is measured at a 12 mm vertex distance. Which is the closest corneal-plane spherocylinder if sphere and cylinder are each rounded to the nearest quarter diopter?
An instrument calibrated at a keratometric index of 1.3375 measures an anterior corneal radius of 6.00 mm. What keratometric power does this radius represent?
41.50 D
48.75 D
52.75 D
56.25 D
Which anterior curvature map is usually more informative for localizing a small region of steepening that looks smoothed on the axial map?
Tangential (instantaneous) curvature map
Axial (sagittal) curvature map
Refractive power Snell's law map
Simulated keratometry (SimK) summary value
During a pre-fitting slit-lamp exam, you observe a fluorescein tear break-up time of , inspissated meibomian gland orifices with toothpaste-like secretions, and a tear meniscus height of . Which interpretation and pre-fitting management plan is most accurate?
Severe aqueous tear deficiency is present; fit a high-water-content conventional hydrogel lens immediately
The ocular surface is normal for extended-wear silicone hydrogel fitting because meniscus height exceeds
Tear instability with a likely meibomian gland contribution. Address the ocular surface and reassess suitability before trial wear.
An absolute contraindication to all contact lens wear exists due to neurotrophic keratopathy
Sections you finish are checked off in the contents.