5.5 Labeling, Identification & Color Codes

Key Takeaways

  • Pipe labels must show the gas name or chemical symbol and its color code, and they must appear at least every 20 ft, at least once in or above every room, on both sides of walls penetrated, and at least once per story on risers.

  • Medical gas piping may not be painted.

  • Standard colors include oxygen green/white (or white/green), medical air yellow/black, nitrous oxide blue/white, nitrogen black/white, vacuum white/black, WAGD violet/white, and instrument air red/white.

  • Shutoff valve labels show the gas name or symbol, color code, rooms or areas served, and a caution not to close or open the valve except in an emergency.

  • Zone valve box labels must be visible through the cover or replicated outside, but never attached to the removable cover.

Last updated: October 2026

The Master Table: Names, Colors and Pressures

NFPA 99 Table 5.1.11 sets the designation colors and standard pressures. Pipe labels, valve labels, and outlets all use it.

Gas serviceAbbreviationColors (background/text)Standard pressure
Medical airMed airYellow/black50-55 psi
Carbon dioxideCO2Gray/black or gray/white50-55 psi
HeliumHeBrown/white50-55 psi
NitrogenN2Black/white160-185 psi (standard for support gas)
Nitrous oxideN2OBlue/white50-55 psi
OxygenO2Green/white or white/green50-55 psi
Oxygen/carbon dioxide mixtureO2/CO2 n%Green/white50-55 psi
Medical-surgical vacuumMed vacWhite/black15-30 in. HgV
Waste anesthetic gas disposalWAGDViolet/whiteVaries with system type
Medical-surgical vacuum/WAGD combinationMed-surg/WAGDWhite/black and violet/white15-30 in. HgV
Instrument air—Red/white160-185 psi (standard for support gas)
Other mixturesGas A%/Gas B%Major gas for background, minor gas for textNone
Nonmedical air and dental air—Yellow and white diagonal stripe/blackNone
Nonmedical vacuum and dental vacuum—White and black diagonal stripe/black boxedNone
Laboratory air—Yellow and white checkerboard/blackNone
Laboratory vacuum—White and black checkerboard/black boxedNone

The valve and outlet labeling rules treat 50-55 psi as standard for patient gases and 160-185 psi as standard for nitrogen and instrument air. Anything outside those ranges is a nonstandard pressure and must be stated on the label.

Note

CGA C-9 sets cylinder color marking, and its colors are not always the same as the pipeline label colors. On a pipeline, the label text, meaning the gas name or symbol, is what identifies the service. Never identify a pipe or a cylinder by color alone.

Pipe Labeling

Content. Piping is labeled with stenciling or adhesive markers that show:

  1. The name of the gas or vacuum system, or its chemical symbol, per Table 5.1.11
  2. The system's color code

Where positive-pressure piping operates at a nonstandard pressure, the label also shows the operating pressure. Where a vacuum system also serves WAGD, the piping in the immediate area of the WAGD system is labeled to show both systems.

Location. Pipe labels must be placed:

RuleExample
At intervals of not more than 20 ft (6.1 m)A 90 ft corridor run needs labels at no more than 20 ft spacing, so at least five labels plus any others the rules below require
At least once in or above every roomA short pass-through above a storage room still gets a label
On both sides of walls or partitions the piping penetratesOne label on each side of a fire-rated wall
At least once in every story height a riser passes throughA riser crossing eight floors gets at least eight labels

Never paint medical gas piping. Paint hides tube markings, labels, and leaks.

Intakes, exhausts and relief vents. Piping for compressor intakes, vacuum exhausts, and relief valve vent lines is labeled with the same content and its specific function, so nobody mistakes it for patient supply piping. The medical air intake itself is labeled so it is clearly identified as a medical air intake.

Shutoff Valve Identification

Every shutoff valve label must include:

  1. The name or chemical symbol of the gas or vacuum system
  2. The system's color code
  3. The rooms or areas served
  4. A caution not to close or open the valve except in an emergency

For nonstandard pressures, the valve label also shows the operating pressure. Where a vacuum system serves WAGD, valves on the source side of the WAGD connection are labeled for both systems.

Required wording by valve type (in substance)

ValveLabel reads in substance
Source valveSOURCE VALVE FOR THE (source or gas name)
Main line valveMAIN LINE VALVE FOR THE (gas or vacuum) SERVING (name of building)
Riser valveRISER FOR THE (gas or vacuum) SERVING (area or building served by that riser)
Service valveSERVICE VALVE FOR THE (gas or vacuum) SERVING (area or building served by that valve)
Zone valve boxZONE VALVES FOR THE (gas or vacuum) SERVING (rooms or spaces served)

The zone valve box label must be visible from outside through the cover, or replicated on the outside, but it must not be fixed to the removable cover.

Station Outlet and Inlet Identification

Each outlet and inlet is identified with:

  1. The name or chemical symbol of the gas or vacuum
  2. The color code

Outlets at nonstandard pressures show the operating pressure in addition to the gas name. In sleep labs, outlets downstream of a flow control device carry a warning not to use them for ventilating patients.

Buried and Concealed Piping

  • Buried piping outside buildings gets a continuous tape or marker immediately above the pipe that names the pipeline.
  • During system inspection before concealment, the inspector confirms that labeling and valve tagging are present and correct for all concealed components and piping (Section 7.2). Labels that go missing behind a finished ceiling are hard to fix later.
  • The verifier checks labeling again and confirms through the cross-connection and valve tests that each outlet and valve label matches the piping.
Test Your Knowledge

What color combination does NFPA 99 assign to waste anesthetic gas disposal (WAGD) piping labels?

A

White/black

B

Violet/white

C

Red/white

D

Brown/white

Test Your Knowledge

A 100 ft run of medical air piping passes above a corridor with no walls or rooms along the way. What is the maximum spacing NFPA 99 allows between pipe labels?

A

10 ft

B

25 ft

C

50 ft

D

20 ft

Test Your Knowledge

Where may the label for a zone valve box assembly be placed?

A

Visible through or copied outside the cover, never on the removable cover

B

Only on the removable cover itself, so it is visible from the corridor

C

Only inside the box, attached to each valve handle where staff will see it

D

On the ceiling grid above the box