1.5 Key NFPA 99 Definitions & Terminology

Key Takeaways

  • A station outlet is the point on a positive-pressure gas system where users connect and disconnect; a station inlet is the matching point on a vacuum system.

  • In this code, the patient or use side of a pressure gas system is downstream, toward the outlets, and the source side is upstream, toward the central supply.

  • A zone valve controls gas or vacuum to one area, and a source valve isolates an entire central supply system, including its dryers and final line regulators.

  • Medical support gas means nitrogen or instrument air used to power equipment; it is never breathed as part of treatment.

  • WAGD is the capture and removal of gases vented from the patient breathing circuit during normal operation of anesthesia or analgesia equipment.

Last updated: October 2026

Why Definitions Get Their Own Domain

NITC weights Definitions at 8% of the written exam, and most other questions assume you read code terms precisely. Many wrong choices on medical gas exams swap one similar term for another: a riser valve for a service valve, a station outlet for a station inlet, or medical air for instrument air. The definitions below are summarized in plain language from NFPA 99's terminology. On the job, the exact wording in the adopted edition controls.

System Hardware

TermPlain-language meaning
Central supply systemThe complete source of a medical gas or vacuum, such as a cylinder manifold, cryogenic system, compressor or pump package, or concentrator system
HeaderA pipe assembly that connects cylinders or containers to a central supply, with its leads, filter, check valves, and shutoff valve
ManifoldEquipment that joins cylinders or containers through headers and switches automatically between the primary and secondary headers
Primary / secondary / reserveSupply in use / supply that takes over when the primary is depleted / emergency supply that serves when the others cannot
Average day's supplyThe facility's normal daily consumption, used to size headers and to trigger the low-contents alarms
Source valveA shutoff valve at the immediate connection of a central supply system that isolates the whole source
Main line valveA shutoff valve in the main supply line inside the building served
Riser valveA shutoff valve in a riser, next to the main line
Service valveA shutoff valve in branch (lateral) piping off a riser, ahead of the zone valve boxes on that branch
Zone valveA valve that controls the gas or vacuum to a particular area
In-line valveAn optional valve that isolates part of a branch, such as individual rooms, for service
Auxiliary connectionA tee, valve, and removable cap or plug for connecting a temporary supply
EOSCEmergency oxygen supply connection: an exterior inlet for a temporary oxygen source

Terminal Devices

TermPlain-language meaning
Station outletThe point on a positive-pressure gas pipeline where the user makes connections and disconnections
Station inletThe point on a vacuum pipeline where the user makes connections and disconnections
Primary / secondary valveIn an outlet, the secondary valve closes automatically when the primary valve assembly is removed
DISSDiameter Index Safety System; noninterchangeable threaded connections covered by CGA V-5
Quick couplerA gas-specific, noninterchangeable connector that is not threaded
Demand checkA self-closing fitting that lets a device be disconnected without releasing gas
Manufactured assemblyA factory-built device such as a headwall, column, boom, or medical gas rail with outlets already piped

Directions and Places

  • Source side vs. patient (use) side: for pressure gases, the source side is upstream and the patient side is downstream, toward the outlets. "Install the pressure indicator on the patient side of the zone valve" means downstream of it.
  • Anesthetizing location: an area the governing body designates for administering inhalation anesthetic agents.
  • Category 1-4 spaces: patient care spaces classified by the governing body's risk assessment.
  • Breached system: a system opened by pipeline intrusion, physical separation, or removing, replacing, or adding a component. It must be inspected and tested again.

Gases and Purity Terms

  • Patient medical gas: a piped gas such as oxygen, nitrous oxide, helium, carbon dioxide, or medical air, used for human respiration or for calibrating respiratory devices.
  • Medical support gas: nitrogen or instrument air used for medical support purposes such as running surgical tools, booms, and pendants. It is not breathed as part of treatment.
  • Nitrogen NF / oil-free, dry nitrogen: nitrogen meeting the National Formulary standard, free of moisture and hydrocarbons. It is the required brazing purge gas and test gas.
  • Oxygen USP vs. Oxygen 93 USP: cylinder or cryogenic oxygen tested to at least 99% purity, versus concentrator oxygen tested to 90-96%.
  • Medical air: air from cylinders, bulk containers, or medical air compressors, or reconstituted from oxygen USP and nitrogen NF, that meets the USP medical air monograph.
  • Medical-surgical vacuum: a method that provides drainage, aspiration, and suction to remove body fluids from patients.
  • WAGD (waste anesthetic gas disposal): capturing and carrying away gases vented from the patient breathing circuit during normal operation of anesthesia or analgesia equipment.

People

  • Authority having jurisdiction (AHJ): the organization, office, or individual that enforces the adopted code.
  • Governing body: the health care organization's leadership, which sets risk categories and designates spaces.
  • Responsible facility authority: the person the facility names to operate and manage its medical gas systems.

Testing Vocabulary

TermMeaning
BlowdownClearing newly installed piping with nitrogen NF before outlets and components are installed
Initial pressure testHolding piping at 1.5 times operating pressure (at least 150 psi) while every joint is checked for leaks
Standing pressure testA long hold (24 hours for installers, 10 minutes for verifiers) to detect leakage
Cross-connection testConfirming that each outlet delivers only its own system's test gas
Piping purge testIntermittent high-volume flow at each outlet until a white cloth shows no discoloration
Particulate testFlowing a measured gas volume through a fine filter to check cleanliness
Purity testComparing hydrocarbon and moisture levels at a remote outlet with the source gas
Final tie-in testLeak testing and purging after new piping is joined to a live system
Operational flow pressure drop testConfirming outlets and inlets deliver the required flow without excessive pressure or vacuum loss
Test Your Knowledge

What is the difference between a station outlet and a station inlet in NFPA 99 terminology?

A

An outlet is mounted in a wall and an inlet is mounted in a ceiling column or boom

B

An outlet is a DISS fitting and an inlet is a quick coupler

C

An outlet has no secondary valve, while an inlet always has both primary and secondary valves

D

An outlet serves a pressure gas system; an inlet serves a vacuum system

Test Your Knowledge

Nitrogen piped at 175 psi to power surgical drills and articulating booms is classified as which of the following?

A

A patient medical gas

B

A medical support gas

C

Medical air

D

A nonmedical laboratory gas

Test Your Knowledge

A drawing note says to install a pressure gauge on the patient side of the zone valve for an oxygen line. Where does the gauge go?

A

Downstream of the zone valve, between it and the outlets

B

Upstream of the zone valve, between the valve and the riser

C

At the source valve in the supply room

D

On the master alarm panel